background image

www.emkometer.com 

 

 

Page   

L21/3 dial type| Emko

meter s.r.o., Na Žižkově 1245, 58401 Ledeč n/S., Czech Republic 

 

1

 of 

10

      tel/fax: 569 721 622 | tel: 569 720 539, 569 721 549 | [email protected] 

 

 

 

 

 

 

 

 

 

Measurement and Regulation

 

 

 

 

 
  

EMKOMETER Ltd.

  

 

  

   

Operating instructions   

   

Liquid level sensor   

  

 

  

      Ledec nad Sazavou 

  

   

Series L21/3-DIAL-xx.xx.xx 

   

   

 

  

 

  

 

  

   

 

  

Contents:    

1.  MEASUREMENT PRINCIPLE AND APPLICATIONS   
2.  DESCRIPTION   
3.  OPERATING CONDITIONS   
4.  SPECIFICATION   
5.  OTHER DATA   
6.  STORAGE, DELIVERY AND TRANSPORT   
7.  INSTALLATION / ADJUSTMENT OF MECHANICAL PARTS, COMMISSIONING   
8.  SETTING THE ELECTRICAL PARTS, MAINTENANCE   
9.  TABLE OF FLOAT SENSORS   
10.  QUALITY WARRANTY   
11.  OTHER INFORMATION REQUIRED FOR PRODUCTION AND SETTING-UP   

    
   

1. Principle and Applications   

   

1.1 Principle  
 

The float sensor consists of a head, neck, wire and a float. Optionally float is coupled to the wire via 
magnetic follower over guide tube. The liquid level is measured by the float that moves freely along the 
tube, actuating the follower via a magnetic coupling. The magnetic coupling consists of magnets placed 
in the float and in the follower. In the case of turbulent levels, extreme temperatures etc., a follower with 
a more powerful magnetic coupling is used. The follower located in the guide tube is connected to the 
instrument head by stainless-steel wire. The follower moves through the magnetic field, transmitting the 
liquid level movement to the instrument head. In simple applications with noncorosive liquids directly 
connected float is used. The transmission mechanism indicates instantaneous liquid levels on a circular 
scale, or it controls liquid levels at the desired height by means of the LG

´s electronic equipment. Using 

a resistance transmitter or a transmitter with a potentiometer, it provides an impedance signal or a 
current output signal.   
   

1.2 Use    
Float sensors (hereinafter FS) are designed to measure and control  liquid levels in closed and open 
tanks  or  in  pressure  tanks.  Some  types  of  instruments,  designed  and  marked  for  explosion-free 
atmospheres,    can  also  be  used  to  measure  various  media  in  Zone  0.    IP  65  allows  for  safe 
measurements  in  the  amospheres  with  aggressive  substances.  The  level  gauge  is  a  mechanical-
electrical  device  whose  modular  construction  allows  you  to  create  various  instruments  with  different 
features.    

   

Types of level gauges and their combinations 
 
  

 

LG head (basic information) 

    

Содержание L21/3-DIAL Series

Страница 1: ...of turbulent levels extreme temperatures etc a follower with a more powerful magnetic coupling is used The follower located in the guide tube is connected to the instrument head by stainless steel wi...

Страница 2: ...te the head diameter in mm The next three digits indicate the head depth 2 2 Guide tube The guide tube is made of stainless steel or polypropylene The guide tube commonly called guide rod is closed at...

Страница 3: ...polypropylene float has been designed for aggressive fluids 3 Operating Conditions 3 1 Operating Conditions 3 11 The instrument design ensures safe operation in the following areas a cold b hot c for...

Страница 4: ...s address B Instrument type LBA XX XX XX C Year of production e g 2002 D Serial No E Country of origin Czech Republic F 5 32 Product plate located on the rear cover or on the instrument side This pla...

Страница 5: ...mounted immediately cover the guide rod flange The place above the DN50 PN16 flange or the screwed connection must allow for inserting the guide rod into the flange hole The guide rod dimensions corr...

Страница 6: ...instrument pointers To set a zero position when the float is at the stop pin loosen the screw and set the pointers as necessary We recommend measuring the actual level Increase the liquid level in the...

Страница 7: ...at the level indicated by the pointers Check the cams for switching The other leading edge of the cam e g S4 should not close the contact on the other side of the leading edge when reaching the h1 lev...

Страница 8: ...rusty or the gasket has been damaged execute a medium repair 8 43 Medium repair after approx 6 years of operation The medium repair is executed by the manufacturer The instrument is dismounted and al...

Страница 9: ...al rail Dial indicator standard code L21 3 DIAL xx x x x x 01 up to 5m 02 up to 10m 03 up to 20m 0 number of level switches 2 2 level switches 4 4 level switches 0 no microswitch 1 1 microswitch 3 NO...

Страница 10: ...21 622 tel 569 720 539 569 721 549 info emkometer cz Measurement and Regulation 10 3 Specifications are subject to change without prior notice Any changes shall be listed in the attached document Manu...

Отзывы: