background image

Silo

Patrol

®

SE

Silo Monitoring Unit (SMU)

B U L L E T I N

344A

I N S T A L L A T I O N   &   O P E R A T I O N

®

Thank  you  for  purchasing  a  quality  product  manufactured  by  Monitor

Technologies  LLC.  We  realize  that  you  do  have  a  choice  of  vendors

when procuring level measurement equipment and we sincerely appre-

ciate your business!
This manual contains the information necessary to ensure a safe and

successful installation. Please read and comply with the section on page

18 of this manual pertaining to SAFETY. Doing so will ensure proper

operation of the equipment and the safety of all personnel.
Before  discarding  the  shipping  container,  please  inspect  it  thoroughly

and verify that all parts ordered are accounted for.  Observe the specifi-

cation label on the SMU cover.  Verify the voltage, output type, cable

type, cable length, flange style, and location approval (ordinary or haz-

ardous).  All SMUs are shipped with a standard plumb bob to ease instal-

lation.  If an alternate plumb bob is ordered, it will be packed separately.

See Mechanical Installation section for method to change plumb bob.
In the event that information contained herein does not completely sat-

isfy  your  requirements  or  answer  your  questions,  you  may  contact

Technical Support on our website www.monitortech.com, by telephone

at 800-766-6486  (630-365-9403), or by e-mail at techsupport@moni-

tortech.com. If your SMU ever requires service either in or out of warran-

ty, please contact us and obtain an RMA number prior to shipping the

unit to us. 

www.monitortech.com

SMU

Important Notice

SMU has wireless

Bluetooth

®

set-up.

See Page 8 for more 

details and instructions 

on downloading

ConfigureSensor app.

   

      

Ordinary

Location

Unit Shown

Summary of Contents for Silo Patrol SE 344A

Page 1: ...label on the SMU cover Verify the voltage output type cable type cable length flange style and location approval ordinary or haz ardous All SMUs are shipped with a standard plumb bob to ease instal lation If an alternate plumb bob is ordered it will be packed separately See Mechanical Installation section for method to change plumb bob In the event that information contained herein does not comple...

Page 2: ...res the amount of cable dispensed via its unique optical sensing system The SMU s optic system is completely sealed from the internal environment of the electronics compartment which is isolated and sealed from material ingress in the mechanical compartment to ensure long term reliable operation The descent of the bob is maintained at an optimal speed by the smart motor control system which contri...

Page 3: ...ECT ELECTR ELECTRI ELECTRIC ELECTRICA ELECTRICAL ELECTRICAL ELECTRICAL I ELECTRICAL IN ELECTRICAL INS ELECTRICAL INST ELECTRICAL INSTA ELECTRICAL INSTAL ELECTRICAL INSTALL ELECTRICAL INSTALLA ELECTRICAL INSTALLAT ELECTRICAL INSTALLATI ELECTRICAL INSTALLATIO ELECTRICAL INSTALLATION FACT ELECTRICAL INSTALLATION F ELECTRICAL INSTALLATION FA ELECTRICAL INSTALLATION FAC ELECTRICAL INSTALLATION FACTO EL...

Page 4: ...G DIAGRAM SMU RS 485 MODELS ADV WIRING DIAGRAM SMU RS 485 MODELS ADV M WIRING DIAGRAM SMU RS 485 MODELS ADV M0 WIRING DIAGRAM SMU RS 485 MODELS ADV M00 WIRING DIAGRAM SMU RS 485 MODELS ADV M006 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 WIRING DIAGRAM SMU RS 485 MODELS ADV M0068 7 2 WIRING DIAGRAM SMU RS 48...

Page 5: ...OU LOCKOUT LOCKOUT N L N N L1 N L1 POWER N L1 P N L1 PO N L1 POW N L1 POWE N L1 POWER DEPL N L1 POWER N L1 POWER D N L1 POWER DE N L1 POWER DEP N L1 POWER DEPLU N L1 POWER DEPLUGGABL N L1 POWER DEPLUG N L1 POWER DEPLUGG N L1 POWER DEPLUGGA N L1 POWER DEPLUGGAB N L1 POWER DEPLUGGABLE N L1 POWER DEPLUGGABLE N L1 POWER DEPLUGGABLE FROM BOT N L1 POWER DEPLUGGABLE F N L1 POWER DEPLUGGABLE FR N L1 POWER...

Page 6: ... 485 network SMU Tx orange When illuminated indicates that the SMU is transmitting data out onto the RS 485 network Status yellow When flashing indicates the general well being of the on board Bluetooth radio When flashing rapidly indicates that the Bluetooth radio is currently being pro grammed and is not available for Bluetooth connections nor is the SMU able to take a sounding Off indicates the...

Page 7: ...and 21 5mA See Page 8 for the configuration process 3 Error Reaction The 4 20mA analog SMU can be config ured to produce an error signal on the 4 20mA signal Upon an operational error the SMU will produce a current value of 2mA or 23mA The receiving device must be configured with a respective threshold to recognize the signal is outside of the standard operational range 3 21 5mA and that some form...

Page 8: ... phone or tablet from Monitor s website at http www monitortech com apps or scan the following QR Code 1 l Install the app from an e mail Consult factory l If device does not have Wi Fi connection or mobile data service other options like transfering the app from a PC to the device via a USB connection might be available Consult factory App must be connected with the SMU through Bluetooth in order...

Page 9: ... dependent on the protocol selected default address 1 RELAY SETTINGS RELAY ACTION FORCE RELAY RELAY THRESHOLD OFF CYCLE OPERATE ERROR LOW THRESHOLD HIGH THRESHOLD NOT ACT on ACT off No relay action energized whenever power is applied Relay relaxed while SMU is sounding ener gized when bob idle default CYCLE OPERATE Relay relaxed when error is detected energized when error dismissed When distance t...

Page 10: ... READ ONLY Defines physical length of cable Sets max value for relay threshold cycle limit Sounding stops upon measure reaching value Also presets the time values in Descend Assist of Zone 1 Range 1Ft 150Ft typc 1Ft 200Ft special cases SMU will periodically report distance measured thus far while still sounding EXISTING MEASUREMENT This option is only viewable when Descend Assist is ON 0 008 168H ...

Page 11: ...Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory Consult Factory SHOW REG MAP For Modbus protocols only Displays the Modbus register map as can be see in Installation Bulletin 344N 11 Menu Selection for ConfigureSensor App RS 485 Continued Factory Defaults are Shown in Bold Denote...

Page 12: ...es relay energized Use distance from latest measure only update on good cycles relaxed if Genesis When distance thres relay energized When distance thres relay relaxed Use distance from latest measure only update on good cycles relaxed if Genesis Relay temporarily not activated on regardless of status Relay temporarily activated off regardless of status Distance threshold range 0 150ft 0 45 7m def...

Page 13: ...tion is only viewable when Descend Assist is ON CABLE LENGTH CSL READ ONLY Defines physical length of cable Sets max value for relay threshold cycle limit Sounding stops upon measure reaching value Also presets the time values in Descend Assist of Zone 1 Range 1Ft 150Ft typc 1Ft 200Ft special cases 0 008 168H ...

Page 14: ... errors were detected they will be indicated on the HMI2 The following error conditions are analyzed and dis played if applicable Com Error Communication between HMI2 and SMU is faulty Return Error The measurement of descent was greater than the ascent If 4 100mm 1 upward travel Index Error The measurement of ascent was greater than the descent If 4 100mm Motion Error The plumb bob did not move up...

Page 15: ...on cable around rollers measuring wheel and through the wiper seal compartment and mounting flange Do not install wiper seals yet 7 Reattach plumb bob Pass cable through two compression sleeves around thimble and back through compression sleeves Crimp compression sleeves 8 Install wiper seals Install wiper seals as explained in the Wiper Seal Replacement section 9 Re establish operation Remove all...

Page 16: ... The plumb bob did not travel as far when ascending as it did when descending Determine if plumb bob is trapped in material Request another measurement A subsequent motion error confirms an entrapment A subsequent index error indicates that plumb bob has freed itself Request another measurement and errors will likely cease 2 Index Error The plumb bob did not travel as far when descending as it did...

Page 17: ... IN BRACKETS Silo Monitoring Unit SMU Digestible Plumb Bob 6 2190 Standard K Flange 6 3105 5 K Flange 6 3125 10 K Flange 6 3128 Standard Aluminum Plumb Bob 6 3103 Standard Aluminum Teflon Coated Plumb Bob 6 3123 Standard Stainless Steel Plumb Bob 6 3136 Inverted Cone Plumb Bob 6 4106 Inverted Cone Teflon Coated Plumb Bob 6 4126 Ball Float Plumb Bob 6 4108 Ball Float Teflon Coated Plumb Bob 6 4128 ...

Page 18: ...manual to highlight important safety issues Please pay particular attention to these items Electrical Shock Caution SMUs are powered with HIGH VOLTAGE No operator serviceable parts are inside All servicing is to be performed by qualified personnel Each SMU is pro vided with a protective conductor terminal which shall be terminated to earth ground potential See Electrical Installation This product ...