Operating Instructions

 

Pharmaceutical Refrigerator 

MPR-721R 

Summary of Contents for MPR-721 Series

Page 1: ...es Please read the operating instructions carefully before using this product and keep the operating instructions for future use See page 32 for all model numbers Operating Instructions Pharmaceutical Refrigerator MPR 721R ...

Page 2: ......

Page 3: ...ETTING OF DELAY OF DOOR ALARM 18 SETTING OF RINGBACK OF ALARM BUZZER 19 OPERATION CHECK AFTER RECOVERY 19 DEFROST CYCLES 20 REMOTE ALARM TERMINAL 20 ALARMS SAFETY FUNCTIONS 21 ROUTINE MAINTENANCE 22 Cleaning of cabinet 22 Cleaning of evaporating tray 22 Replacement of fluorescent lamp 23 Replacement of glow starter 23 TROUBLESHOOTING 24 DISPOSAL OF UNIT 24 Decontamination of unit 24 OPTIONAL COMPO...

Page 4: ...ment of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or page order is incorrect Contact our sales representative or agent if any point in the operating instructions is unclear or if there are any inaccuracies No part of the operating instructions may be reproduced in any form without the expressed written permission of PHC Cor...

Page 5: ...ons in a place accessible to users of this unit Label on the unit This mark is labeled on the cover in which the electrical components of high voltage are enclosed to prevent the electric shock The cover should be removed by a qualified engineer or service personnel only USA Only This product has a fluorescent lamp that contains mercury Disposal may be regulated in your community due to environmen...

Page 6: ...ower supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may cause electric shock in the case of an incomplete circuit Connect the unit to a power source as indicated on the rating label attached to the unit Use of any other voltage or frequency other than th...

Page 7: ...ay cause electric shock or fire Do not position this unit and the other unit so that it is difficult to operate the disconnection of the power supply plug Failure to disconnect the power supply plug may cause fire if there is something wrong with the unit When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric...

Page 8: ...is in the unit except in a sealed container These may be detrimental to health and may cause corrosion of internal components cooling circuit or electrical components Check the setting when starting up of operation after power failure or turning off of power switch The stored items may be damaged due to the change of setting Be careful not to tip over the unit during movement to prevent damage or ...

Page 9: ... 40 o C Maximum relative humidity 80 for temperature up to 31 o C decreasing linearly to 50 relative humidity at 40 o C Mains supply voltage fluctuations up to 10 of the nominal voltage Transient overvoltages up to the levels of OVERVOLTAGE CATEGORY II Temporary OVERVOLTAGES occurring on the mains supply Applicable pollution degree of the intended environment POLUTION DEGREE 2 in most cases ...

Page 10: ...8 REFRIGERATOR COMPONENTS 1 8 9 2 13 4 12 6 11 MPR 721 7 5 3 16 15 17 18 10 14 19 MPR 721R 13 Back side 17 18 6 ...

Page 11: ...unit is level 12 Castor When installing the unit the castors can be raised from the ground by using the leveling feet above 13 Shelf MPR 721 only Items to be stored in the chamber must be placed on the shelves The maximum storage weight for each shelf is 50 kg Do not put stored items directly onto the interior floor of the chamber See page 15 Drawer MPR 721R only Items to be stored in the chamber ...

Page 12: ...is is for the fluorescent light It is recommended that the glow starter is also replaced when the fluorescent light is replaced Refer to page 23 2 Remote alarm terminal This is used to connect the unit to an exterior alarm to notify users of any malfunction Refer to page 20 1 2 ...

Page 13: ...Pressing this key in the set mode alters the digit to be set Key lock setting mode is led by pressing this key for more than 5 seconds in the temperature display mode Refer to page 16 for the key lock 6 Set key SET Temperature setting mode is initiated by pressing this key Once the key is pressed the digit to be changed will flash Pressing this key again after setting the desired temperature store...

Page 14: ...on or noise WARNING Install the unit on a sturdy floor If the floor is not strong enough or the installation site is not adequate this may result in injury from the unit falling or tipping over Select a level and sturdy floor for installation This precaution will prevent the unit from tipping Improper installation may result in water spillage or injury from the unit tipping over A location not pro...

Page 15: ... leveling feet by rotating them to contact them to the floor Ensure the unit is horizontal See Fig 1 3 Fixing the unit Two fixtures are attached to the rear of the frame Fix the frame to the wall by attaching a rope or chain between the wall and the fixtures See Fig 2 4 Ground earth The ground earth is for preventing the electric shock in the case of unexpected deterioration of the electrical insu...

Page 16: ... light switch to check the light After checking turn off the switch if the light is not necessary 6 Set the desired temperature When the chamber temperature gets to the set temperature begin slowly placing items into the chamber to minimize the temperature rise CAUTION The chamber is refrigerated by the forced circulation of cooled air inside the chamber Ensure that the air intake and exhaust vent...

Page 17: ... into the fan Air exhaust vent Do not block this vent with containers as this can impede air circulation Items stored near the exhaust air vent can freeze when the refrigerator is operating at temperature settings lower than 5 o C Load line In this refrigerator storage space is limited to allow the passage of cool air around the chamber The label load line is fixed to both sides of the chamber Alw...

Page 18: ...when 90 seconds has passed without any key operation In this case the chamber temperature setting is not accepted The available set range of chamber temperature is between 2 o C and 23 o C The partial freezing may occurred if the set temperature is lower than 3 o C KEY LOCK FUNCTION This unit is provided with a key lock function When the key lock is ON change of temperature setting through the key...

Page 19: ...oll the figure to 1 When pressed the figure of settable digit changes 4 Press set key SET The first digit is flashed 5 Set the temperature to 003 with the digit shift key and numerical value shift key Pressing the key shifts the digit which can be set When pressed the figure of settable digit changes 6 Press set key SET Alarm temperature is memorized and the current chamber temperature is displaye...

Page 20: ...nge the delay time to 3 minutes from 2 minutes Description of operation Key operated Indication after operation 1 The current chamber temperature is displayed 2 Press numerical value shift key for 5 seconds F00 is displayed and the first digit flashed 3 Press numerical value shift key 4 times The display is changed to F04 from F00 4 Press set key SET The current setting 002 is displayed and the fi...

Page 21: ...ess numerical value shift key for 5 seconds F00 is displayed and the first digit is flashed 3 Press numerical value shift key 5 times The display is changed to F05 from F00 4 Press digit shift key once The second digit is flashed 5 Press numerical value shift key twice The display is changed to F25 from F05 6 Press set key SET The current setting 030 is displayed and the second digit is flashed 7 ...

Page 22: ...on the digital temperature display Once the forced defrost cycle is completed normal operation resumes CAUTION The unit may collect excessive frost on the evaporator if it is installed in high temperature and high humidity location For example the unit starts to defrost once a week with 2 o C setting in the ambient of 35 o C and 80 R H The chamber temperature goes up to approximately 10 o C tempor...

Page 23: ...yed alternately Intermittent tone Remote alarm Unit keeps continuous running If the defrost sensor is disconnected Alarm lamp is flashed E03 and chamber temp are displayed alternately Intermittent tone Remote alarm If the defrost sensor is short circuited Alarm lamp is flashed E04 and chamber temp are displayed alternately Intermittent tone Remote alarm If the compressor sensor is disconnected Ala...

Page 24: ...nit and all accessories If the outside panels are dirty clean them with a diluted neutral dishwashing detergent Undiluted detergent can damage the plastic components For the dilution refer to the instruction of the detergent After the cleaning with the diluted detergent always wipe it off with a wet cloth Then wipe off the cabinet or accessories with a dry cloth Never pour water onto or into the u...

Page 25: ...er with the wiring on the both sides of the lamp cover 5 Take out the fluorescent lamp and insert a new fluorescent lamp into the lamp cover Fig 2 6 Replace the water proof cover on the both sides of the lamp cover and then replace the lamp cover to the stopper 7 Replace the stored items on the top shelf and in the top drawer and connect the power supply cord to the outlet Replacement of glow star...

Page 26: ...he unit The air exhaust vent is blocked up with containers The unit is in direct sunlight The door is frequently opened There is a nearby heat source The ambient temperature is too high The door is not securely closed The heat source is put in the unit The door seal is damaged or foreign substance is inserted between door gaskets Note If the malfunction is not eliminated after checking the above i...

Page 27: ...εν πρέπει να απορριφθούν μαζί με οικιακά απορρίμματα Προκειμένου να πραγματοποιηθεί η κατάλληλη επεξεργασία και ανακύκλωση αποσύρετε αυτά τα προϊόντα σε ανάλογα σημεία περισυλλογής όπου θα γίνουν δεκτά χωρίς χρέωσή σας Εναλλακτικά σε ορισμένες χώρες μπορείτε να επιστρέψετε τα προϊόντα σε τοπικά καταστήματα με σκοπό την αγορά ανάλογου νέου προϊόντος Η ορθή απόρριψη αυτού του προϊόντος βοηθάει στην ...

Page 28: ... pertenecientes a la Unión Europea Este símbolo sólo es válido en la Unión Europea Si desea desechar este producto póngase en contacto con las autoridades locales o con su distribuidor para que le informen sobre el método correcto de eliminación Informacja dla użytkowników o pozbywaniu się urządzeń elektrycznych i elektronicznych dotyczy gospodarstw domowych Przedstawiony symbol umieszczony na pro...

Page 29: ...aardevolle hulpbronnen en voorkomt u potentiële negatieve gevolgen voor de volksgezondheid en het milieu die anders kunnen ontstaan door een onjuiste verwerking van afval Neem contact op met uw gemeente voor meer informatie over het dichtstbijzijnde inzamelingspunt of raadpleeg www nvmp nl www ictoffice nl of www stibat nl Voor zakelijke gebruikers in de Europese Unie Neem voor het weggooien van e...

Page 30: ... produkt Hvis du kasserer dette produktet på riktig måte bidrar til du til å bevare verdifulle ressurser og til å motvirke de negative virkningene på miljøet og den menneskelige helse som kan forårsakes av feilaktig avfallsbehandling Ta kontakt med de lokale myndigheter hvis du ønsker ytterligere informasjon om ditt nærmeste innsamlingspunkt Feilaktig kassering av dette utstyret kan kanskje bøtele...

Page 31: ... that each recorder needs a different fixture See the following for the details Temperature recorder Type Recording range Recorder fixing MTR 0621LH 30 days split recording paper type 6 o C to 20 o C MPR S30 MTR G04A MTR G04C 7 days circular recording paper type 10 o C to 40 o C MPR S7 WARNING Always disconnect the power supply to the unit prior to attachment of a temperature recorder in order to ...

Page 32: ...ure At the power failure the alarm activates blink of alarm indicator sound of buzzer to prompt an action for preventing the storage items Contact our sales representative or agent for the purchase of battery mounting box After installing the optional battery mounting box MPR 48B1 setting on the main unit is required Contact our sales representative or agent for setting ...

Page 33: ...d tube type Refrigerant R 134a Internal heater 153 W Temperature controller Microprocessor control system Temperature display Digital display Alarm High temperature Low temperature Door Power failure with optional MPR 48B1 installed Sensor Remote alarm contact Allowable contact capacity DC 30 V 2 A Memory back up Nonvolatile static memory Fluorescent lamp FL20SD FL20SSEXD for MPR 721 PE and MPR 72...

Page 34: ...to 14 o C ambient temperature 5 o C to 0 o C no load 2 o C to 23 o C ambient temperature 0 o C to 35 o C no load Temperature control range 2 o C to 23 o C Rated voltage AC 115 V AC 220 V AC 220 V AC 230 V 240 V Rated frequency 60 Hz 60 Hz 50 Hz 50 Hz Rated power consumption 400 W 330 W 280 W 305 W Noise level 41 dB A background noise 20 dB Maximum pressure 1516 kPa Note The unit with CE mark compl...

Page 35: ...Contamination of the unit Unit interior No contamination Yes No Decontaminated Yes No Contaminated Yes No Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on Yes No b There is some danger see below Yes No Procedure to be adhered to in order to reduce safety risk indicated in b below Date Signature Address Division Telephone Product name Pharmaceutic...

Page 36: ...Printed in Japan LDCL036300 1 S0418 11118 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018 ...