SeaISO™ USB Isolator

 

User Manual

 

 
 
 
 

 

 

 

 
 

 

Item# 

ISO-1R 
 

Item# 

ISO-1 
 

Summary of Contents for SeaISO ISO-1

Page 1: ...SeaISO USB Isolator User Manual Item ISO 1R Item ISO 1 ...

Page 2: ...tallation Instructions 5 Specifications 7 Dimensions 7 Power 7 Environmental Specifications 7 Certifications Approvals 7 Product Marking 7 Manufacturing 7 Appendix D Drawings 8 Appendix E How to Get Assistance 10 Sealevel Systems 10 Technical Support 10 Warranty 11 Warranty Policy 11 Non Warranty Repair Retest 11 How to Obtain an RMA Return Merchandise Authorization 11 ...

Page 3: ... or a situation that could cause the product to fail Note The lowest level of importance used to provide background information additional tips or other non critical facts that will not affect the use of the product IMPORTANT SAFETY INSTRUCTIONS SAVE THIS MANUAL It contains important safety and operating instructions for the installing and using the SeaISO USB Isolator Prior to beginning the insta...

Page 4: ...version featuring cable bump design and sealed TPE overmold for environmental protection in medical applications Features Medical grade 4000Vac isolation data and power Meets 60601 1 3rd Edition requirements for 1x MOPP at working voltage up to 250Vac Meets 60601 1 3rd Edition requirements for 2x MOPP at working voltage up to 42 4Vac High retention connectors USB bus powered High speed host port 1...

Page 5: ...USB A B cable 1m CA 539 INSTALLATION INSTRUCTIONS Connect B type USB connector on the isolator to the host computer via the included A B type 1m long USB cable CA 539 Connect CA 536 cable to the A type female USB connector on the isolator Connect your desired USB device to the A type female end of the CA 536 cable Host computer should detect a new device and start enumeration After enumeration com...

Page 6: ...ed in a bump in the cable INSTALLATION INSTRUCTIONS Connect A type male USB connector of the isolator to the host computer Connect your desired USB device to the A type female connector of the isolator Host computer should detect a new device and start enumeration After enumeration completes and appropriate device drivers are loaded if required your USB device is ready to use ...

Page 7: ...Temperature Range 20ºC to 60º C 40º to 85º C Humidity Range 10 to 90 R H Non Condensing 10 to 90 R H Non Condensing CERTIFICATIONS APPROVALS UL Recognized in accordance with 60601 1 3rd Ed E474084 FCC Part15 Subpart B ICES 003 CE marking o EMC Directive 2004 108 EC o RoHS Directive 2011 65 EU PRODUCT MARKING MANUFACTURING All Sealevel Systems printed circuit boards are built to UL 94V0 rating and ...

Page 8: ... Sealevel Systems Inc SeaISO USB Isolator 8 SL9261 8 2015b Appendix A Drawings2 ISO 1 CA 536 1 All dimensions in millimeters ...

Page 9: ... Sealevel Systems Inc SeaISO USB Isolator 9 SL9261 8 2015b CA 539 ISO 1R ...

Page 10: ...uals links located under Technical Support Manuals and software can also be downloaded from the product page for your device The FAQ section of our website answers many common questions Refer to this helpful resource by visiting www sealevel com faq asp TECHNICAL SUPPORT Monday Friday 8 00 am to 5 00 pm EST Phone 1 864 843 4343 Email support sealevel com RETURN AUTHORIZATION MUST BE OBTAINED FROM ...

Page 11: ...he Product failure to adhere to any specifications or instructions or failure resulting from neglect abuse accidents or acts of nature are not covered under this warranty Warranty service may be obtained by delivering the Product to Sealevel and providing proof of purchase Customer agrees to insure the Product or assume the risk of loss or damage in transit to prepay shipping charges to Sealevel a...