Module E 505 for intubation equipment

Miele Company Ltd., Fairacres, Marcham Road, Abingdon, Oxon, OX14 1TW, Tel 0330 160 6600,
Internet: http://www.miele.co.uk /

!

0297

Alteration rights reserved / Publication date 2015-11-01

,

Please read the Warning and Safety instructions in the Operating

Instructions for your washer-disinfector carefully. You should only process
instruments/medical products in this module if they have been declared as
suitable for machine processing by the manufacturer. Please also observe the
manufacturer's care instructions. The programme selected must be suitable for
the materials from which the items to be processed are made. Where possible
the programme should include thermal disinfection.

Areas of application:

This module is designed for use in mobile unit E 501 (which is supplied with
separate operating instructions). It is designed for cleaning, disinfecting and
drying hollow anaesthetic instruments and equipment.
Special holders and injector jets are used for securing breathing tubes, breathing
bags, Guedel tubes and endotracheal tubes, laryngeal masks and bellows.

,

To ensure a suitable standardised spray pressure for all connections, all

screw connectors must be fitted with jets, adapters, sleeves or blanking screws.
Damaged jets, adapters or sleeves etc. must not be used.

Empty jets, etc. do not have to be replaced with blanking screws.

,

To ensure that hollow instruments are thoroughly cleaned and disinfected

internally, it is essential to connect them to the correct holders/jets in the mobile
injector unit. They must not detach themselves from their holders during the
programme.
If, however, a hollow instrument does become dislodged from its holder during
a programme, it must be processed again.

,

To avoid damage to certain items, observe the maximum permitted drying

temperature.
The maximum temperature permitted for the elastomers in some breathing bags
and masks is below 85°C.
Drying results must be inspected after the cleaning programme has finished.
If necessary, adjust the drying time at the end of the cleaning programme.

^

Expel air from the balloon lead.

^

Endotracheal tubes are fitted onto the
injector jets so that spring clips
secure them to the jets.

^

Place the balloon lead with the valve
behind the spring clip.

^

Fit the breathing bags onto the
injector jets.

^

Make sure the opening of the
breathing bag fits securely over the
plastic cone.
This will prevent the breathing bag
from sticking together during the
cleaning and drying process and
ensure it is thoroughly cleaned
internally.

The breathing bag is attached
correctly when it cannot dislodge
itself from the plastic cone.

^

Fit module E 505 into mobile unit
E 501.

To do this:

^

Place the two positioning guides over
the middle water inlet.

The inlet pipe needs to be positioned in
front of the cover in the water inlet.

^

Press the bracket on the left hand
side down until the handle engages
behind the rail on the mobile unit.

^

Attach Guedel tubes to injector jets
with spring clips.

The illustration above shows module E 505 in mobile unit E 501.

^

Remove the stopper from breathing
masks which have a closable air
cushion. Expel as much air as
possible and then replace the
stopper.

^

Several breathing masks can be
placed onto one breathing bag
injector jet.