Model LBP30

Smart

Pulsewave Blood Pressure Monitor

Model LBP30

Smart

Pulsewave Blood Pressure Monitor

INSTRUCTION MANUAL

Shenzhen RSP Medical Technology Co., Ltd

Summary of Contents for LBP30

Page 1: ...Model LBP30 Smart Pulsewave Blood Pressure Monitor Model LBP30 Smart Pulsewave Blood Pressure Monitor INSTRUCTION MANUAL Shenzhen RSP Medical Technology Co Ltd ...

Page 2: ... use the device correctly and efficiently please read this Instruction Manual before use In addition please take good care of the manual after your reading to ensure you can find it easily when necessary Visual number V1 1 0 This product is suitable for people who are older than 12 years of age in hospital clinic and social medical organizations etc ...

Page 3: ...NS 4 COMPONENTS OF THE PRODUCT 6 RECHARGE THE BATTERY 7 APPLYING THE CUFF 8 POSTURES INSTRUCTIONS 8 SETTING UNITS 9 TAKING A MEASUREMENT 10 CARE AND MAINTENANCE 12 ERROR INDICATORS 12 TROUBLESHOOTING TIPS 12 PRODUCT SPECIFICATIONS 13 APPENDIX A PACKING LIST 14 APPENDIX B EMC 14 ...

Page 4: ...values are for reference 3 Pregnant women pre eclamptic mental disorder or arrhythmia patients should use this device under practitioner s guidance 4 When common arrhythmias such as atrial premature ventricular premature ventricular premature and atrial fibrillation appear it cannot ensure that the blood pressure monitor will achieve the performance it claims to It is advised to use the product un...

Page 5: ...ff when arm is wet or sweaty please use it after wiping up 8 Do not subject the monitor or the AC adapter to strong shocks such as dropping on the floor otherwise it may damage the unit 9 Do not soak the unit or cuff in water 10 Operate the device only as intended Do not use the device for any other purpose 11 Do not store or use this device outside the specified temperature or humidity range or i...

Page 6: ...Pulsewave Blood Pressure Monitor Otherwise the device may reach less than the claimed accuracy COMPONENTS OF THE PRODUCT Main unit Display Systolic blood pressure Measurement data Pulse rate Diastolic blood pressure Unit ...

Page 7: ...te the measurement please recharge in time after the measurement 3 In charging process the charging indicator light is orange when the battery has been fully charged the charging light indicator turns green CAUTION 1 Please use the standard power adapter 5V 1A or other power supply equipment 5V 1A to recharge products 2 In the process of charging please do not use the product please do not plug in...

Page 8: ...gh the cuff loop with the monitor installed on the cuff be at the middle of the arm After placing the cuff correctly straining the tail of the cuff to fasten the cuff NOTES 1 Do not put your thick clothes into cuffs and the bottom of the cuff should be approximately 1cm above the elbow 2 It will get an incorrect result when the cuff is applied too tightly or too loosely it is advisable to put in a...

Page 9: ...the normally before measurement NOTE Your arm cuff and heart should be on the same level otherwise measurement accuracy will be affected SETTING UNITS In shutdown status press and hold the button for over 5 seconds to enter unit setting mode When mmHg displays on the screen it means the unit is mmHg if kPa displays on the screen it means the unit is kPa ...

Page 10: ... for 2 3 minutes Users are advised to be measured at the same time daily 1 Apply the DualCuff correctly refer to the section APPLYING THE DUAL CUFF and set right measuring postures refer to the section POSTURES INSTRUCTIONS Do not move your hands in the process of measurement 2 Press the button to start the monitor and then it shows 0 thus enter standby mode 3 Press the button to take a measuremen...

Page 11: ...be shown on the display circularly 8 Press the button to turn off the monitor If you want to quit the measurement in the process of measurement please press the button NOTES 1 The monitor will automatically turn off after 60 seconds without operation 2 Do not repeat the measurements in a short time or it may cause the arm blood can t timely return which may result in incorrect results A new measur...

Page 12: ... or direct sunlight 6 Do not forcefully bend the DualCuff or airway tubes Do not fold tightly 7 Do not start measurement when the DualCuff is not applied or else the DualCuff may be damaged 8 Changes or modifications not approved by RSP Medical will void the user warranty Do not disassemble or attempt to repair the unit or components 9 Remove the batteries if the unit will not be used for three mo...

Page 13: ...measurement to relax yourself and keep quiet PRODUCT SPECIFICATIONS Name Smart Pulsewave Blood Pressure Monitor Model LBP30 Measurement Range Pressure 0 270 mmHg 0 36 kPa Pulse Rate 40 min 180 min Accuracy Pressure 3mmHg 0 4kPa Pulse Rate 5 Power Supply DC3 7v Li batteries Operating Temperature Relative Humidity 5 40 15 RH 80 RH Storage and Transportation Temperature Relative Humidity 20 55 93 RH ...

Page 14: ...warranty card 1 3 USB cable 1 APPENDIX B EMC Please install and use this instrument according to the EMC information provided in this Instruction Manual The portable and mobile RF communications equipment can affect this instrument s normal operation Please use the accessories sold by our company the inappropriate one may result in increased emission or decreased immunity of this instrument The in...

Page 15: ...nitor is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes 6 Harmonic emissions EN 61000 3 2 Class A 7 Voltage fluctuations flicker emissions EN 61000 3 3 Complies Table 2 Guidance and manufacturer s declaration electromagnetic immunity The Smart Pul...

Page 16: ...or 5 sec Mains power quality should be that of a typical commercial or hospital environment If the user of the Smart Pulsewave Blood Pressure Monitor requires continued operation during power mains interruptions it is recommended that the Smart Pulsewave Blood Pressure Monitor be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field EN 61000 4 8 3 A m 3 ...

Page 17: ... range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless te...

Page 18: ... below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output of transmitter W 150 kHz to 80 MHz P V d 5 3 1 80 MHz to 800 MHz P E d 5 3 1 800 MHz to 2 5 GHz P E d 7 1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not li...

Page 19: ...mart Pulsewave Blood Pressure Monitor Model LBP30 Shenzhen RSP Medical Technology Co Ltd 2203 Taibang Technical Building Gaoxin S 4th Rd Nanshan District Shenzhen 518000 China Website www lepuhomecare com Tel 86 755 86952278 Email sales lepuhomecare com Made in China ...