background image

II. User manual

1. Warnings

1) The long-term effects of chronic electrical stimulation are unknown.

2) Do not use stimulation over the main arteries around your neck. 

3) Do not use stimulation over the neck or mouth. Severe spasm of muscles may 

occur and the contractions may be strong enough to close the airway or cause 

difficulty in breathing.

4) Do not apply stimulation directly to the heart area across the chest or on the chest.

5) Do not use stimulation on your head or across your head.

6) Do not use stimulation over swollen, infected, or inflamed areas or skin eruptions, 

e.g., phlebitis, thrombophlebitis,varicose veins. 

7) Do not use stimulation over, or in proximity to, cancerous lesions.

8) Do not use stimulation when charging the device.

2. Precautions

1) Do not use the muscle stimulator during pregnancy.

2) Do not use if you have heart problems. 

3) Do not use if you have epilepsy.

4) Do not use the device in the presence of the following:

a. When there is a tendency to hemorrhage following acute trauma or fracture;

b. Following recent surgical procedures when muscle contraction may disrupt the 

healing process;

c. Over the menstruating or pregnant uterus; 

d. Over areas of the skin which lack normal sensation.

5) Do not use the products near the heart, it may cause fast or irregular heartbeat.

6) Some patients may experience skin irritation or hypersensitivity due to the 

electrical stimulation or electrical conductive medium. The irritation can usually be 

reduced by using an alternate conductive medium (like gel), or alternate electrode 

placement.

7) Pch Life TENS & PMS must be kept out of the reach of children.

8) Pch Life TENS &PMS should be used only with the leads and electrodes 

recommended for use by the manufacturer.

9) Never apply the pads to your skin with the power on, which will result in sudden 

shock. If, during application, you want to move the pads to another body part, please 

shut down the device first, and then move the pads to the place that you want to 

stimulate.

10) Never use this product near the following devices: pacemakers or any other 

embedded electronic medical devices, heart-lung machine and any other life keeping 

electronic medical devices, electrocardiograph and any other medical screening and 

monitoring devices. Simultaneous use of the Pch Life TENS & PMS and any of the 

above devices will cause malfunction and can be very dangerous to the users.

11) Two pads should be used together as a pair. Always peel off the protective film on 

the pads before use. To avoid an electrical short, do not connect two pads to each 

other.

12) Do not apply pads to the same spot for over 60 minutes at a time.

13) Do not use the Pch Life TENS & PMS while driving, operating machinery, or 

during any activity in which involuntary muscle contractions may put the user at 

undue risk of injury.

14) Do not use the product in bathroom or a moist environment. Do not use while 

bathing.

15) Do not use the product in the condition of air mixed with flammable gases. 

16) Never use this product in concurrence with high frequency surgical equipment; it 

may result in burns at the site of the stimulator electrodes and possible damage to the 

control unit.

17) Never use this product near a microwave oven, or other high frequency 

equipment.

18) Never use this product while it is being charged.

19) Do not operate the product in close proximity (e.g. 1m/3ft)) to shortwave 

therapy equipment, it may produce instability in stimulator output.

3. Contraindication

1) Do not use this device on patients who have a cardiac pacemaker, implanted 

defibrillator, or other implanted metallic or electronic device, because this may 

cause electric shock, burns, electrical interference, or death.

2) Do not use this device on patients whose pain syndromes are undiagnosed. 

3) Do not use this device during pregnancy.

4) Do not use this device on babies or infants who cannot express themselves.

4. Adverse Reactions

1) Patients may experience skin irritation and burns beneath the stimulation 

electrodes applied to the skin;

2) Patients may experience headache and other painful sensations during or 

following the application of electrical stimulation.

3) Patients should stop using the device and should consult with their physicians if 

they experience adverse reactions from the device.

5. Indications for Use

TENS:

To be used for temporary relief of pain associated with sore and aching muscles in 

the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) 

due to strain from exercise or normal household work activities. Choose Mode 1, 2

3, 5, 6.

PMS:

It is intended to be used to stimulate healthy muscles in order to improve and 

facilitate muscle performance. Choose Mode 1, 4, 5, 6.

6. Description of the Device

6.1 General Description of the Device

Pch Life TENS & PMS is a portable and DC 3.7V battery powered multifunction 

device, offering both Transcutaneous Electrical Nerve Stimulation (TENS) and 

Powered Muscle Stimulation (PMS) qualities in one device.

Pch Life TENS & PMS has 6 operation modes, which can give certain electrical 

pulses through electrode adhesive pads to the suggested area of the body where the 

electrodes are placed. 

The electronic stimulatory module has the operating elements of an ON/OFF Switch, 

Display screen, Mode Selection key, Intensity Modification keys, Timing key, Pause 

key, Output socket, and USB port for battery charging.

The display screen can show battery power, selected mode, current intensity, time 

remaining of an application mode, and indication of a pause (Page 10, No. 10).

The device is equipped with accessories of electrode pads, electrode cables, a battery 

charger, and one USB cable. The electrode cables are used to connect the pads to the 

device; the USB cable is used to connect the charger and the built-in lithium battery. 

All accessories, including USB cables, electrode pads, electrode cables, chargers can 

only be changed or replaced by a qualified person. 

The electrodes are interchangeable. The application area of electrode pads must be 

larger than the smaller electrode pads. The electrode pads are provided by GMDASZ 

Manufacturing Co., Ltd. with 510(k) cleared Number K092546.

6.2 Specifications and Essential performance:

(Essential performance: The values of pulse duration, amplitudes, and repetition 

frequencies do not deviate by more than ±30% when measured with an error not 

exceeding ±10% into a load resistance (500Ω) within the range specified by the 

manufacturer.)

- Power supply: DC 3.7V

- Output voltage: 42V@500Ω

- Output current: 84mA@500Ω

- Consumed current: 40mA

- Pulse width: 100μS

- Frequency: 1~110Hz

- Smallest are of electrode pad: 4cm2

- Timer: 10~60 minutes

- Strength level adjustment: 20 grades

- Charger: Input: 100-240V 50/60Hz 0.18A Max; Output: DC5V 250mA; Class 

II, not applied part, not suitable for use in the presence of a flammable gases, 

with oxygen or nitrous oxide, or for continuous operation..

6.3 Features

* A large, easy to read LCD display

* Adjustable timer

* 6 Modes

* Rechargeable lithium battery

6.4 Mode Description

We suggested that you initially experiment using each of the 6 modes. The mode that 

gives you the most desirable sensations and comfort is the most appropriate one to use 

for your current condition.

1) Intensity should be set at a level where you will feel some muscular vibration and 

involuntary muscle movements. 

It should be set to give a comfortable feeling and should not be painful.

2) Timing of application should be at least 10 to 20 minutes.

3) Frequency of the application should be at least one to three times per day.

Summary of Contents for PCHLIFE

Page 1: ...Power supply DC 3 7V Output voltage 42V 500 Output current 84mA 500 Consumed current 40mA Pulse width 100 S Frequency 1 110Hz Smallest are of electrode pad 4cm2 Timer 10 60 minutes...

Page 2: ...Troubleshooting 24 14 Disposal of the Unit 25 15 Warranty period 25 16 Electromagnetic Compatibility 25 17 Date 30 III Labels on the device 31 I Foreword Before operation please read this user s manu...

Page 3: ...electrical conductive medium The irritation can usually be reduced by using an alternate conductive medium like gel or alternate electrode placement 7 Pch Life TENS PMS must be kept out of the reach o...

Page 4: ...cription of the Device 6 1 General Description of the Device Pch Life TENS PMS is a portable and DC 3 7V battery powered multifunction device offering both Transcutaneous Electrical Nerve Stimulation...

Page 5: ...Rechargeable lithium battery 6 4 Mode Description We suggested that you initially experiment using each of the 6 modes The mode that gives you the most desirable sensations and comfort is the most ap...

Page 6: ...requency changes among 12 5Hz to 55 5Hz and there are 64 periods on 220ms off 100ms then the device stops for 1 second The device repeats this cycle all the time Mode 3 Mode 4 1 Timing 10 60 minutes 2...

Page 7: ...ith flammable gases 16 Never use this product in concurrence with high frequency surgical equipment it may result in burns at the site of the stimulator electrodes and possible damage to the control u...

Page 8: ...pplying the electrodes 5 Apply the whole surface of the electrodes firmly to the skin Do not use electrodes that do not stick properly to the skin or only partially stick to the skin 6 In case of skin...

Page 9: ...nd Mode 4 for 10 20 minutes twice or 3 times a day Figure 7 Mode 1 for 20 minutes and Mode 3 for 20 minutes twice or 3 times a day Figure 8 Mode 4 for 10 20 minutes and Mode 6 for 10 20 minutes twice...

Page 10: ...it becomes uncomfortable Always stay below that point of discomfort 3 Recommended application duration and Mode selection The modes to be used for improving the muscle performance are Modes 1 4 5 and...

Page 11: ...last approximately 1 hour When charging is finished the LCD will show a full battery cell 4 The battery should be charged for 10 hours or so before first use Notes Only charge the unit when battery i...

Page 12: ...store the electrodes in the reusable bag in a shady place Stockroom temperature 5 C 27 C 41 80 F and humidity of 30 80 No need to sterilize 3 Never apply the pads to any other surface other than your...

Page 13: ...of pad is not sticky Are you using pad when perspiring Use when not perspiring in a cool room Were the pads stored under high temperature high humidity or direct sunshine Replace the pad 25 24 14 Dis...

Page 14: ...meters do not deviate the essential performance requirement The device displays abnormally in LCD 2 As to patients The sensitivity of stimulation may be weaker or stronger but it does not produce safe...

Page 15: ...s equipment should be used no closer to any part of the Pch Life TENS PMS including cables than the recommended separation distance calculated from the equation applicable to the frequency of the tran...

Page 16: ...23 333 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the trans...

Reviews: