JPD-100S4

Fetal Doppler

INSTRUCTION MANUAL

Manual Version: V1.3

Issuing Date: 2018/05/22

Summary of Contents for AngelSounds JPD-100S4

Page 1: ...JPD 100S4 Fetal Doppler INSTRUCTION MANUAL Manual Version V1 3 Issuing Date 2018 05 22 ...

Page 2: ...or organization shall not copy modify or translate this book All contents described in this book are consistent with the actual situation of the related product Shenzhen Jumper Medical Equipment Co Ltd has right to revise all contents of this book if needed without prior notice Shenzhen Jumper Medical Equipment Co Ltd reserves the right of final interpretation of this book JUMPER and ANGELSOUNDS a...

Page 3: ...NS 3 2 5 SAFETY ALERT DESCRIPTIONS 4 2 6 SYMBOL DESCRIPTIONS 5 SECTION 3 USING THE PRODUCT 6 3 1 UNPACKING AND INSPECTING 6 3 2 CONTROLS AND DISPLAY 7 3 2 1 Structure and shape 7 3 2 2 Keys 8 3 2 3 Display panel 10 3 3 BATTERY INSTALLATION AND REMOVAL 12 3 4 AUDIO OUT 12 3 5 PREPARATION 13 3 6 USING PRODUCT TO DETECT 13 SECTION 4 MAINTENANCE AFTER SALES SERVICE 14 4 1 MAINTENANCE 14 4 2 RECOMMENDE...

Page 4: ...ON 5 SPECIFICATIONS AND SAFETY 18 5 1 SPECIFICATIONS 18 DISPLAY 19 ULTRASOUND 19 FHR PERFORMANCE 19 BATTERY 19 5 2 MODE OF OPERATION 20 5 3 PHYSICAL DIMENSIONS 20 5 4 ENVIRONMENTAL REQUIREMENTS 20 OPERATING CONDITIONS 20 STORAGE AND SHIPPING CONDITIONS 20 SECTION 6 ACCESSORIES 20 OVERVIEW 21 PRODUCT ACCESSORIES 21 APPENDIX A EMC INFORMATION GUIDANCE AND MANUFACTURE S DECLARATION 21 ...

Page 5: ...t portable detector It is designed to meet your detecting and hearing needs by providing advanced detecting functions and a full range of sound of the fetal heartbeat The product is mainly used to detect the fetal heartbeat rate FHR and the sound of the fetal heartbeat SFH The growth and development of a fetus can be found out through examination of these indices It is applicable for department of...

Page 6: ... formed It is transmitted an output signal of cardiac valves vibrating and it is sent to the loudspeaker for getting a rhythmical sound with the fetal heartbeat Simultaneously it is sent to a counter which calculate periods of heartbeat that is fetal heartbeat rate bpm beat per minute SECTION 2 SAFETY GUIDANCE 2 1 INDICATIONS FOR USE The product is normally applied to fetus above twelve weeks grow...

Page 7: ...rder complied with the safety requirements of IEC 60601 1 2 2 CONTRAINDICATIONS FOR USE Normally none as a particular case please consult your doctor 2 3 NOTE FOR HOME USE This device cannot replace a professional fetal monitor If the fetal heart rate is abnormal or cannot be located using this monitor immediately go to the hospital to seek the doctor s help If fetal movement is not felt by the pr...

Page 8: ...Medical Ultrasonic Instrument Co Ltd which may cause the Product to function improperly during a rescue CAUTION Possible Radio Frequency RF Susceptibility RF susceptibility from cellular phones CB radios and FM 2 way radio may cause interference with the product Do not operate wireless radiotelephones in the vicinity of the product turn power OFF to the radiotelephone and other like equipment near...

Page 9: ...r glutaraldehyde based cleaning solution to avoid damage to the metal connectors CAUTION Environment of use The product is designed for indoor use Operator must confirm that the environment of use meets the required operating environmental specifications before using CAUTION Cold Environments If the product is stored in an environment with a temperature below the operating temperature the unit sho...

Page 10: ...n for separate collection after its useful life Type BF applied part Ingress Protection SECTION 3 USING THE PRODUCT This section presents information on unpacking and setting up the product 3 1 UNPACKING AND INSPECTING Every attempt is made to ensure your accurate and complete order However to be sure that your correct order verifying the contents of the box against your packing slip The product i...

Page 11: ...ing shipment Check the components according to the packing list Check for any damage or defects Do not attempt to assemble the product if anything is damaged or defective Contact Shenzhen Jumper Medical Equipment Co Limited Customer Service immediately if anything is damaged or defective 3 2 CONTROLS AND DISPLAY 3 2 1 Structure and shape ...

Page 12: ...8 3 2 2 Keys There are one knob and four push buttons on JPD 100S4 Volume on off knob ...

Page 13: ...from 5 to 1 gradually and circle If in Warn sound on or off mode press the button can open or close the warn sound If in FHR restrict configure mode press the button can decrease the restrict value Mode 3 No function in Playback mode the functions in the other modes are same as mode 2 Mode 4 Same as mode 1 Right Key Functions Different in every mode Mode 1 Close or open the backlight in the Auto c...

Page 14: ...ct configure mode Press it again the working mode back to Play music mode Mode 3 Playback mode Same as mode 2 Mode 4 Audio record mode Press the button once Benign recording the fetal heartbeat sound The most recording time is 60 seconds If recording reach 60 seconds the working mode is changed to mode 1 automatically If recording not reach 60 seconds press the function key the working mode change...

Page 15: ...laying back the fetal heartbeat sound Audio record Twinkling when recording the fetal heartbeat sound Play music Appeared when playing music Average FHR mode Appeared in Average FHR mode Transient FHR mode Appeared in Transient FHR mode Battery indicator Display the quantity of electricity in batteries FHR display In mode 1 ever in Auto calculating mode or in Manual calculating mode it displays th...

Page 16: ...not likely to be used for some time WARNING Irregular treatment of batteries may be result in hazards to health and environment 3 4 AUDIO OUT The headset jack socket can be connected to headset or recorder A socket for audio output can only be connected with a recorder complied with the requirements of IEC 60601 1 ...

Page 17: ... heartbeat Adjust the probe to obtain an optimum audio signal ideally by angling the probe around Generally the site of heart of fetus is 1 3 below of navel line at its earlier stage it then moves upward with increasing of gestational period and the site of heart of fetus will be a little deviation to left or right with different fetus Pls make sure that the surface of probe should be contacted fu...

Page 18: ...manner assume the responsibility for performing the recommended maintenance schedule The sole responsibility rests with the individuals hospitals or institutions utilizing the product 4 1 1 The transducer acoustic surface is frangible and must be handle with care Gel must be wiped off from the transducer after use These precautions will prolong the life of the unit The user must check that the equ...

Page 19: ...fully inspected prior to installation once every 12 months thereafter and each time the equipment is serviced Carefully inspect the equipment for physical damage Inspect all external connections for loose connectors or frayed cables Inspect the graphics display for marks scratches or other damage Verify that the Safety label on back of the Product is clearly legible INSTRUCTION INSPECT FOR RECOMME...

Page 20: ...lution in water Mild soap and water Sodium hypochlorite chlorine bleach 3 solution in water Quaternary ammonium compounds such as Lysol 10 solution in water Do not use abrasive cleaners or strong solvents such as acetone or acetone based cleaners Do not use mixing disinfecting solutions such as bleach and ammonia as hazardous gases may result Do not clean electrical contacts or connectors with ble...

Page 21: ...the product or accessories CAUTION Cleaning liquids DO not submerge the device in liquids or pour cleaning liquids over into or onto the device Don t use strong solvent for example acetone Never use an abrasive such as steel wool or metal polish Do not allow any liquid to enter the product and do not immerse any parts of the device into and liquids Avoid pouring liquids on the device while cleanin...

Page 22: ... in this chapter If you are unable to resolve the problem contact Shenzhen Jumper Medical Equipment Co Limited Service NOTE The warranty will be void upon unauthorized disassembly or service of the product 4 9 CONTACT INFORMATION Shenzhen Jumper Medical Equipment Co Limited D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26696279 Fax 86 755 26852025 Website ht...

Page 23: ...0 50Hz Targe velocity 10cm s 40cm s 90dB Spatial peak temporal peak acoustic pressure 1MPa Output power 20mW Effective area of the ultrasonic transducer active element 4 92cm2 The acoustic coupling medium for normal use ph 5 5 8 Acoustic impedance 1 5 106 1 7 106Pa s m FHR PERFORMANCE FHR Test Range 50 210 bpm Beat Minute Resolution 1 bpm Accuracy 2 bpm AUDIO OUTPUT Audio Output Power 0 5 W Audio ...

Page 24: ...RAGE AND SHIPPING CONDITIONS Temperature 20 C to 55 C Humidity 10 93 RH non condensing Atmospheric pressure 50kPa to 106kPa CAUTION Environment of use Product is designed for indoor use Operator must confirm that the environment of use meets the required operating environmental specifications before using CAUTION Cold Environments If the Product is stored in an environment with a temperature below...

Page 25: ...eds special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING OCUMENTS Portable and mobile RF communications equipment can affect Fetal Doppler The Fetal Doppler should not be used adjacent to or stacked with other equipment The Fetal Doppler is intended for use in the electromagnetic environment specified bel...

Page 26: ...kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 The Fetal Doppler is intended for use in the electromagnetic environment specified below The customer or the user of the Fetal Doppler should assure that it is used in such an environment Immun ity test IEC 606 01 test le...

Page 27: ...ection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If t...

Page 28: ...transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the tra...

Page 29: ......

Page 30: ...Shenzhen Jumper Medical Equipment Co Ltd Address D Building No 71 Xintian Road Fuyong Street Baoan Shenzhen Guangdong China Tel 86 755 26696279 Fax 86 755 26852025 Website http www jumper medical com ...