background image

ZX3030B AC/DC Three 

Phase Standard Power Source

Summary of Contents for ZX3030B

Page 1: ...ZX3030B AC DC Three Phase Standard Power Source ...

Page 2: ...external voltage or current signal to the signal output of the device 3 It is prohibited to access signals beyond the measurement range to the signal measure end of the device 4 It is prohibited to connect any connection plugs to the ground end of the front panel 5 When disconnecting the cable always disconnect the power supply first 6 When the output voltage is more than 36V attention should be p...

Page 3: ...acteristics 1 III Specifications 1 IV Familiar with the device 3 Front Panel Features 3 rear panel Features 4 V Operation instructions 5 Attention 5 Main menu 6 Ac output operation 6 Electric energy error detection 8 DC output operation 9 VI Standard configuration 10 List 10 ...

Page 4: ...quency meter power factor meter phase meter can be checked Voltage current waveform power factor and other factors caused by the changes The power supply part can generate the distortion wave with 2 32 order harmonics The number of harmonics frequency the amplitude and the phase of harmonics to the fundamental wave can be programmed Using high power precision operational amplifier the working freq...

Page 5: ...acy Grade 0 05 accuracy in the range of rated power ranges Reduction of reactive power accuracy by one level Output waveform distortion At the AC voltage and current output waveform distortion is 0 3 Watt hour meter measurement Comprehensive difference 0 05 FS pulse output error 90kHz Three phase rated value 60kHz DC output DC voltage range 75mV 150mV 300mV 3V 6V 10V output capacity 100mA per file...

Page 6: ...reading 0 04 full Grade 0 1 Other Working power 220Vac 10 50Hz 5 Use environment Temperature 20 10 Deg C relative humidity is 85 RH Volume weight 460 X 480 x 200mm about 18kg IV Familiar with the device Front Panel Features Figure 1 1 DC output terminal 2 AC voltage output terminal 3 AC current output terminal need short connection when not working 4 Display screen 5 Function selection key 6 Numer...

Page 7: ...nd output terminals for watt hour meter calibration See figure 3 Figure 3 Pulse Input GND grounding 5V providing optoelectronic head power supply FL low frequency pulse input Pulse Output GND grounding FH high frequency pulse output FL low frequency pulse output ...

Page 8: ...ation Turn it on at least 30 seconds after shutdown If a continuous beep is emitted while the installation is working or the amplitude does not rise the computer must be turned off for inspection The sequence of shutdown is to press the disconnect button before pressing the power switch When a three phase two element meter is calibrated the common black end of the voltage of the B phase connection...

Page 9: ...on the right side of the LCD display to enter the corresponding submenu Figure 5 Main menu window Ac output operation In the main menu press the AC output key to enter the AC Standard Source Operation interface and on the right side of the AC Standard Source Operation interface enter the AC Source output setting through key Figure 6 Ac operation window ...

Page 10: ...f the current range shortcut directly select the range 1 0 1A 2 0 5A 3 1A 4 5A 5 10A and 6 20A through the digital key Voltage and current amplitude adjustment press the corresponding key the voltage amplitude is 0 highlighting changing the setting fineness by moving the cursor on the left and right keys changing the magnitude through the up and down key and setting the amplitude through the outpu...

Page 11: ...power frequency total power cycle to count sampling pulse Err error Pre last error Samps cumulative number of cycles and S values measured by the built in standard The lower half is used for shortcut key settings The setting of the power error detection operation area is the same as that of the AC output operation area Output low frequency power constant Measuri ng range 50V 100V 200V 400V 0 1A 72...

Page 12: ...4 3V 5 6V 6 10V 7 75V 8 150V 9 300V 0 500V Current path shortcut key definition Select range directly through numeric keys 1 30uA 2 300uA 3 3mA 4 30mA 5 300mA 6 1A 7 5A 8 20A Voltage and current amplitude adjustment According to the corresponding key the voltage amplitude 00 000 is highlighted The right and left keys move the cursor to change the setting fineness and the up and down key to change ...

Page 13: ...X3030B manual 12 VI Standard configuration 1 Hosts 1 2 Test line 1 3 RS232 communication line 1 4 Pulse line 1 5 power cable 1 6 CD 1 7 Aluminum alloy box 1 8 manual 1 9 Test Report 1 10 Certification 1 ...