background image

IDM 98IV

 & 

99IV / EN

 

1  

Warning

  

. . . . . . . . . .

 . . . . . . . . . .
 . . . . . . . . . .
 . . . . . . . . 

Safety sheet

 . . . . . . . . .  
 . . . . . . . . . .
 . . . . . . . . . .
 . . . . . . . .

 . . . . . . . . .

 

 

Read First

 

 

 Safety Information

 

 

Understand and follow operating instructions carefully. 
Use the meter only as specified in this manual; otherwise,  
the protection provided by the meter may be impaired. 

 

 

 WARNING

 

Identifies hazardous conditions and actions that could cause  

BODILY  HARM 

or 

DEATH 

 

 

 CAUTION 

Identifies conditions and actions that could 

DAMAGE

 the meter  

or equipment under test 

 

 

 WARNING 

˙

 When using test leads or probes, keep your fingers behind  

     the finger guards. 

˙

 Remove test lead from Meter before opening the battery  

     door or Meter case. 

˙

 Use the Meter only as specified in this manual or the protec- 

     tion by the Meter might be impaired.   

˙ 

Always use proper terminals, switch position, and range for  

     measurements. 

˙ 

Verify the Meter’s operation by measuring a known voltage.  

     If in doubt, have the Meter serviced. 

˙

 Do not apply more than the rated voltage, as marked on  

     Meter, between terminals or between any terminal and  

     earth ground.  

˙

 Only replace the blown fuse with the proper rating as  

     specified in this manual. 

˙

 Use caution with voltages above 30 Vac rms, 42 Vac peak,  

     or 60 Vdc. These voltages pose a shock hazard.   

˙

 To avoid false readings that can lead to electric shock and  

     injury, replace battery as soon as low battery indicator. 

˙

 Disconnect circuit power and discharge all high-voltage  

     capacitors before testing resistance, continuity, diodes, or  

     capacitance. 

˙

 Do not use Meter around explosive gas or vapor. 

˙ 

To reduce the risk of fire or electric shock do not expose this  

     product to rain or moisture.  

˙ 

Wear suitable personal protective equipment when working  

    around or near hazardous live conductors which could be  

    accessible. 

 Warning

 

Safety sheet

 

Summary of Contents for 873-2367

Page 1: ...measurements Verify the Meter s operation by measuring a known voltage If in doubt have the Meter serviced Do not apply more than the rated voltage as marked on Meter between terminals or between any...

Page 2: ...ter in mA A function to measure the voltage of a power supply circuit in equipment that could result in damage the meter and the equipment under test Symbols as marked on the Meter and Instruction man...

Page 3: ...isplay 6000 counts 2 Push buttons 3 Rotary switch for turn the Power On Off and select the function 4 Input Terminal for A 5 Input Terminal for V functions 6 Common Input Terminal 7 Input Terminal for...

Page 4: ...function AC DC function Frequency Counter on AC mode Capacitance Measurement Temperature for 99IV only Low battery indicator with segments Auto Power Off 20 minutes Shock proof from 4 feet drops CAT...

Page 5: ...easurements Caution When connecting the test leads to the DUT Device Under Test connect the common test leads before connecting the live test leads when removing the test leads remove the live test le...

Page 6: ...nd press the Function button to select the measuring function See Using the Function Button Measuring AC DC Current Frequency Disconnect AC Disconnect Rotate the switch and press the Function button t...

Page 7: ...e Resistance Capacitance Rotate the switch and press the Function button to select the measuring function See Using the Function Button Measuring Continuity Diode Rotate the switch and press the Funct...

Page 8: ...IDM 98IV 99IV EN 8 Measuring Temperature C F for 99IV only Rotate the switch and press the Function button to select the measuring function C F See Using the Function Button...

Page 9: ...IV EN 9 Using The Function Buttons Function Button Switch Position Function Press AC Press button 2sec to Auto Range Mode Press the Function button to change the function on the same switch position R...

Page 10: ...r 99IV only Min Reading Max Reading Press 2 sec Present Reading Press button over 2 sec then exit this function In the Peak HOLD function the meter records the peak min value and the peak max value wh...

Page 11: ...arger than the display reading by 50 counts However it can not detect across the AC and DC Voltage Current VoltSeek Press the VoltSeek button to activate the VoltSeek function Caution The number of da...

Page 12: ...o turn the backlight on off Auto Power Off 2 1 Reactivate the meter by rotating the switch or by pressing any button Auto Power Off Disable After idle 20 min Rotate the switch to the OFF position Pres...

Page 13: ...ate the switch to the OFF position Press and hold down the SmartHOLD button and turn the meter on Default Temperature Units for 99IV only 2 1 2 1 2 1 Rotate the switch to the OFF position Press and ho...

Page 14: ...nserting the batteries 4 Do not short circuit used batteries disassemble them or throw them in a fire Doing so may cause the batteries to explode 5 Dispose of the used batteries in accordance with loc...

Page 15: ...eplace the battery as soon as the low battery indicator appears to avoid false reading Refer to the following figure to replace the batteries 9V Battery Battery Cover Screw Caution Remove test leads f...

Page 16: ...3 times per second Altitude 6561 7 ft 2000m Safety Complies with EN61010 1 UL61010 1 IEC 61010 1 CAT IV 600V CAT 1000V Measurement Category Application I Measurements on circuits not directly connect...

Page 17: ...80 RH AC Function ACV and ACA specifications are ac coupled true R M S The crest factor may be up to 3 0 at 4000 counts For non sinusoidal waveforms Additional Accuracy by Crest Factor C F Add 3 0 for...

Page 18: ...0 01mA 1 5 3d 1 2 3d 600 0mAAC 0 1mA 1 5 3d 1 2 3d AC 600 0mV ranges are specified from 1 of range to 100 of range Input Impedance 10M less than 100pF Overload Protection AC DC 1000V Range Resolution...

Page 19: ...s rest time 10A for max 30 seconds with at least 10 minutes rest time AC Frequency Response 45 1kHz Sine Wave Overload Protection AC DC 11A 7 Peak Hold For 99 only Specified accuracy 150 digits for 60...

Page 20: ...1000V 12 Frequency Counter Range Resolution Accuracy 100 00 Hz 0 01 Hz 0 1 2d 1000 0 Hz 0 1 Hz 10 000 KHz 0 001 KHz 100 00 KHz 0 01 KHz Minimum Sensitivity 6V for ACV 1Hz 10kHz 12V for ACV 10kHz 50kHz...

Page 21: ...lution For 98IV Accuracy For 99IV Accuracy 60 00mA 0 01mA 2 5 5d 2 5d 600 0mA 0 1mA Range Resolution For 98IV Accuracy For 99IV Accuracy 6 000A 0 001A 2 5 5d 2 5d 10 00A 0 01A Additional specification...

Page 22: ...repair alteration contamination or abnormal conditions of operation or handling Any implied warranties arising out of the sale of this product including but not limited to implied warranties of mercha...

Page 23: ...Road Shanghai 200001 China www rs components com Europe RS Components Ltd PO Box 99 Corby Northants NN17 9RS United Kingdom www rs components com Japan RS Components Ltd West Tower 12th Floor Yokoham...

Page 24: ...EN Instruction Manual IDM 98IV IDM 99IV Digital Multimeter...

Reviews: