New GentleYAG

 Cryogen Spray Alignment Procedure  

Candela Corporation 

 

 

8503-01-0845, Revision A 

Candela Corporation Proprietary 

 

Page 1 of 10 

New GentleYAG                                        

Cryogen Spray Alignment

 

 Procedure 

Summary of Contents for New GentleYAG

Page 1: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 1 of 10 New GentleYAG Cryogen Spray Alignment Procedure...

Page 2: ...late 1301 00 8291 available from Candela Corporation VPYAG distance gauge 050 inch Allen Wrench The following tests are described in this section 1 Cryogen Alignment Verifies the cryogen spray nozzle...

Page 3: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 3 of 10...

Page 4: ...3109 4 Select 30ms of DCD Spray 5 Point hand piece away from objects and personnel toward the floor View the contact ring of distance gauge looking from the hand piece 6 Press purge button DCD Spray s...

Page 5: ...oprietary Page 5 of 10 Results Acceptable alignment no further action needed Unacceptable alignment repeat the test with a different distance gauge If the test shows acceptable alignment with the new...

Page 6: ...ended to represent treatment parameters but rather provide a check on proper functionality of the handpiece and provide a reference for the user to help identify changes in the handpiece operation Pro...

Page 7: ...Proprietary Page 7 of 10 Distance gauge spot size Typical spray duration to minimally fill the inside ring spot 3 mm 30 ms 8 mm 30 ms 10 mm 30 ms 12 mm 40 ms 15 mm 50 ms 18 mm 60 ms Results If the spo...

Page 8: ...empty Warning Put on appropriate laser safety glasses 2 Turn on the system and allow the WARM UP to complete 3 With the Electrical Connector installed disconnect the Cryogen Connector see Figure 3 El...

Page 9: ...You may continue use of this handpiece If PURGE REQUIRED does not occur This particular handpiece failed the test Discontinue use of this handpiece and call Candela Tech Support at 508 358 7637 ex 336...

Page 10: ...New GentleYAG Cryogen Spray Alignment Procedure Candela Corporation 8503 01 0845 Revision A Candela Corporation Proprietary Page 10 of 10...