background image

Technical Specifications

PILATUS3 R CdTe 100K-M

Detector System

Document Version v1.0.0

DECTRIS Ltd.

5405 Baden-Daettwil
Switzerland
www.dectris.com

Summary of Contents for PILATUS3 R

Page 1: ...Technical Specifications PILATUS3 R CdTe 100K M Detector System Document Version v1 0 0 DECTRIS Ltd 5405 Baden Daettwil Switzerland www dectris com ...

Page 2: ...Document Version v1 0 0 Copyright 2019 DECTRIS Ltd ...

Page 3: ...ty 8 4 DETECTOR DIMENSIONS AND CONNECTORS 9 4 1 PILATUS3 R CdTe 100K M Detector 9 4 1 1 Technical Drawing 9 4 1 2 Detector Head 12 4 1 3 Detector Electronics Unit 13 4 1 4 Vacuum Feedthroughs Optional 15 4 2 Detector Control Unit 16 4 2 1 Configuration 16 4 2 2 Samba Share 16 4 3 Thermal Stabilization Unit 16 4 3 1 In Vacuum Operation for Detectors with Optional Vacuum Compatibility 17 5 INSTALLING...

Page 4: ...24 7 4 Vacuum Operation for Detectors with Optional Vacuum Compatibility 25 7 5 Storing the Detector 25 7 6 Cleaning and Maintenance 25 7 7 Safety Instructions Cadmium Telluride 26 8 TROUBLESHOOTING 27 9 CERTIFICATION TESTS 28 10 SERVICE FORM 29 PILATUS3 R CdTe 100K M Technical Specifications v1 0 0 ii 29 ...

Page 5: ...ocument Version Date Status Prepared Checked Released v1 0 0 2019 06 28 release DJ LG TD BL SB Changes Table 2 Changes to this Document Version Date Changes v1 0 0 2019 06 28 First Release of PILATUS3 100K M PILATUS3 R CdTe 100K M Technical Specifications v1 0 0 iii 29 ...

Page 6: ...se please contact us via telephone e mail or fax 1 2 Explanation of Symbols Danger 0 Danger blocks are used to indicate immediate danger or risk to personnel or equipment Warning 0 Warning blocks are used to indicate danger or risk to personnel or equipment Caution 0 Caution blocks are used to indicate danger or risk to equipment Information 0 Information blocks are used to highlight important inf...

Page 7: ...rom the use of this system DECTRIS is the sole owner of all user rights related to the contents of the manual in particular information images or ma terials unless otherwise indicated Without the written permission of DECTRIS it is prohibited to integrate the protected contents in this publication into other programs or other websites or to use them by any other means DECTRIS reserves the right at...

Page 8: ...ctor please refer to the order confirmation or check for the presence of the label vacuum tested on the serial number sticker at the detector backside A check in the box of vacuum tested implies the available vacuum option of the PILATUS3 R CdTe 100K M detector Please avoid organic materials and highly out gassing compounds inside the vacuum chamber as they tend to deposit on sensitive detector com...

Page 9: ...overflow state 1 048 573 Maximum count rate 1 107 photons s pixel Energy range2 15 keV to 80 keV Adjustable threshold range2 8 keV to 40 keV Number of thresholds 1 Maximum frame rate 20 Hz Information 1 When using the external trigger or external enable mode the detector will not acquire an image if the ef fective frame rate is above 20 Hz 1 The sensor thickness of your actual system can be found i...

Page 10: ... Supply Unit The PILATUS3 R CdTe 100K M is delivered with the power supply unit TreNew SINPRO MPU130 105 It is a switching power supply Use only the included power supply Please consult the user documentation of the TreNew SINPRO MPU130 105 power supply unit for details 3 1 4 Detector Control Unit The PILATUS3 R CdTe 100K M is delivered with the detector control unit DELL PowerEdge R230 It is a ra...

Page 11: ...low level Caution 4 Absolute maximum is 5 V Applying a higher voltage will damage the detector External trigger input impedance 50 Ω Detector trigger output 5 V max current 100 mA 3 2 2 Power Supply Unit Table 3 4 Power Supply Unit Ratings Power input 100 VAC to 240 VAC 47 Hz to 63 Hz 1 58 A to 0 64 A Power output 12 V DC max 10 84 A 130 W AC connector IEC 320 C14 input inlet Dimensions 89 5 mm x ...

Page 12: ...ors may not prevent damage temperature and humidity should be monitored to avoid breaching the operation limits Information 2 The relative humidity within the module chamber must be lower than 30 during operation and lower than 25 during start up use of dry air or nitrogen advised The PILATUS3 R CdTe 100K M detector is designed for indoor use only The ambient conditions shown in table 3 7 must be ...

Page 13: ...see if a detector is vacuum compatible check whether the vacuum compatibility option is included in the order confir mation received from DECTRIS Warning 3 Only systems with a vacuum compatibility option are allowed to be operated in vacuum Please contact support dectris com for information regarding vacuum compatibility upgrade For in vacuum operation of the detector following conditions must be f...

Page 14: ...ction with this document Without our agreement they may be neither copied nor made accessible to a third party DECTRIS LTD 5400 Baden Switzerland A2 1 1 gek 1 1 0 27 01 2017 prepared for air Zeichnung 162700 M 00 L02 Pilatus3 100K M Air Outline ISO 2768 mK ISO 13920 BF General tolerances Prepared for air views without cover 83 764 487x0 172 33 54 195x0 172 88 20 4 37 17 to center of sensor 44 to c...

Page 15: ...d A2 1 2 erb 1 1 0 13 01 2017 Massbild Zeichnung 121400 M 80 L10 Electronics for P3 K Systems ISO 2768 mk ISO 13920 BF General tolerances 50 125 38 5 M6x1 9 5 TIEF 206 6 202 151 155 5 102 95 56 51 3 16 Center of gravity Center of gravity Rev Beschreibung Description Datum Date genehm approv 0 Drawing created 13 01 2017 erb Figure 4 2 Drawing of the PILATUS3 R CdTe 100K M Detector electronics unit ...

Page 16: ...ugness europ Projektion Bezeichnung Zeichnungsnummer Title Drawing No Datei File Format 17 06 2013 Revision Page Gewicht Pages 1 of 25pin 1of 50pin SUB D F T weight JB 1 210S D25 D50 ISO100 DEC Standard Toleranz F Products SEC02 21 218 F Products SEC02 21 218 All informations provided in this drawing is considered to be confidential proprietary and secret information It is intended only for the us...

Page 17: ...4 5 The PILATUS3 R CdTe 100K M detector head with the protective cover in place front view Danger 1 Danger of electric shock Do not touch the Mylar foil The sensors behind the Mylar foil are operated at high voltages Touching the Mylar foil can cause an electrical shock and damage of the sensors Caution 5 The cover may not protect the detector from a direct beam Figure 4 6 The PILATUS3 R CdTe 100K...

Page 18: ...able to the detector electronics unit vacuum feedthrough blue Samtec SQCD ribbon cable DRY AIR Dry air or nitrogen for humidity control For details see section 5 4 Pipe Use a pipe with outer diameter of 4 mm IN Coolant inlet OUT Coolant outlet 4 1 3 Detector Electronics Unit Figure 4 8 The PILATUS3 R CdTe 100K M detector electronics unit viewed from the front left and back right PILATUS3 R CdTe 10...

Page 19: ...e is a power failure or if the de tector temperature or humidity is out of the limits Table 4 4 Connectors for cables on the back plane of detector electronics unit Connectors Description DATA RJ45 Cat 6A S FTP cable for data transfer Caution 6 Note that there must be a 1 Gbit point to point con nection between detector and PC Gb2 12 V DC Main voltage 12 V DC from the external power supply EXT IN ...

Page 20: ...length of 2 m is not guaranteed The cooling hoses are convoluted stainless steel tubes suitable for static applications but not for dynamic flexing applications Caution 7 The cable of the feedthrough set should not be combined with the default cables of the detector system Figure 4 9 Picture of the optional PILATUS3 R CdTe 100K M electrical vacuum feedthrough viewed from the vacuum left and air sid...

Page 21: ...rewall and SE Linux are disabled by default otherwise the following ports must be open for UDP 52010 52011 52012 The following firewall port must be open if you want to connect to Camserver with a TCP IP socket connection from the outside 41234 4 2 2 Samba Share There is a Samba share configured on the detector PC The Samba service is enabled by default and running after the system boots The storage...

Page 22: ... of thermal stabilization unit Use opaque hoses to avoid the growth of algae Do not set the temperature of the thermal stabilization unit below the recommended operating temperature Condens ing moisture can develop and damage the detector 4 3 1 In Vacuum Operation for Detectors with Optional Vacuum Compatibility For in vacuum operation unscrew the self sealing valves from the detector and use vacu...

Page 23: ... are described below 5 2 1 Mounting the Detector Head Figure 5 1 Picture of the PILATUS3 R CdTe 100K M detector head with extended mounting brackets viewed from below left and front right Mounting the Detector Head from Above Use the mounting brackets as depicted in figure 5 2 These mounting brackets have to be mounted on the base plate of the detector The detector should be mounted using both oute...

Page 24: ...nted using the two internal M6x1 threads as shown in figure 5 3 indicated with yellow circles Figure 5 3 Drawing of the PILATUS3 R CdTe 100K M detector head base plate a printed copy is supplied in the user documentation folder Left View from below with retracted mounting brackets for mounting from below Right Cross section view through the internal thread to show the 7 mm thread less and the 9 mm ...

Page 25: ...detector control unit and the power supply of the detector have to be connected to a grounded power outlet Although the detector might be already grounded via the mounting bolts the detector should be grounded additionally via the functional ground connector at the back in order to establish a defined grounding 5 4 Connection to Dry Air or Nitrogen Caution 12 Humidity might damage the detector Make...

Page 26: ...ow is necessary Warning 6 When venting with nitrogen take proper precaution against the risk of asphyxiation caused by oxygen displacement from nitrogen Ensure sufficient ventilation and oxygen level monitoring Use compressed dry air for venting large vessels or in confined spaces 5 5 Connection to Thermal Stabilization Unit The PILATUS3 R CdTe 100K M detector is water cooled and must be connected t...

Page 27: ...PILATUS3 R CdTe 100K M cooling vacuum feedthrough set section 4 1 4 5 6 Mounting the Detector Control Unit Caution 13 Make sure that the detector control unit has adequate ventilation The detector control unit can be mounted in a standard 19 inch rack which has to be properly grounded PILATUS3 R CdTe 100K M Technical Specifications v1 0 0 22 29 ...

Page 28: ...ed up thermal stabilization unit temperature set to 25 C prior to pumping down venting and opening the chamber Only use dry air or nitrogen for venting For stable operation a constant detector temperature is recommended After switching on the detector its temperature will reach a constant value after about 30 60 min If the temperature or humidity is out of range following error message appears _ C...

Page 29: ...ot at room temperature wait until the thermal stabilization unit has reached stable operation Connect the power cable the local network cable and the detector data cable to the detector control unit 7 2 Start up Procedure Please use the following start up procedure Turn on the dry air or nitrogen at least 30 min before turning on the detector Then turn ON the power switch at the back of the detect...

Page 30: ...t pumping down a warm detector prevents condensation issues After the thermal stabilization unit has reached the set value switch on the detector electronics unit Venting the Vacuum Chamber Turn OFF the power switch on the detector electronics unit Set the temperature to 23 C on the thermal stabilization unit and let the detector warm up at least 30 min to prevent condensation inside the vacuum ch...

Page 31: ... R CdTe 100K M detector does not require any maintenance 7 7 Safety Instructions Cadmium Telluride The sensor material is built up from Cadmium Telluride CdTe In order to prevent injuries please follow these safety pre cautions Do not inhale CdTe dust when sensor is broken or burnt Do not ingest CdTe Avoid skin contact For recycling and see Section 2 2 PILATUS3 R CdTe 100K M Technical Specification...

Page 32: ...tor control unit and the detector are powered and correctly connected the LINK LED should be green Takes up to 30 s after power up The detector fails to turn on The power cord is not connected or the plug is incompletely inserted Connect the power cord firmly Check the green LED on the external power supply The fuse is blown Replace the fuse see table 3 3 The temperature is over the critical limit ...

Page 33: ...fication Tests for PILATUS3 R CdTe 100K M EN 61000 6 2 2016 IEC 61000 6 2 2005 ed2 0 industrial EN 61000 6 4 2007 A1 2011 IEC 61000 6 4 2011 ed2 1 industrial EN 61010 1 2010 IEC 61010 1 2010 AMD1 2016 EN 61326 1 2013 IEC 61326 1 2012 ed2 0 PILATUS3 R CdTe 100K M Technical Specifications v1 0 0 28 29 ...

Page 34: ...10 SERVICE FORM PILATUS3 R CdTe 100K M Technical Specifications v1 0 0 29 29 ...

Reviews: