background image

January 2017 

© 2017 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. 

All product names are trademarks of their respective companies. 

64 Max 

IR Thermometer 

Users Manual 

1.888.610.7664 

[email protected]

Fluke

-

Direct

.com

Summary of Contents for 64 max

Page 1: ...ration All rights reserved Specifications are subject to change without notice All product names are trademarks of their respective companies 64 Max IR Thermometer Users Manual 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 2: ... your nearest Fluke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUS...

Page 3: ...Features 5 Overview 6 Settings Menu 6 Basic Features 10 Advanced Features 12 EMS setting 12 ALARM 13 Select MAX MIN AVG DIF 14 SAVE and Review the Records 16 Lock the Measurement 19 Trigger Lock 19 Auto Capture 20 Set Temperature Units 22 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 4: ...ii Set Date and Time 23 Aiming and Measurement 24 Maintenance 27 Change the Battery 27 Clean the Product 27 Specifications 29 Safety 30 Electromagnetic Compatibility 31 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 5: ...er the Product can determine the surface temperature by measuring the amount of infrared energy radiated by the target s surface Warning Read all safety information before you use the Product 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 6: ... on the Product and in this manual Warning To prevent eye damage and personal injury Read all safety Information before you use the Product Do not use the Product if it operates incorrectly Use the Product only as specified or the protection supplied by the Product can be compromised Before you use the Product inspect the case Do not use the Product if it appears damaged Look for cracks or missing...

Page 7: ...pes Optical tools can focus the laser and be dangerous to the eye Do not look into the laser Do not point laser directly at persons or animals or indirectly off reflective surfaces Replace the batteries when the low battery indicator shows to prevent incorrect measurements Do not use the Product around explosive gas vapor or in damp or wet environments Use the Product only as specified or hazardou...

Page 8: ...suant to Laser Notice 50 dated June 24 2007 In addition the following pattern on the label will indicate wavelength and optical power λ xxxnm x xxmW This product complies with the WEEE Directive marking requirements The affixed label indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE ...

Page 9: ...perature displays 1 AA battery Adjustable emissivity Auto off Printed 64 Max Instructions 24 hour clock Appointment measurement Monitoring measurement Ambient alarm Lower battery alarm IP54 3 meter drop Last reading Hold 7 seconds High and low alarm Data storage and review Trigger lock Trigger protection 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 10: ...e Product see Figure 1 For a view of the laser and LCD display see Figure 2 Settings Menu To put the Product into HOLD status pull and release the trigger shows on the display When the Product is in HOLD status push the SEL button once to enter the first level settings Top level settings include L Light Backlight Laser ALARM HI and LO EMS Emissivity MAX MIN DIFF AVG R Record 1 888 610 7664 info Fl...

Page 11: ...IR Thermometer Overview 7 iba02 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 12: ...64 Max Users Manual 8 iba01 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 13: ...tings include TRIG L Trigger Lock L Lock AUTO CAPTURE TIME AUTO CAPTURE INTERVAL DATE TIME TEMPERATURE UNIT From either menu push SEL to step through each feature The last push goes back to HOLD status To set a feature or parameter pull the trigger to save the option and return to HOLD status 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 14: ... status push SEL to enter the first or second level menu 2 Push SEL several times until the feature shows on the display 3 Push the SET button to change the ON or OFF status 4 Push SEL several times or pull the trigger to quit the process and return to HOLD status Note The Product laser is for aiming purposes only The laser turns off when you release the trigger 1 888 610 7664 info Fluke Direct co...

Page 15: ...IR Thermometer Overview 11 iba14 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 16: ...curacy To adjust the EMS value See Figure 6 and Table 2 1 In HOLD status push SEL several times until EMS and its value are shown in the upper right corner 2 Push or to adjust the EMS value as necessary 3 Push SEL several times or pull the trigger to quit the process and return to HOLD status 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 17: ...s to the higher or lower extremes of the ambient temperature range In this condition the readings do not reflect actual temperature and should not be trusted 3 Push SEL to change to a high or low alarm the display shows HI or LO 4 Push SET to turn the alarm feature On or OFF When the HI LO alarm is on the alarm value shows in the lower right of the display 5 Push or to change the setting value as ...

Page 18: ...D status push SEL until MAX MIN AVG DIF shows 2 Push SET to select MAX MIN AVG or DIF 3 Push SEL several times or pull the trigger to exit the setup The feature selected is shown on the display Note You can select only one of MAX MIN AVG or DIF 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 19: ...IR Thermometer Overview 15 iba15 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 20: ...means that the reading has been saved into memory at the two digit location The Product also saves a time and date stamp and the EMS value To review saved readings 1 In HOLD status push SEL to step through the first level menu until Rshows on the display The latest record shows on the display The record includes the reading EMS value and the date and time stamp 2 Push or to review each record A tw...

Page 21: ...is deleted After the deletion all remaining records move one position forward For example if the deleted record was in position 08 then the record on 09 becomes the new 08 record To delete all saved readings In record review status push and hold SET for approximately 10 seconds At 3 seconds the display flashes to warn that all data will be lost Continue to push SET to delete all the records To can...

Page 22: ...30 Iron 0 70 Asbestos 0 95 Lead 0 50 Asphalt 0 95 Oil 0 94 Brass 0 50 Paint 0 93 Ceramic 0 95 Plastic 0 95 Concrete 0 95 Rubber 0 95 Copper 0 60 Sand 0 90 Food Frozen 0 90 Steel 0 80 Food Hot 0 93 Water 0 93 Wood 0 94 Oxidized Opaque over 20 mils Natural Factory Setting 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 23: ... stop the measurement as necessary The Product returns to setup status Trigger Lock This feature conserves battery power when the Product is stored For example if the Product is stored inside a toolbox so that the trigger is pulled for 10 minutes the Product turns itself off to save the battery To use this feature 1 In HOLD status push SEL for approximately 3 seconds to enter the second level menu...

Page 24: ...easurement 1 In HOLD status push SEL 3 seconds to enter the second level menu 2 Push SEL until AUTO CAPTURE and TIME show on the display 3 Push SET to switch the feature on or off 4 When it is on push or to adjust the time 5 Push SEL several times or pull the trigger to exit the setup Note When the Auto Capture feature is on AUTO CAPTURE shows on the display when the Product is in SCAN and HOLD st...

Page 25: ...cond level menu 2 Push SEL until AUTO CAPTURE and INTERVAL show on the display 3 Push SET to switch the feature on or off 4 When the feature is on push or to adjust the displayed value as necessary The displayed value is the start value of the countdown counter 5 Push SEL several times or pull the trigger to exit the setup Note If the Interval monitoring feature is turned on when Auto Capture is a...

Page 26: ...data will be saved starting from the first empty memory location until all 99 positions are used Set Temperature Units To set the temperature units 1 In HOLD status push the SEL button for 3 seconds to enter the second level menu 2 Push SEL several times until F or C shows on the display 3 Push SET to toggle the units 4 Push SEL or pull the trigger to accept the format 1 888 610 7664 info Fluke Di...

Page 27: ...te adjustment 5 The active item flashes push or to adjust the value and then push SEL to move to the next item 6 When the date is complete push SEL until TIME shows on the display 7 Push SET to toggle a 12 hour or 24 hour display 8 Push SEL to accept the time format and start the time adjustment 9 The active item flashes push or to adjust the value and then push SEL to move to the next item 10 Pus...

Page 28: ...64 Max Users Manual 24 Aiming and Measurement Figures 10 through 13 show how to aim and take measurements with the Product iba17 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 29: ...IR Thermometer Aiming and Measurement 25 iba03 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 30: ...64 Max Users Manual 26 gtt04 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 31: ...hown in Figure 14 Note To prevent the loss of the set date time and status of appointment and monitoring remove and insert a new battery when the Product is in sleep mode and complete the operation within 10 seconds Clean the Product Use soap and water on a damp sponge or soft cloth to clean the Product case Carefully wipe the lens surface with a moist cotton swab The swab may be moistened with wa...

Page 32: ...64 Max Users Manual 28 gtt13 eps 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 33: ...00 ms 95 of reading Spectral Response 8 microns to 14 microns Emissivity 0 10 to 1 00 Temperature Coefficient 0 1 C C or 0 1 C of reading whichever is greater Optical Resolution 20 1 calculated at 90 energy Display Resolution 0 1 C Repeatability of reading 0 5 of reading or 0 5 C whichever is greater Power 1 AA IEC LR06 Battery Battery Life 30 hours with laser and backlight on 1 888 610 7664 info ...

Page 34: ...Non Condensing 10 C 90 RH at 10 C to 30 C 75 RH at 30 C to 40 C 45 RH at 40 C to 50 C Operating Altitude 2000 meters above mean sea level Storage Altitude 12 000 meters above mean sea level Safety General IEC 61010 1 Pollution Degree 2 Laser IEC 60825 1 Class 2 650 nm 1 mW Ingress Protection IEC 60529 IP54 1 888 610 7664 info Fluke Direct com Fluke Direct com ...

Page 35: ...tential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances Caution This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments Korea KCC Class A Equipment Industrial Broadcasting Communication Equipment Class A Equipment meets requirements for ...

Reviews: