background image

 

2915 133

rd

 Street West • Shakopee, MN 55379 • (p) 952.445.4841 • (t) 800.237.3603• (f) 952.445.7615 

www.gammavacuum.com  • [email protected]

 

 

 
 
 
 
 
 

Titanium Sublimation Pump & Remote 
TSP Pump Controller Users Manual 

 
PN 900016, Rev C 
 
 
 
 
 

Specification information is located on our website at: 

 

www.gammavacuum.com 

 
 
 
 
 
 
 
 
 
 

 

 
 
 
 

ISO 9001:2000 Certified

Summary of Contents for TSP

Page 1: ...41 t 800 237 3603 f 952 445 7615 www gammavacuum com info gammavacuum com Titanium Sublimation Pump Remote TSP Pump Controller Users Manual PN 900016 Rev C Specification information is located on our...

Page 2: ...ration 12 Description 12 Menu Display 12 TSP Configuration Config Menu 13 TSP Settings Menu 15 TSP Sublimation Screen 17 Procedures and Additional Information 18 Cold Start Procedure 18 Degassing Fila...

Page 3: ...the filaments The titanium serves to capture chemically reactive non noble gases Sublimation requires heating of the elements which raises the vacuum system temperature temporarily raising system pres...

Page 4: ...trol of a TSP titanium sublimation pump or NEG non evaporable getter The remote controller operates in conjunction with the DIGITEL Ion Pump Power Supply Line It has the capability of operating up to...

Page 5: ...5 titanium 15 molybdenum alloy wire Total length of each wire mm in 219 8 6 Coiled length of wire mm in 117 4 6 Weight per filament 3 1 to 3 5g Filament life up to 20 hours depending on operating cond...

Page 6: ...s per TSP Control Modes Manual Programmed Auto Filament Advance Disabled Next Rotate Power Control Watts or Amps Power Output max 75 Amps 1000 watts Voltage Output max 14 volts Power Ramp Rate amps se...

Page 7: ...lowing specifications as listed in the Official Journal of the European Communities EN 50081 1 Emissions EN 55011 Class A radiated and Conducted Emissions EN 50082 1 Immunity IEC 801 2 Electrostatic D...

Page 8: ...MMA VACUUM CONTROL UNITS DESIGNED FOR ION PUMP OPERATION ARE CAPABLE OF DELIVERING 7000 VDC UNDER OPEN CIRCUIT OR LOW PRESSURE OPERATING CONDITIONS GAMMA VACUUM PRODUCTS ARE DESIGNED AND MANUFACTURED...

Page 9: ...ZE O CONSIDERE REEMPLAZARLO ACHTUNG SCHWERES OBJEKT ZUR VERMEIDUNG VON MUSKELZERRUNGEN ODER R CKENSCH DEN BEIM TRANSPORT GEEIGNETE HUBVORRICHTUNGEN UND HEBETECHNIKEN VERWENDEN WARNING READ AND UNDERST...

Page 10: ...ontroller Installation Procedure The Remote TSP Controller requires a communication cable with the DIGITEL MPCe for proper operation Required Items You need the following items to install the TSP DIGI...

Page 11: ...s in the preferred order using the M3 set screw The orientation of the wires to the connectors is not relevant for proper TSP operation and is governed by technician preference 10 Reinstall the protec...

Page 12: ...The TSP parameters are controlled through the MPCe Menu screen The TSP functions are separated into 2 areas 1 TSP Config configuration This section displays and controls possible hardware configuratio...

Page 13: ...ly overwritten Re Scan The Re Scan button initiates a scan of the TSP hardware and automatically sets the TSP Configuration selected based on the hardware detected In the event that a scan does not se...

Page 14: ...etected on that high current wire filament 4 in the above example A forward slash indicates that the filament is installed but is not achieving the desired level of current This often occurs during th...

Page 15: ...independently and maintains the set amperage for the set on time Note Degassing will activate all installed filaments TSP Settings Menu The TSP settings menu allows control over software parameters f...

Page 16: ...advances to the next sequential filament 4 Level A when in edit mode this value determines the maximum amperage or wattage applied to the filament Amps or watts can be selected in the edit screen 5 O...

Page 17: ...rammed and Manual Mode Once the TSP has been properly configured and the settings entered the TSP will show the appropriate parameters in the TSP firing screen This screen is entered by pressing the T...

Page 18: ...w filaments or ones that have been up to air must be outgassed into the roughing system The following procedure forces the major gas load from the filaments out of the system A slight release of gas o...

Page 19: ...ur 2 and three 3 3 8 silver plated gaskets for use with either a 3 or 4 filament TSP respectively Wrench a special wrench designed to prevent bending of posts during filament installation To replace t...

Page 20: ...volts by observing product labeling This information is available on the serial number block of the unit and near the J1 connection Contact Gamma Vacuum if a change in input power is required MPC MPC...

Page 21: ...of cycles Z ZE ZZ is the upper pressure A AE AA is the lower pressure DDD is the duration of each firing is seconds W is set to 1 to ignore pressure automatic cycle period 28 TSP OFF Disables TSP fir...

Page 22: ...el and the controlling parameter either Amps or Watts MPCe only R XXX P R the corresponding TSP 1 or 2 XXX watts or amps P W watts or A amps 31 GET QUAL PRESSURE 1 Reads the programmed TSP qualifying...

Page 23: ...ent 6 Single TSP 8 filament 7 Dual TSP 3 filaments 8 Dual TSP 4 filaments 88 SET IND MODE Enables and disables the TSP independent mode YES or NO 89 GET IND MODE Returns the TSP independent mode statu...

Page 24: ...al investment To do this our service technicians require as much information as possible about the vacuum system in need of support To assist in this process please provide the following information i...

Page 25: ...of it s products other than in accordance with proper operating and safety procedures EXCEPT AS STATED HEREIN GAMMA VACUUM MAKES NO WARRANTY EXPRESSED OR IMPLIED EITHER IN FACT OR BY OPERATION OF LAW...

Page 26: ...ess may result A standard restocking fee may apply Note Prior to issuance of an RMA the required documents must be submitted to Gamma Vacuum Required Documentation During a lifetime of system operatio...

Page 27: ...defined as any substance that because of its properties is not compatible with ultra high vacuum UHV operation Some of these are silicon in the form of silicones sulfur cadmium fluorine and chlorine...

Reviews: