background image

 

 

 

 

    

fundus

 

link

 

model: 20D001 

 

 

instructions 

manual

 

 
 

 

Содержание fundus link 20D001

Страница 1: ...funduslink model 20D001 instructions manual...

Страница 2: ...fundus for the purposes of general visualization for the retina or back of the eye It can be operated by health care professional be used for patient as adult Contraindications Because prolonged inte...

Страница 3: ...temporary visual disturbances particularly in patients with a history of light sensitivity or migraines Multiple flashes per eye in a single exam may increase the likelihood of occurrence Clean and d...

Страница 4: ...y C MER RainsOptics Limited or by a centre approved by C MER RainsOptics Limited Do not the device in direct sunlight Strong ambient light may affect the results Do not use automatic cleansing machine...

Страница 5: ...tery Do not disassemble modify crush or destroy the battery pack Doing so can cause battery fluid leakage heat generation burns fire and or explosion Do not replace of battery with an incorrect type t...

Страница 6: ...observed to verify that they are operating normally Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inche...

Страница 7: ...t generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications However there is...

Страница 8: ...at comply with Innovation Science and Economic Development Canada s licence exempt RSS s Operation is subject to the following two conditions 1 This device may not cause interference 2 This device mus...

Страница 9: ...Cet quipement est conforme aux limites d exposition aux radiations d finies par le Canada pour des environnements non contr l s Cet metteur ne doit pas tre install au m me endroit ni utilis avec une...

Страница 10: ...compliance 6 Trademarks 8 Table of Contents 9 Parts and Feature of Fundus Link 10 Procedures for before use 12 Procedures for Mounting the Fundus Link onto the iPhone 14 Investigation 15 Cleaning Fun...

Страница 11: ...IFU 20D001 20220913 Page 10 Parts and Feature of Fundus Link 1 2 3 4 5 6...

Страница 12: ...or will be flashing Bluetooth Connection Blue LED indicator will be constant on Preview Mode Green LED indicator will be constant on Low battery Amber LED indicator will be flashing Battery charging A...

Страница 13: ...ging should take approximately 4 hours Please connect the charger to the Fundus Link and to the mains During the charging the YELLOW indicator light is constant on The YELLOW indicator light will turn...

Страница 14: ...nnection To initiate the Bluetooth connection put both Fundus Link and the smartphone within 0 3 meter 1 foot from each other 1 Power on the Fundus Link the GREEN indicator is flashing 2 Switch on you...

Страница 15: ...Fundus Link onto the iPhone 1 Pull out the bottom clamp 2 Insert the iPhone and make sure the alignment bar on the mount have aligned to the side wall of the iPhone Note Make sure the camera on the i...

Страница 16: ...IFU 20D001 20220913 Page 15 Investigation WARNING Ensure the Fundus Link DO NOT contact with the subject eye Or use the Eyecup for protect the subject eye...

Страница 17: ...ark environment allows pupils naturally dilated during investigation 1 Ensure the Bluetooth connection is well connected between Fundus Link and Smartphone BLUE indicator is constant ON 2 Press the SH...

Страница 18: ...ding the device and pressing the shutter with your Left hand Ask the patient to look at your Right ear for fixation Ask the patient to look at your Left ear for fixation Touching the patient s forehea...

Страница 19: ...dilation 7 Release the SHUTTER button for flash lighting and remote control the smartphone to capture the image CAUTION Capturing failures could be due to limited pupil size blinking and alignment is...

Страница 20: ...m the device Remove the eyecup from the device Using a commercially blower to blow away dust when it adheres to the lens And using a soft cotton cloth wipe the lens gently in a circular motion Fundus...

Страница 21: ...or The eyecup should be replaced when Discolored Deteriorated Shattered cracked or disintegrated CAUTION Do not use cleaners that contain organic solvents to clean the Fundus Link body or lens Fundus...

Страница 22: ...blow away dust when it adheres to the lens To remove dirt grease or fingerprints clean with a soft cotton cloth rubbing in a circular motion Use of coarse cloth or unnecessary rubbing may scratch the...

Страница 23: ...ing Diode LED Operating Environment 10 C to 40 C 30 to 75 700 1060hpa Storage Transport Environment 10 C to 55 C 30 to 75 700 1060hpa Compatible smartphones model Apple iPhone 12 Power Source Recharge...

Страница 24: ...yecup Soft pouch Battery charger USB charging cable Instruction for use Soft cloth Dust sweeper Lens cover Activation Card Device Class This product has been tested to meet the requirements under ISO...

Страница 25: ...IFU 20D001 20220913 Page 24 Troubleshooting TBD...

Страница 26: ...electrical equipment designed primarily for indoor use Symbol for Consult Instructions for Use or Consult Instruction Manual Symbol for Follow instruction for use Do not operate or leave this assembl...

Страница 27: ...tering the European market to indicate conformity with the essential health and safety requirements set out in European Regulation US Federal Communication Commission Waste Electrical and Electronic E...

Страница 28: ...ct Information Manufacturer C MER RainsOptics Limited Manufacturer Address Room 607 6 F 17W of Hong Kong Science Park No 17 Science Park West Avenue Shatin N T Hong Kong Tel 852 35905755 www cmerrains...