Page 1: ...07 97 0327 7324 Rev 01 1997 Hans Pausch Rontgengeratebau Graf Zeppelin Str 1 D 91056 Erlangen ALL RIGHTS RESERVED Ru...

Page 2: ...nception 7 General Brief Description 8 Range of Application 8 Installation 9 Floor Space Required 9 Room Height 9 AC Power Mains 9 Operating Elements 10 Arrangement 10 Meaning of Symbols Function 11 M...

Page 3: ...vals must by all means meet the requirements of the respective legislation or government regulations Changes and additions to the product must comply with the relevant legislation as well as with the...

Page 4: ...aws and regulations have priority Mechanical Safety It is the responsibility of the operator to ensure the safety of the patient while the unit is in operation by visual check proper patient positioni...

Page 5: ...radiation protection measures must be observed In addition we recommend the following 1 Set the tube current as low as possible 2 Limit the radiation field to the maximum possible extent 3 Keep as far...

Page 6: ...42 EEC part 11 para 5 according the procedure in annex VII The CE mark is only applicable for the product without X ray components and Bucky The declaration of EC conformity can be sent to you by requ...

Page 7: ...rtical carriage E column base movable F floor rail G x ray tube H controls K collimator L GE tube support arm M GE collimator N GE model without control arm 07 97 7 0327 7324 Rev 01 1997 Hans Pausch R...

Page 8: ...m max down to 41 cm middle support arm The solid column stand is mounted on the column base by means of screws and contains the counterweight The travel range of column in the floorrails is 243 cm The...

Page 9: ...235 cm AC Power Mains Connection 24V DC 1 5 A are required for the electromagnetic brakes and the indicator light beam AC Power Mains The AC power mains connection requires a 30mA circuit breaker to...

Page 10: ...3 push button for light beam indicator 3a push button for field light 4 push button for rotational movement of tube collimator assembly 4a push button for rotational movement of tube collimator assem...

Page 11: ...this position Push on handle 8 after having pressed any of the functional push buttons 4a la 2a to allow movement Goniometer 5 indicates tilting angle of X ray tube to exposure object Optical pointer...

Page 12: ...ons for traces of wear Periodic Maintenance For trouble free operation of the RS 85 as well as safety for patient and user it is necessary to carry out a technical maintenance by the service company e...

Page 13: ...e Polish all enamelled metal surfaces Disinfection The equipment has to be switched off before disinfection Only disinfection methods can be used that correspond to the relevant regulations and rules...