International version

DELTA

Portable Workstation

Quick Start Guide

 

103158-01EN [U8998465]— Revision A

April 2012

Summary of Contents for Delta

Page 1: ...International version DELTA Portable Workstation Quick Start Guide 103158 01EN U8998465 Revision A April 2012 ...

Page 2: ...990865 Any DELTA handheld analyzer model including the Premium Standard or Classic In this configuration the DELTA is controlled by the Olympus DELTA PC software The open beam handheld analyzer is then converted into a safe closed beam workstation 1 1 Test stand CONTENTS Product Overview Page 2 Safety Information Page 3 Physical Planning Page 5 Unfolding the Test Stand Page 6 Configuring the Works...

Page 3: ...y trained and authorized operators in accordance with proper safety procedures Improper use may impair safety protection and cause potential harm to the user Read all warning signs and labels 5 I O panel back of the test stand a Serial 9 pin D Subminiature connector b 15 pin D Subminiature connector for the probe adaptor c USB port and AC power Component Key DELTA Portable Workstation All models 1...

Page 4: ...est button is disabled grayed out It is not possible to force the X ray On condition If the lid is opened during an active test the X ray tube turns off immediately and a Test Aborted message is displayed The instrument turns off 2 3 X Ray Indicator The X ray indicator is located on the top of the test stand see Figure 2 1 on page 4 This indicator consists of a four element amber LED array and has...

Page 5: ...nt The DELTA Portable Workstation weighs 8 9 kg 19 6 lb with the probe adaptor analyzer and battery included To make a minimum footprint for the test stand add at least 5 1 cm 2 in to the actual 52 1 cm 20 5 in width and 37 5 cm 14 75 in depth Plan for a 70 cm 27 5 in height when the lid is fully open see Figure 3 1 on page 5 You should be able to comfortably access the analyzer s lid latch and te...

Page 6: ...ging accessories require an AC power source 4 Unfolding the Test Stand To unfold the test stand To avoid equipment damage make sure that there is sufficient clearance when unfolding the test stand see section 3 1 on page 5 1 Place the test stand on a table or a bench with its legs facing toward you see Figure 4 1 on page 7 2 Make sure that the locking levers are open see Figure 4 1 on page 7 3 Cen...

Page 7: ...Figure 4 2 on page 7 Figure 4 2 Legs locked into position 5 Place the test stand upright 6 If necessary use leveling feet to stabilize the test stand 7 Before opening the lid slide the latch see Figure 4 3 on page 7 to the right Figure 4 3 Lid latch Locking lever in the release position Leg Leg Locking lever in the release position Locking lever in the locked position Locking lever in the release ...

Page 8: ...e DELTA measurement window is flush with the probe adaptor faceplate see Figure 5 2 on page 8 Figure 5 2 DELTA measurement window flush with the probe adaptor faceplate 4 Tighten the two thumbscrews making sure that the measurement window stays flush with the probe adaptor faceplate 5 Install the analyzer adaptor assembly on the test stand see To install the analyzer adaptor assembly onto the test...

Page 9: ...at the probe adaptor faceplate is flush with the test chamber floor 5 Carefully turn the analyzer adaptor assembly counterclockwise until it locks The probe adaptor plastic housing must be centered with the locking fixture alignment screw see Figure 5 5 on page 9 Figure 5 5 Analyzer adaptor assembly installed on the test stand 6 Place the workstation in the upright position Locking fixture Locking...

Page 10: ...or to the test stand s I O panel see Figure 5 6 on page 10 and Figure 5 7 on page 10 Figure 5 6 Probe adaptor interface cable Figure 5 7 Test stand I O panel 2 Plug the USB cable mini USB B connector into the DELTA data port see Figure 5 8 on page 10 Figure 5 8 Mini USB B connector plugged into the analyzer data port 3 Plug the USB cable USB A connector into a PC USB port 15 pin D Subminiature con...

Page 11: ...tly performed or adhered to could result in death or serious personal injury Do not proceed beyond a DANGER signal word until the indicated conditions are fully understood and met The WARNING signal word indicates a potentially hazardous situation It calls attention to a procedure practice or the like which if not correctly performed or adhered to could result in death or serious personal injury D...

Page 12: ...ipment Immediately notify the carrier making the delivery of any damage because the carrier is normally liable for damage during shipment Retain packing materials waybills and other shipping documentation needed in order to establish a damage claim After notifying the carrier contact Olympus for assistance with the damage claim and equipment replacement if necessary This instruction manual attempt...