background image

User manual for

DATAMAN-848PRO2

Extremely fast universal 8x 48-pindrive Stand-Alone concurrent multiprogramming
system

Internal PC is driven by Microsoft Windows XP Embedded

September 2017

Summary of Contents for DATAMAN-848PRO2

Page 1: ...User manual for DATAMAN 848PRO2 Extremely fast universal 8x 48 pindrive Stand Alone concurrent multiprogramming system Internal PC is driven by Microsoft Windows XP Embedded September 2017 ...

Page 2: ... modified in any form on any medium for any purpose Information provided in this manual is intended to be accurate at the moment of release but we continuously improve all our products Please check for an updated manual on our website at www dataman com Dataman Programmers Ltd assumes no responsibility for misuse of this manual Dataman Programmers Ltd reserves the right to make changes or improvem...

Page 3: ...test version from our website www dataman com Conventions used in the manual References to the control program functions are in bold e g Load File Device etc References to control keys are written in brackets e g F1 Terminology used in the manual Device any kind of programmable integrated circuits or programmable devices ZIF socket Zero Insertion Force socket used for insertion of target device Bu...

Page 4: ...ming at the moment the chip is detected to be inserted in the socket properly independently of the status of other programming modules This results in seven programming modules working while you replace the programmed chip in the eighth Graphical control unit with touch screen provides basic control and easy monitoring of the programming flow Modular construction of hardware the programming module...

Page 5: ...perate down to 1 8V so you ll be ready to program the full range of today s advanced low voltage devices DATAMAN 848PRO2 performs device insertion test based on the check of proper signal path between the programmer and programmed device before it programs each device In dependence on programming configuration it identifies missed or poor contact between programmed device and the ZIF socket of the...

Page 6: ...e a three year warranty on parts and labor for the programmer limited 25 000 cycle warranty on ZIF sockets Although we take as much care as possible to integrate high quality components in the built in computer we can t provide a higher warranty as we get from suppliers So there is a limited warranty of one year on all computer parts such as touch screen motherboard CPU HD etc DATAMAN 848PRO2 elem...

Page 7: ...et device due to incorrect user selected programming parameters Target device may not be destroyed by forced interruption of the control program turn off programmer but the content of programmed cells may remains undefined Don t unplug the target device from the ZIF socket during work with device BUSY LED on In system serial programming by DATAMAN 848PRO2 Optimized advanced pindrivers deliver prog...

Page 8: ...cription of ISP connector 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 Front view at ISP connector H L read driver RA1 pin of ISP connector RA2 RA3 Pull up Pull down Read Float H L drivers in programmer CC1 RC1 pin of ISP connector Pull up Pull down RC2 Read Float H L drivers in programmer RB1 RB2 pin of ISP connector drivers in programmer pins 3 5 7 9 11 13 of ISP connector pin 14 of ISP co...

Page 9: ...ip and target system Specification of ISP connector pins depends on the device which you want to work with You can find it in the control SW for programmer PG4UW menu Device Device Info Ctrl F1 Note the ISP programming ISP option of respective device must be selected It is indicated by ISP suffix after the name of selected device These specifications correspond with application notes published by ...

Page 10: ...tors 2 enclosed in the standard delivery package Selftest of programmer Insert 48 pin diagnostic POD type I into ZIF socket of the programmer Take care about orientation of 48 pin diagnostic POD type I see top side of POD Run selftest of programmer in PG4UW Programmer Selftest plus Selftest of ISP connector Insert Diagnostic POD for ISP connector 2 into ZIF socket of the programmer Take care about...

Page 11: ... VPP2 range 0 26V 1A autocalibration selftest selftest capability Socket pindriver 48 pin DIL ZIF Zero Insertion Force socket accepts both 300 600 mil devices up to 48 pin pindrivers 48 universal VCCP VPP1 VPP2 can be connected to each pin perfect ground for each pin FPGA based TTL driver provides H L CLK pull up pull down on all pindriver pins analog pindriver output level selectable from 1 8 V u...

Page 12: ...x 29Cxxx 29Fxxx 29GLxxx 29BVxxx 29LVxxx 29Wxxx 49Fxxx series Samsung s K8Fxxxx K8Cxxxx K8Sxxxx K8Pxxxx series EPROM NMOS CMOS 27xxx and 27Cxxx series EEPROM NMOS CMOS 28xxx 28Cxxx 27EExxx series 3D Plus 3DEExxxxxxxx mDOC H3 SanDisk ex M Systems SDED5xxx SDED7xxx MD2533xxx MD2534xxx Hynix HY23xxx FRAM Ramtron MRAM Everspin MRxxxxx8x 3D Plus 3DMRxxxxxxxx NV RAM Dallas DSxxx SGS Inmos MKxxx SIMTEK ST...

Page 13: ...ntrollers Cypress CY7Cxxxxx CY8Cxxxxx Microcontrollers ELAN EM78Pxxx Microcontrollers EPSON S1C17 series Microcontrollers Explore Microelectronic EPF01x EPF02x series Microcontrollers Generalplus GPM8Fxxx series Microcontrollers GreenPeak GPxxx series Microcontrollers Infineon Siemens XC800 C500 XC166 C166 series Microcontrollers MDT 1xxx and 2xxx series Microcontrollers Megawin MG87xxx MPC82xxx s...

Page 14: ...xxxxx ZLP323xxxxxxx series Microcontrollers other EM Microelectronic Spansion Fujitsu Goal Semiconductor Hitachi Holtek Novatek Macronix Princeton Winbond Samsung Toshiba Mitsubishi Realtek M Square ASP Coreriver Gencore EXODUS Microelectronic Topro TinyARM VersaChips SunplusIT M Square QIXIN Signetic Tekmos Weltrend Amic Cyrod Technologies Ember Ramtron Nordic Semiconductor Samsung ABOV Semicondu...

Page 15: ...TM8A S L series SPC5 Power Architecture Microcontrollers Silicon Laboratories Cygnal C8051 series Microcontrollers Programmable System Memory STMicroelectronics uPSD PSD series Microcontrollers TI MSP430 series both JTAG and BSL MSC12xxx series CC430 series LM4F series TM4C series Microcontrollers ZILOG Z8Fxxxx Z8FMCxxxxx Z16Fxxxx series ZLF645x0xx Various PLD also by Jam VME SVF STAPL Player JTAG...

Page 16: ...crease programming times considerably If you plan to compare ask always which pattern they use The programming speed depends on the PC speed only slightly SOFTWARE Algorithms only manufacturer approved or certified algorithms are used Algorithm updates software updates are available regularly approx every 4 weeks free of charge Internet download OnDemand version of software is available for high p...

Page 17: ...andard JESD 71 VME ispVME file VME2 0 VME3 0 GENERAL operating voltage AC 90 264V 47 63Hz power consumption max 300W active dimensions 625x465x115 mm 24 6x18 3x4 5 inch weight programmer 14 9kg 32 85 lbs temperature 5 C 35 C 41 F 95 F humidity 20 80 non condensing Package included Standard accessories DATAMAN 848PRO2 programmer diagnostic POD for ZIF socket selftest of the programmer 1x diagnostic...

Page 18: ... use all properties of this software and can make a project for mass production Standard device related commands read blank check program verify erase are boosted by some test functions insertion test connection check signature byte check and some special functions autoincrement production mode start immediately after insertion of chip into socket All known data formats are supported Automatic fil...

Page 19: ...t external display keyboard and mouse to PC connectors on back side of programmer Turn on programmer 2 Run the control program double click on icon 3 Find DATAMAN 848PRO2 Site programmer Ctrl F or right click on panel Programmer ...

Page 20: ...20 Select DATAMAN 848PRO2 Site and then click on Connect button 4 Select site Select desired DATAMAN 848PRO2 Site and then click on OK button 5 Select device click on icon and select desired device ...

Page 21: ...fer from file click on icon 7 Set Device operation options Set desired operations for the device and then click on OK button 8 To customize device use menu Alt S Set desired settings for the device and then click on OK button ...

Page 22: ...opy project file s from external drive for example from network drive USB key etc to the standalone s local Projects folder see the progress of device operation including statistics information serialization Step by step instruction Starting Stand Alone multiprogramming system 1 Check if the power switch placed at back side of programmer near power supply connector is on position 0 2 Plug the powe...

Page 23: ...ator is displayed during connection of sites Additional information about site connection progress is displayed for each site in panel Status on main screen 13 Programmer sites are numbered from 1 to 8 in direction of physical placement of programmer sites from left to right 14 When connection process of programmer sites is complete indicator Connecting Sites will be hidden Panel Status will conta...

Page 24: ...Selected device operation including sub operations that are part of predefined device operation will start Progress of operations is displayed in panel status for each programmer site You can use some special features of Stand Alone multiprogramming system for example Count down or Statistics ...

Page 25: ...e screen is shown during booting up of operating system and starting of standalone control software PG4UWMC Welcome screen About screen Welcome screen is displayed on Stand Alone build in LCD screen after successful starting of standalone control software The screen can also be displayed by using menu command Help About for more information see Help About Welcome screen can be closed by clicking o...

Page 26: ...for each Site If serialization is not used the Serialization column remains empty Sites that are not connected have displayed in the Status column Connected Sites have their Status displayed Following are typical values Ready means the programmer site is connected and ready to start any operation Present special status of connected Site means the site is connected but it is not selected checked in...

Page 27: ...ecent Project name Currently loaded project file name Checksum of data stored in data buffer for currently selected device Button Run can be operation name for example Program Verify Erase Click on button Run will start selected device operation for all connected programmer sites Button can be used when no device operation is running on any programmer site Button Stop Click on button Stop will sto...

Page 28: ...king on Load button Change dir button is used to change current directory to directory selected in Directory list on the left side of dialog When folder is selected Change dir button jump up from current directory one level one folder up Note The drive combo box is automatically refreshed each time it is dropped down It is useful for loading project files from USB key which can be plugged temporar...

Page 29: ... can be plugged temporarily to USB port of standalone programmer and consequently the new drive letter assigned with USB key will appear in list of available drives When option 2 is selected the dialog with a list of project files present in Projects folder will appear User can select desired project and delete selected project by clicking on Delete button The dialog can be closed by Cancel button...

Page 30: ...e To switch back to production mode just close PG4UW control program and run program PG4UWMC by double clicking on PG4UWMC icon on the Desktop Note Engineering mode requires connection of external VGA monitor keyboard and mouse Do not use this menu option without these peripherals connected and working File Shut down computer Command is used to close control program and to shut down built in compu...

Page 31: ...s used to reinitialize software configuration and to re find available programmer sites There is no need to use this menu command in normal operation Situations when the command is useful are 1 when first starting DATAMAN 848PRO2 Stand Alone multiprogramming system 2 when error message Programmer Sites not found is displayed In the dialog Search Programmers use button Search to search for attached...

Page 32: ...ation Present special status of connected site means the site is connected but it is also not selected checked in dialog Select sites User selected device or project operations will not run on Present unselected sites User can set site selection in dialog Select sites accessible by menu command Programmer Select Enabled sites No project file is loaded means the site is successfully connected but n...

Page 33: ...atic YES function is displayed on the bottom left side of LCD main screen with label Automatic YES Is On by following If the Automatic YES function is Off no label about Automatic YES is displayed For more details about Automatic YES feature see documentation chapter Programmer Automatic YES Programmer Select Enabled Sites Command displays dialog Select Enabled Sites This dialog is used to select ...

Page 34: ... recent Operator Mode settings obtained from ini file are used If you wish to see more information about Protected mode of single application PG4UW please look at Options Protected mode There is menu command Options Switch to Operator Mode that allow using Operator Mode in PG4UWMC After selecting the menu Switch to Operator Mode dialog appears User has to enter password twice to confirm the passwo...

Page 35: ...etimes when Administrator Mode is switched from Operator Mode some commands for example command Load project may remain disabled This can be resolved by clicking on button Stop ALL Operation Load project for protected project files uses project defined allowed device operations to enable disable device operations in PG4UWMC Also Load project operation can be automatically disabled in Operator Mode...

Page 36: ...ed on built in disk of standalone programmer The Log file names used are reportmc log PG4UWMC main Log file reportmc_ 1 log Site s 1 Log file reportmc_ 2 log Site s 2 Log file reportmc_ 3 log Site s 3 Log file and so on Following options can be set for Log file creation option Append Log file sets usage of Log file Log file will be created after the first restart of PG4UWMC For all following execu...

Page 37: ...unchecked the size of Log file is unlimited respectively is limited by free disk space only option Maximum Log file size specifies the maximum size of Log file in kBytes option Amount of truncated text specifies the percentage of Log file text which will be truncated after Maximum Log file size is reached The higher value means more text will be truncated removed from Log file Job report Job Repor...

Page 38: ... few points at different positions Each of the point has to be clicked by user After clicking required number of points touchpad calibration is complete which will be announced by information message displayed Sounds Dialog Sounds settings allows user to select the sound mode of program Program generates sounds after some activities e g activities on device programming verifying reading etc Progra...

Page 39: ...Erase and or Verify operations is counted as one operation The next function of statistics is Count down Count down allows checking of the number of device operations and then the number of devices on which device operations have to be done After each successful device operation the value of count down counter is decremented Count down has user defined start number for devices to do When count dow...

Page 40: ...lp Index Menu command displays short help on built in touch screen The short help contains brief information about menu commands and basic operations with usage of touch screen Help Problem report Menu command Create problem report is used for writing more particular diagnostic information to Log window and consequently copy Log window content to problem report file Problem report is useful when a...

Page 41: ... USB key which can be plugged into USB port of Stand Alone programmer When USB key is plugged to USB port of programmer new drive letter assigned with USB key will appear in list of available drives in dialog Select installation package User can than select this drive and run installation package from the USB key When user confirms another confirmation dialog update process will start During updat...

Page 42: ...There are two major kinds of administration of the OS network and local OS is configured with two users accounts admin operator Credentials for DATAMAN 848PRO2 Stand Alone multiprogramming system users are Login Password Administrator admin saadmin Operator operator saoperator Network administration Connecting to network OS in programmer is configured for operating with TCP IP network To obtain co...

Page 43: ...Connection program on your computer In Computer field enter IP address from Welcome or About screen of standalone programmer When communication is established enter administrator credentials Because current OS allows only one user connected at a time next message may be showed Warning Connecting to programmer in time when programming devices is running programming will be immediately terminated yo...

Page 44: ...aved log files from operations with devices When OS update files are available they should be saved to Updates directory Your project files should be in Projects directory Exact update steps will be distributed with operating system updates For programmer reliability operating system is configured with write filters on Recovery and System disk As a side effect of these filters any configuration ch...

Page 45: ...you accept this agreement Accepting agreement starts formatting of disk D and installation of control program recovery deletes Programmer directory on disk E and creates new one Also directories Updates and Temp has been erased and recreated When recovery is completed you should see System recovery completed Then press any key and system will restart First start of the operating system after recov...

Page 46: ...er being relevant about the programmer software and the target device Mail fax Print and complete the RETURNS FORM available from our website www dataman com Include everything that you consider being relevant about the programmer software and the target device Send the completed form by mail or fax to DATAMAN fax number in the control program menu Help About or to your local dealer Phone Call you...

Page 47: ...cal injury you must connect ground wire to black banana jack for connection to ground marked with ground symbol on the back site of programmer If you have an unsupported target device If you need to operate on a target device not supported by the control program for programmer please do not despair and follow these next steps Look in the device list of the latest version of the control program on ...

Page 48: ... and display A lot of dust or grime may cause an unwanted touch on touch screen Perform the calibration test using 48 Pins Calibration test POD Type I Biannual maintenance Check amount of dust or dirty in programmer If you feel that a lot of dust or dirty is in programmer clear it by vacuum cleaner The warranty does not apply to any damage of programmer or its part caused by dust or dirt Warning T...

Page 49: ...e warranty period the customer must prove the date of purchase These warranty terms are valid for customers who purchase a programmer directly from Dataman The warranty conditions of Dataman sellers may differ depending on the target country law system or Dataman seller s warranty policy The warranty does not apply to products that are of wear and tear or mechanically damaged Equally the warranty ...