background image

 

MES

-

CK08-075-11EN-0 

PREPARED: (2022-10-18) (Version 3) 

Instruction for Use 

 

Trade Name: Base Unit Attachment 

1/2 

Be sure to read this instruction manual before using this product. 
Keep this manual available for reference when needed. 

Warning 

When  using  operating  table  accessories,  always  pay  close 
attention to the patient. Patients left in the same posture for an 
extended period of time are at risk for developing nerve paralysis 
and/or bedsores.

 

 

Contraindication/Prohibition 

1.  This  product  is  only  validated  for  use  with  Mizuho  operating 

tables. 

2. Use only appliances suitable for the intended use of this   

product. Incorrect use could cause breakage of this product as   
well as that of tools and appliances used in combination with   
it. 

Symbol mark for labeling 

  MD    : Medical Device 

Specifications 
1.

 

Shape 

 
 
 
 
 
 

 
 
 
 
 
 
 
 

Frame 

Insertion hole of the Attachment 

②Insertion shaft 

⑤Fixing handle of the Attachment 

③Fixing handle 

 

Code No. 

Product Description 

08-075-11 

Attachment for Mayfield Head Support 

2.

 

Weight 
2.1 kg 

3.

 

Material 

Frame: Stainless steel 

Insertion shaft: Stainless steel 

Fixing handle: Synthetic resin 

Insertion hole of the Attachment: Stainless steel 

Fixing handle of the Attachment: Synthetic resin 

Note: The numbers correspond to those used in 1. Shape. 

 
 

Intended purpose 

This is an accessory for an operating table. This product is used to   
attach the base unit such as the Head Supporter. 

Intended user 

This product is to be used by health care professionals, including 
but not limited to surgeons, nurses and biomedical technicians. 

Instructions for use 
 

 

 
1. Remove the head plate from the operating table. 
2. Attach the Base Unit Attachment to the back plate insertion 

holes of the operating table. 

3. Tighten the fixing handles of the back plate. 
4. Attach the base unit to the insertion holes of the Attachment. 
5. Tighten the fixing handles of the Attachment. 
6. Attach the component parts to the base unit according to 

the instructions for use of the base unit. 

Note: Refer to the instruction manual of the base unit for more   

usage information of the base unit. 

Warning/Caution 

1.

 

Warning 
Ensure all fixing handles are tight. Loose condition can cause 
the products to come off or move, which may result in a patient 
injury. 

2.

 

Important caution 

2-1.When working with an operating table, take care not to allow 

this product to make contact or interfere with the table top 
or with other tools and appliances used in combination with 
this product. 

2-2.For hygiene, be sure to use sterilized drapes on the areas on 

this product where the patient comes into contact with it. 

Storage/Life 

1.

 

Do not store the device in high temperatures or in areas with 
high humidity where the temperature has drastic variations. 

2.

 

Service life of this product: 7 years 
(Subject  to  the  specified  maintenance  and  inspection  and  is 
stored properly.) 

 

 

Unit

mm 

Fixing handle 

Fixing handle

 

Fixing handle of 
the Attachment

 

Base Unit 

Insertion hole of 

the Attachment 

Insertion hole 

 

Insertion hole 

 

Operating table back plate 

Base Unit Attachment

 

Reviews: