myAirvo™ 2 Quick Start Guide

60

60

90

a

b

c

Nasal cannula

Auto-fill chamber

60

a. Remove the blue port 

tabs and water supply 

tube holder. 

b. 

Install the provided 

elbow adapter to the 

MR290 water chamber. 

c. Clip the water supply 

tube into position.

Please refer to Section 2, “Setting up myAirvo 2”, and Section 3, “Using myAirvo 2”, of your 

 

myAirvo 2 User Manual for information on setup.

Please refer to Section 4, “Cleaning and maintenance”, of your myAirvo 2 User Manual for information on 

consumable parts. myAirvo 2 has five parts which you will need to replace regularly to keep your device 

as clean as possible. The times recommended below assume your myAirvo 2 is only being used by one 

person, and rely on the correct cleaning procedures being followed.

1

2

Op nasal cannula

Replace every 60 days.
Product  code:  MYOPT9SMALL  (2-pack)                           

  MYOPT9MEDIUM (2-pack)

   

  MYOPT9LARGE (2-pack)

AirSpiral™ tube and auto-fill chamber kit

Replace every 60 days.

Product code: MYAIRVOKIT1

Reorder code:

Air filter

Replace every 90 days or 1000 hours.

Product code: 900PT913 (2-pack)

Water bag

Replace every 60 days.

Product code: 900PT401 (2-pack)

Important note: This is a Quick Start Guide only and should not be considered in any way a replacement 
for your myAirvo 2 User Manual. Before using your myAirvo 2 device you must read your myAirvo 2 User 
Manual to understand the device’s intended use, warnings and cautions. myAirvo 2 is not intended for life 
support.

Some products may not be available in your country. 

Please contact your local Fisher and Paykel Healthcare representative.

Place your myAirvo 2 on a 

low shelf or near the floor 

beside your bed. Make 

sure it is placed below your 

head height and flat.

REF 6

129
70 REV D 20

21-
10 
©

 20
21 Fisher & P

ayk

el Healthcar

e Limit

ed

Contact details for support and ordering

www.fphcare.com/myairvo

4. Consumable parts

1. How to set up your device

Fit the water chamber onto 

your device by depressing 

the finger guard, pushing 

forward until the finger 

guard clicks into place.

3

a

b

a. Hang water bag from 

the home stand. 

b. Push water supply tube 

spike into the fitting at 

the bottom of the water 

bag. Open vent cap. 

c. Water will enter MR290 

water chamber and be 

contained below the 

max water level line.  

4

Attach the breathing tube 

to your myAirvo 2.

5

Plug myAirvo 2’s power 

cord into the mains/utility 

power socket and into the 

rear of the unit.

6

Switch on myAirvo 2 by 

pressing and holding 

the On/Off button for 5 

seconds.

7

myAirvo 2 will begin to 

warm up, displaying a 

warm-up symbol on the 

screen.

8

When myAirvo 2 is ready 

for you to use, a tick 

symbol will be displayed.

9

Connect your interface 

and begin therapy. During 

use, if excess condensate 

forms in the tube, refer to 

Condensate Management 

in Section 2, “Setting up 

myAirvo 2”. 