eNodTouch 

Touch screen for eNod4 controllers 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 

 

 

 

U

se

M

a

n

u

a

Summary of Contents for eNodTouch

Page 1: ...NU eNodTouch E 0413 SCAIME SAS Technosite Alt a BP501 74105 ANNEMASSE Cedex T l 33 0 4 50 87 78 64 www scaime com eNodTouch Touch screen for eNod4 controllers User Manual...

Page 2: ...tartup screens 9 3 2 Main screen 10 3 2 1 Transmitter mode eNod4 T eNod4 C eNod4 D 10 3 2 2 Checkweigher mode eNod4 C 10 3 2 3 Dosing mode eNod4 D 10 3 2 4 Keyboard functions 11 3 3 Result displaying...

Page 3: ...prox Panel cut dimension W105 x H66 mm Panel thickness area 1 to 5 mm Surrounding air temperature 0 to 50 C Ambient humidity 10 to 90 RH Wet bulb temperature 39 or less no condensation Dust 0 1 mg m o...

Page 4: ...anel of a cabinet or enclosure Surrounding air temperature indicates both the et or enclosure s internal temperature eNodTouch installation Panel thickness eNodTouch has two projections on the top to...

Page 5: ...nnection to eNod4 The serial interface is not isolated Always connect the SG Signal Ground of the unit to the connected device SG Signal Ground and FG Frame Ground are connected internally RS422 RS485...

Page 6: ...n after the startup screen is displayed When Operating Touch the upper right then bottom left corners or the upper left and bottom right corners of the panel within a 40 pixel area in 0 5 seconds When...

Page 7: ...le the in built Buzzer sound upon a touch Disable Display setting Show Brightness contrast control bar Specify whether to display on touch the Control Bar to adjust brightness and contrast Enable D Sc...

Page 8: ...fo Manufacturer Schneider Electric SA Series MODBUS SIO Master N of Device 1 Note Never change the parameters in grey 2 2 4 Password menu Menu Input Parameter Description Default System Password Syste...

Page 9: ...es of network communication 3 On Line Functioning 3 1 Startup screens At Power On the connection screen is displayed while eNodTouch try to communicate with connected eNod4 device If the connection sc...

Page 10: ...ouch 3 2 3 Dosing mode eNod4 D To switch from the Bar graph to the Status bar touch the Weight value area To switch from the main keyboard to the 2 nd keyboard touch Bar graphe Gross Net indicator Sta...

Page 11: ...t stability condition START DOSING Start the dosing cycle if starting conditions are satisfied dosing cycle is ongoing STOP DOSING Stop the dosing cycle the dosing counter is not incremented SUSPEND D...

Page 12: ...ore information about parameters consult eNod4 software manuals Setup menu access can be secured by password see 3 4 1 In that case the following screen is displayed After entering a valid password or...

Page 13: ...tion loads used during the physical calibration load 1 2 3 1 1 000 000 Weight values corresponding to each calibration segment Zero Calibration Allows acquiring stable measurement with no load on the...

Page 14: ...logical inputs before their activation Output 1 4 Parameters Function Function assigned to the related output None Thresholds Stability Coarse feed High feed fine feed Cycle ongoing result available O...

Page 15: ...ter with frequency rejection of 50 Hz or 60 Hz harmonics Frequency 6 1600 Hz Filtered measurement frequency is adjustable from 6 25 to 1600 Hz Bessel low pass filter parameters Filter Order 0 2 3 4 Fo...

Page 16: ...ng measurements during the specified time Automatic Dyn Zero Yes No Automatic zero correction allows to dynamically track the Zero evolution When enabled an average Zero value is continuously calculat...

Page 17: ...mode Auto Manual In filling mode define the emptying mode In manual mode emptying start after input Manual Emptying activation Emptying end level 0 1 000 000 The emptying output is disabled if the net...

Page 18: ...Page 1 3 Start cycle delay 0 65535 ms This stabilization time precedes the tare execution Meas Stab delay This stabilization time follows the FF stop After this delay a new dosing result is calculate...

Page 19: ...o FF level Page 2 2 Correction 0 100 If Automatic correction enabled the correction formula is In flight In flight Dosing result Target weight x Correction X3 correction if Out of Tolerance Yes No If...

Page 20: ...represents the value to reach at the end of the dosing process Used in sequences using at least 2 or 3 feeds Defines the transition threshold relative to the target weight Coarse feed and Fine Feed p...