1.2 General advice
• Check if the contents of the package is umdamaged and complete.
• Remove the protection foil above the start button and the two LEDs.
• For cleaning the instrument please do not use an abrasive cleaner only a dry or
wet piece of soft cloth. Do not allow any liquid to the interior of the device.
• Please store the measuring instrument in a dry and clean place.
• Avoid any force like shocks or pressure to the instrument.
• No responsibility is taken for irregular or incomplete measuring values and their
results, the liability for subsequent damages is excluded!
1.4 Scope of delivery and usage
LOG32TH/LOG32T/LOG32THP series loggers are suitable for recording, alarm
tracking, and display of temperature, humidity*, dew point* (*only LOG32TH/THP) and
barometric pressure (only LOG32THP) measurements. Areas of application include
the monitoring of storage and transport conditions or other temperature, moisture
and / or pressure-sensitive processes. The logger has a built-in USB port can be
connected without cables to all Windows PCs. The USB port is protected by a
transparent plastic cap. The green LED flashes every 30 seconds during recording.
The red LED is used to display limit alarms or status messages (battery change ...
etc.). The logger also has an internal buzzer that supports the user interface.
1. Start button, 2. LED green, 3. LED red, 4. battery case, 5. USB-connector, 6. USB-
cover, 7. wall holder, 8. Slits ... this is where the sensor is located, 9. protective foil
1.3 Overview
see Fig.A
Press button for 2 seconds,
beeper sounds for 1 second
LED lights green for 2 sconds - logging
has started!
LED blinks green every 30 sec.
2. First Start & Begin Recording
1.6 Ready to use
The logger is already preset (see 5 default settings) and ready for start. It can be used
immediately without any software!
1.5 For your safety
This product is exclusively intended for the field of application described above.
It should only be used as described within these instructions.
Unauthorized repairs, modifications or changes to the product are prohibited.
2.1 Restart recording
see 2. The logger is started by default by button and stopped by USB port plug-in.
The measured values are plotted automatically to the PDF file.
NOTE: When you restart the existing PDF file is overwritten.
Important! Always secure the generated PDF files on your PC.
3. Stop recording / Create PDF
Connect logger to USB port.
Beeper sounds for 1 second. The
recording stops.
LED blinks green until result PDF is
created (can take up to 40 seconds).
Beeper sounds and LED stays green.
Logger is shown as removeable drive
LOG32TH/LOG32T/ LOG32THP.
View PDF and save.
PDF will be overwritten with next
log start!
5. Standard settings / Factory settings
Note the following default settings of the data logger before first use. By using the
LogConnect* software, the setting parameter can easily be changed:
Interval: 5 min. LOG32TH/ LOG32THP, 15 min. LOG32T
Start possible by: Key press
Stop possible by: USB connect
Alarm: off
4. Description of PDF result file (see Fig. B)
X
A
Filename: e.g.
LOG32TH_14010001_2014_06_12T092900.DBF
LOG32TH:
Device
14010001:
Serial
2014_06_12:
Start of recording (date)
T092900:
time: (hhmmss)
Description:
Log run info, edit with LogConnect* software
Configuration:
preset parameters
Summary:
Overview of measurement results
Graphics:
Diagram of measured values
Signature:
Sign PDF if required
:Measurement OK :Measurement failed
B
C
D
E
F
G
* Download free LogConnect software:
www.dostmann-electronic.de/home.html ->Downloads ->Software/<your logger>/
Software/LogConnect_XXX.zip (XXX choose latest version)
GB
1. Introduction
Thank you very much for purchasing one of our products. Before opera-
ting the data logger please read this manual carefully. You will get useful
information for understanding all functions.
Schnellstart
mit Standardeinstellungen
Aufzeichnung starten. Gedrückt halten bis
Signalton ertönt.
LED blinkt grün
(alle 30 Sek.)
Logger in USB port
warten
PDF anschauen und sichern
1.
2.
3.
4.
5.
6. Battery replacement
Twist rear cap (about 10°),
the battery lid opens.
Remove empty battery and insert
new battery like shown.
Battery change ok:
both LEDs light for 1second, beep sounds.
7. Alarm signals
Logger in record mode
Beeper sounds once each 30 seconds for
1 second, red LED blinks each
3 seconds - measured values exceeds
selected measurement range
(not with standard settings).
Alarm limits can be changed using
LogConnect* software.
Logger in standby mode (not in record mode)
Red LED blinks once each 4 seconds.
Replace battery.
8. Waste disposal
This product has been manufactured using high-grade materials and components which can
be recycled and reused.
Never dispose of empty batteries and rechargeable batteries in household waste.
As a consumer, you are legally required to take them to your retail store or to an
appropriate collection site depending on national or local regulations in order to
protect the environment. The symbols for the heavy metals contained are:
Cd=cadmium, Hg=mercury, Pb=lead
This instrument is labelled in accordance with the EU Waste Electrical and
Electronic Equipment Directive (WEEE). Please do not dispose of this instrument in
household waste. The user is obligated to take end-of-life devices to a designated
collection point for the disposal of electrical and electronic equipment, in order to
ensure environmentally-compatible disposal
This sign certifies that the product meets the requirements of the EEC directive and
has been tested according to the specified test methods.
Technical changes, any errors and misprints reserved. Stand05_CHB1806
Logger in standby mode (not in record mode)
Red LED blinks once each 4 seconds.
Replace battery.
Red LED blinks twice or more
each 4 sconds. Hardware fault!
* Download free LogConnect software:
www.dostmann-electronic.de/home.html ->Downloads ->Software/<your logger>/
Software/LogConnect_XXX.zip (XXX choose latest version)
NOTE: Check Logger status: Press start button for appr. 1 second. If the green
LED flashes twice the logger is recording! This procedure can be done as often as
you want.
1.1 Delivery contents
• Data logger LOG32
• Wall holder
• Battery 3,6 Volt
• USB protection cap
• 2x screws and dowels (already inserted)
ATTENTION! Please observe our battery recommendation strictly. Use only the battery
type LS 14250 3.6 volt of the manufacturer SAFT or DYNAMIS Lithium Batt. LI-110 1/2
AA/S, respectively only batteries authorized by the manufacturer.
5005-0170 / 5005-0172 / 5005-0173
LOG
32
T
/
LOG
32
TH
/
LOG
32
THP
Bedienungsanleitung
Operating Instructions
Datenlogger für Temperatur, Feuchte
*
und Druck
**
Data logger for temperature, humidity
*
and pressure
**
DOSTMANN
electronic GmbH · Waldenbergweg 3b
D-97877 Wertheim · www.dostmann-electronic.de
*nur LOG32TH/LOG32THP / *only LOG32TH/LOG32THP
**nur LOG32THP / **only LOG32THP
9. Marking
only LOG32T
CE-conformity, EN 12830, EN 13485, Suitability for storage (S) and transportation
(T) for food storage and distribution (C), Accuracy classification 1 (-30..+70°C),
according to EN 13486 we recommend a recalibration once per year.