INSTRUCTION MANUAL

HI96752

Calcium and Magnesium 

High Range ISM

Warning:

 Do not validate or calibrate the instrument with standard solu-

tions other than the Hanna Instruments 

CAL Check™

 Standards, otherwise 

erroneous results will be obtained.

For accurate validation and calibration results, please perform tests at room 

temperature (18 to 25 °C; 64.5 to 77.0 °F).

3

Validation

Note: The validation is performed only for 

the selected parameter. For full validation 

of the instrument, the following procedure 

must be performed for each parameter. 

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  Place the 

CAL Check™

 Standard Cuvette 

A into the cuvette holder and ensure that 

the notch on the cap is positioned securely 

into the groove.

4•

 Press 

ZERO/CFM

 and the lamp, cuvette and 

detector icons will appear on the display, 

depending on the measurement phase.

5•

  After a few seconds the display will show 

-0.0-

”. The meter is now zeroed and ready 

for validation.

6•

  Remove the cuvette.

7•

  Place the specific 

CAL Check

TM

 Standard 

Cuvette B into the cuvette holder, for:

 

Calcium

: B, 

HI96752-11

 

Magnesium:

 B, 

HI96754-11

  Ensure that the notch on the cap is 

positioned securely into the groove.

8•

 Press 

CAL CHECK

 key and the lamp, 

cuvette and detector icons together with 

CAL CHECK

” will appear on the display, 

depending on the measurement phase.

9•

  At the end of the measurement the display will 

show the validation standard value.  

The reading should be within specifications 

as  reported on the 

CAL Check™

 Standard 

Certificate. If the value is found out of 

specifications, please check that the cuvettes 

are free of fingerprints, oil or dirt and repeat 

validation. If results are still found out of 

specifications then recalibrate the instru-

ment.

4-5

7

8

Calibration

Note: It is possible to interrupt the calibra-

tion procedure at any time by pressing  

CAL CHECK or ON/OFF keys. When calibrat-

ing, only the selected range is affected.

1•

  Turn the meter on by pressing 

ON/OFF

.

2•

  When the beeper sounds briefly and the 

LCD displays dashes, the meter is ready.

3•

  To change the range, simply press 

RANGE/GLP

.

4•

  Press and hold 

CAL CHECK 

for three seconds 

to enter calibration mode. The display will 

show 

“CAL”

 during calibration procedure. 

The blinking 

“ZERO”

 asks for instrument 

zeroing.

5•

  Place the 

CAL Check™

 Standard Cuvette 

A into the cuvette holder and ensure that 

the notch on the cap is positioned securely 

into the groove.

6•

 Press 

ZERO/CFM

 and the lamp, cuvette and 

detector icons will appear on the display, 

depending on the measurement phase.

7•

  After a few seconds the display will show 

“-0.0-”

. The meter is now zeroed and ready 

for calibration. The blinking 

“READ”

 asks 

for reading calibration standard.

8•

  Remove the cuvette.

9•

  Place the specific 

CAL Check

TM

 Standard 

Cuvette B into the cuvette holder, for:

 

Calcium

: B, 

HI96752-11

 

Magnesium

: B, 

HI96754-11

  Ensure that the notch on the cap is 

positioned securely into the groove.

10•

 Press 

READ /TIMER

 and the lamp, 

cuvette and detector icons will appear on 

the display, depending on the measurement 

phase.

11•

 The instrument will show for three seconds 

the 

CAL Check™

 standard value.

Note: If the display shows “STD HIGH”, 

the standard value was too high. If the 

display shows “STD LOW”, the standard  

value was too low. Verify that both CAL 

Check™ Standard Cuvettes, A and B  are 

free of fingerprints or dirt and that they are 

inserted correctly.

12•

  Then the date of last calibration (e.g.: 

“01.08.2009”) appears on the display, 
or “01.01.2009” if the factory calibration 
was selected before. In both cases the year 
number is blinking, ready for date input.

13•

 Press 

RANGE/GLP

 to edit the desired 

year (2009-2099). If the key is kept 
pressed, the year number is automatically 
increased. 

9

4

5

6-7

9

10-11

Last Calibration Date

1•

  Press and hold 

RANGE/GLP

 

for three 

seconds to enter GLP mode. The calibration 
month and day will appear on the main 
display and the year on the secondary 
display.

2•

  If no calibration was performed, the factory 

calibration message, 

“F.CAL”

 will appear on 

the main display and the instrument returns 
to measurement mode after three seconds.

14•

 When the correct year has been set, press 

ZERO/CFM

 or 

READ /TIMER

 to confirm. 

Now the display will show the month blink-
ing.

15•

 Press 

RANGE/GLP

 

to edit the desired 

month (01-12). If the key is kept pressed, 
the month number is automatically in-
creased.

16•

 When the correct month has been set, press 

ZERO/CFM

 or 

READ /TIMER

 to confirm. 

Now the display will show the day blinking.

17•

 Press 

RANGE/GLP

 to edit the desired 

day (01-31). If the key is kept pressed, 
the day number is automatically increased.

Note: It is possible to change the editing 

from day to year and to month by pressing 

READ

/TIMER.

18•

 Press 

ZERO/CFM

 

to save the calibration 

date.

19•

 The instrument displays “

Stor

” for one 

second and the calibration is saved.

20•

 The instrument will return automatically to 

measurement mode by displaying dashes 
on the LCD.

12-14

 or

15-16

 or

17

18

In GLP mode, the last calibration date can be verified and the factory 

calibration can be restored.

1

3•

 Press 

ZERO/CFM

 to restore the factory 

calibration or press 

RANGE/GLP

  

again 

to abort factory calibration restore. 

4•

  The instrument briefly indicates 

“donE” 

upon restoration of factory calibration prior 
to returning to measurement mode.

To save the battery, the instrument shuts down after 10 minutes of non-
use in measurement mode and after 1 hour of non-use in calibration mode.

 

If a valid measurement was displayed before 
auto-shut off, the value is displayed when the 
instrument is switched on. The blinking 

“ZERO” 

means that a new zero has to be performed.
One fresh battery lasts for around 750 measurements, depending on the 
light level.
The remaining battery capacity is evaluated at the instrument startup and 
after each measurement.
The instrument displays a battery indicator with three levels as follows:
•  3 lines for 100 % capacity
•  2 lines for 66 % capacity
•  1 line for 33 % capacity
•  Battery icon blinking if the capacity is under 10 %.
If the battery is empty and accurate measurements can’t be taken any 
more, the instrument shows 

“dEAd bAtt”

 and turns off.

To restart the instrument, the battery must be replaced with a fresh one.
To replace the instrument’s battery, follow the steps:
•  Turn the instrument off by pressing 

ON/OFF

.

•  Turn the instrument upside down and remove the battery cover by 

turning it counterclockwise.

•  Extract the battery from its location and replace it with a fresh one.
•  Insert back the battery cover and turn it clockwise to close.

2

3-4

19

2

VALIDATION AND CALIBRATION PROCEDURES

 

Use the Hanna Instruments CAL Check™ cuvettes (see 
“Accessories”) to validate or calibrate instruments.

Validation

 

Calibration

 

GLP

Factory Calibration 
Restore

BATTERY MANAGEMENT

Last Calibration 
Date

 

3''

Before using these products, make sure that they are entirely suitable for your 
specific application and for the environment in which they are used.
Operation of these instruments may cause unacceptable interferences to other 
electronic equipments, this requiring the operator to take all necessary steps 
to correct interferences. Any variation introduced by the user to the supplied 
equipment may degrade the instrument’s EMC performance.
To avoid damages or burns, do not put the instrument in microwave oven. 
For yours and the instrument safety do not use or store the instrument in  
hazardous environments. 

RECOMMENDATIONS FOR USERS

Factory Calibration Restore

It is possible to delete the calibration and 
restore factory calibration.

1•

  Press and hold 

RANGE/GLP

 

for three 

seconds to enter GLP mode.

2•

 Press 

READ /TIMER

 to enter in the 

factory calibration restore screen. The 
instrument asks for confirmation of user 
calibration delete.  

Summary of Contents for HI96752

Page 1: ...INSTRUCTION MANUAL HI96752 Calcium and Magnesium High Range ISM ...

Page 2: ...port contact your local Hanna Instruments Office or e mail us at tech hannainst com Find your local Hanna Instruments Office at www hannainst com PRELIMINARY EXAMINATION For more details about spare parts and accessories see Accessories SPECIFICATIONS Range Calcium 0 to 400 mg L Magnesium 0 to 150 mg L Resolution 1 mg L Accuracy 25 C 77 F Calcium 10 mg L 5 of reading Magnesium 3 mg L 3 of reading ...

Page 3: ... measuring scheme lamp cuvette detector appears during different phases of zero or reading measurement 2 Error messages and warnings 3 The battery icon indicates the charge state of the battery 4 The hourglass appears when an internal check is in progress 5 Status messages 6 The chronometer appears when the reaction timer is running 7 The month day and date icons appear when a date is displayed 8 ...

Page 4: ...Oxalate reagent Magnesium Using another 1 mL syringe add to the cuvette exactly 0 5 mL of sample Note Do not mix up the two 1 mL syringes 9 Calcium Replace the cap and invert the cuvette 10 times to mix Magnesium Replace the cap and invert several times to mix 10 Replace the cuvette into the holder and en sure that the notch on the cap is positioned securely into the groove 11 Press READ TIMER for...

Page 5: ... measurement Please check the preparation of the zero cuvette Light High There is too much light to perform a measurement Please check the preparation of the zero cuvette Inverted cuvettes The sample and the zero cuvette are inverted Zero A zero reading was not taken Follow the instructions of the measurement procedure for zeroing the meter Under range A blinking 0 indicates that the sample absorb...

Page 6: ...n 14 days from the date of purchase WARRANTY IST96752 01 18 Hanna Instruments reserves the right to modify the design construction or appearance of its products without advance notice All rights are reserved Reproduction in whole or in part is prohibited without the written consent of the copyright owner Hanna Instruments Inc Woonsocket Rhode Island 02895 USA Reagent Sets HI93752 01 Reagents for 1...

Page 7: ...the measurement phase 5 After a few seconds the display will show 0 0 The meter is now zeroed and ready for validation 6 Remove the cuvette 7 Place the specific CAL CheckTM Standard Cuvette B into the cuvette holder for Calcium B HI96752 11 Magnesium B HI96754 11 Ensure that the notch on the cap is positioned securely into the groove 8 Press CAL CHECK key and the lamp cuvette and detector icons to...

Page 8: ...blinking READ asks for reading calibration standard 8 Remove the cuvette 9 Place the specific CAL CheckTM Standard Cuvette B into the cuvette holder for Calcium B HI96752 11 Magnesium B HI96754 11 Ensure that the notch on the cap is positioned securely into the groove 10 Press READ TIMER and the lamp cuvette and detector icons will appear on the display depending on the measurement phase 11 The in...

Page 9: ...linking 17 Press RANGE GLP to edit the desired day 01 31 If the key is kept pressed the day number is automatically increased Note It is possible to change the editing from day to year and to month by pressing READ TIMER 18 Press ZERO CFM to save the calibration date 19 The instrument displays Stor for one second and the calibration is saved 20 The instrument will return automatically to measureme...

Page 10: ... empty and accurate measurements can t be taken any more the instrument shows dEAd bAtt and turns off To restart the instrument the battery must be replaced with a fresh one To replace the instrument s battery follow the steps Turn the instrument off by pressing ON OFF Turn the instrument upside down and remove the battery cover by turning it counterclockwise Extract the battery from its location ...