Operation: 

1. Insert the mouthpiece in the opening on the left side. 

2. Press the Start button. 

  

3. A countdown takes place from 150. 
4. Once "

blo

" is displayed, evenly blow into the 

    mouthpiece until you hear a signal tone. 
    

  If you do not blow within 20 seconds,

 

      the unit automatically switches off.  

5

After the measurement, the result of the breath test is shown  

    for approx. 20 sec. on the display. 
    After approx. 15 seconds, the display becomes dark. By 

    pressing the Start button again,

 the unit prepares 

    for a new test (s. point 3). Another breath test can be 
    performed. 
       

 If the test result is over 0.50‰, the

 

        LCD display illuminated red and "WARN" is displayed. A

 

        warning tone is emitted. 

6. The CA 2020 automatically switches off after  
    approx. 20 seconds. 

"OFF" 

appears in the display and 

    a signal tone can be heard. 
7. Please start at point 1 for more tests. 
 

We recommend waiting at least 2 minutes between each test. 
The device is not designed for a series consisting of more 
than ten consecutive measurements. 

The measuring principle

 

Blood and breath alcohol have a close correlation with 
each other. In the depths of the lungs, in the alveolar, 
blood and breathing air are separated by a thin, gas-
permeable membrane. This is where the gas exchange 
takes place. 

Oxygen is absorbed by the blood 

and carbon

 

dioxide and alcohol, if present, is emitted into the 
breathing air. To obtain this air from the depth of the 
lungs (alveolar air) for the analysis, a specific minimum 
quantity of breathing air must be provided. 
 (10 litres per minute of air volume = 5 seconds of strong 
blowing). 
A correct measurement is only possible if the person 
blows really strongly and sufficient breathing air is 
exhaled. 

 

Technical data 
 

Sensor  

 

Electrochemical sensor 

Accuracy 

 

±0.05 ‰ BAC 

 

 

 

 at 1.00 ‰ BAC 

Detection range 

0.00 ~ 4.00‰ BAC 

Warm-up time  

15 ~ 35 seconds 

Response time  

3 ~ 22 seconds 

Dimensions   

125 mm x 56 mm x 24 mm 

Weight  

 

87 g (131 g with battery) 

Power supply   

9V alkaline battery 

Temperature range  5  ~ 40

 

Operating temperature

 

Storage

 

temperature  0  ~ 40  

Calibration 

 

Every 6 months  

 

 

 

or after approx. 500 tests 

Warranty 

 

12 months for device 

 

 

 

6 months for sensor 

 
 

COSMOS ALKOHOLTESTER GMBH 

AM IHLSEE 18 - D 23795 BAD SEGEBERG 

Tel.: 0049 (0) 4551 / 83838   
Fax: 0049 (0) 4551 / 87171

 

E-mail: [email protected]

 

The device is ready for use 
when 

READY

 appears in t

he display.

BATT

 Signal indicates

 

low battery voltage. 

Display in ‰ 

The warning display
appears with results 

 

over 0.50‰.