background image

 

 
 
 

 

 

 

 
 

 

 

YBC-5  

Passive Vaccine Storage Device 

Maintenance Manual

 

 

 
 

 

 

Summary of Contents for YBC-5

Page 1: ...YBC 5 Passive Vaccine Storage Device Maintenance Manual ...

Page 2: ...al 2 Device Features 2 Device Information 2 Safety precautions 2 MAINTENANCE CLEANING 3 General instructions 3 To clean the exterior of the device 3 To clean the interior of the device 3 TROUBLESHOOTING 4 DETAILED PARTS ASSEMBLY VIEW 5 LIST OF PARTS 6 ...

Page 3: ...o 43 C Vaccine storage temperature range Internal temperature in vaccine payload area 0 C to 10 C Capacity Volume of internal payload space 5 4 L Unit Dimension OD x H Outside diameter and height φ528 x 747 mm Net Weight Unloaded 26 5 Kg Gross Weight Fully loaded with packages 32 5 Kg Safety precautions Device is designed to store vaccine Do not use for other purposes e g do not use for storing li...

Page 4: ...ior of the device To clean the interior of the Arktek take these steps 1 Remove the cap 2 Remove the vaccine stacks Follow established cold chain vaccine handling practices with any removed vaccines 3 If needed clean the vaccine cup stack containers with soap and water and allow them to dry 4 Remove and set aside the ice blocks 5 Use mild soap and water or a 1 chlorine bleach solution to clean the...

Page 5: ...came off or wire broken 2 LCD display and or main PCB assembly have failed 3 Battery is dead 1 Reconnect or replace with new components 2 Replace with new components 3 Replace with a new battery The Err message displays for a long time 1 The connection of the data cable has come off 2 The data cable is broken 1 Reconnect 2 Replace with a new one Visual and or audible alarm The interior temperature...

Page 6: ...5 DETAILED PARTS ASSEMBLY VIEW ...

Page 7: ...s Pan Head Screw ST3 5 9 5 19 YBC 5 6 2 Battery Clamp 20 YBC 5 6 5 Battery Clamp Foam 21 YBC 5 6 6 Battery 22 91400A107 Stainless Pan Head Screw ST3 5 9 5 23 YBC 5 6 7 Loudspeaker 24 YBC 5 6 9 Snap Button Female 25 YBC 5 2 4 Nylon Rope 2 26 YBC 5 6 8 Shoulder Screw M5 0X10 Shoulder φ 6 35 Length 38mm 27 90945A760 Flat Washer IDφ6 6 ODφ12 Thickness0 8mm 28 YBC 5 6 3 Latch 29 YBC 5 6 2 Lock Inner Br...

Page 8: ...ivider Male 50 YBC 5 3 3 2 Vaccine Cup Divider Female 51 YBC 5 5 1 Ice Block 52 YBC 5 5 5 Thermometer Plug 53 YBC 5 5 6 Thermometer 54 YBC 5 5 2 Ice Block Cap 55 91771A539 Flat Head Screw M6X16 56 YBC 5 0 7 Vaccine Stack Plate Fixed Axis 57 YBC 5 0 5 Vaccine Stack Plate 58 YBC 5 0 10 Stop Compression Sleeve For Wire Rope 59 YBC 5 0 11 Wire Rope 60 YBC 5 1 Dewar ...

Reviews: