Page 1: ...ROPOX Toilet lifter Electric User manual This manual should always be in close proximity of the product TF 200 11 0006 ENG Date 03 05 2023...

Page 2: ...ir the Bluetooth remote 15 4 1 7 Unlock the handcontrol 15 4 1 5 continued 16 4 2 Operating the product 19 4 3 Residual risks 19 4 4 Installation of accessories 20 4 4 1 Installation of height adjusta...

Page 3: ...3 TF 200 11 0006_eng 10 1 Suitable Environments 31 10 2 Adjacent and stacked use 31 10 3 Cables 31 10 4 RF portable equipment 31 11 Accident reporting 32...

Page 4: ...e and installation Always keep this manual in close proximity of the product The use installation and service of this product must be in compliance with this manual to avoid accidents and serious pers...

Page 5: ...the appliance Cleaning and user maintenance shall not be made by children Do not use the product if it has defects or have become damaged before being repaired or replaced If the control box makes unu...

Page 6: ...ectrically operated equipment The use of electrical height adjustable equipment can run at maximum 2 minutes followed by an 18 minute break Type B applied part The product complies with IEC 60601 1 re...

Page 7: ...temperature range 10 C to 50 C Transport and storage 5 C to 40 C Operation Relative humidity 20 to 80 non condensing Materials in contact with patient Components Covers Material description ABS PMMA...

Page 8: ...ly it is straight forward to set to the optimum height for both user and helper It is a solution using modern welfare technology to help users become more self reliant in their bathroom routines The T...

Page 9: ...socket It is recommended to always place the water and electrical connections as shown below In case of deviations it s recommended to place the water and electrical sockets within the hatched areas 4...

Page 10: ...e toilet lifter with a wall drain it s recommended to place the drain within the hatched area not deviating further than the shown tolerances This ensures the best flushing Ropox supplied toilets Shor...

Page 11: ...d rawlplugs suitable for mounting in a concrete wall The fitter should always consider the material condition and strength of the wall and use screws and rawlplugs suitable for the specific wall type...

Page 12: ...lugs Ensure that the center of the bottom 2 holes is at least 130mm above the floor b Connect water hose on wall outlet and electric outlet before securing frame to the wall c Secure frame to wall If...

Page 13: ...e drain bellow as instructed below The examples given below to the left are configurations that will work but as a general rule the installation interface in 4 1 2 should be followed Make sure the bel...

Page 14: ...11 0006_eng d2 For 40 46027 If drain is in wall mount the drain bellow in wall drain mount the accompanying brackets on the wall so that the toilet flange is held in place Make sure the bellow is not...

Page 15: ...nnection travel freely The electrical connection shall travel unhindered and may not get entangled in other installations Note If the control doesn t drive after pairing it means it needs to be unlock...

Page 16: ...16 TF 200 11 0006_eng 4 1 5 continued e Cut threaded rods to length and secure them in the frame f Mount inner cover 1 and 2 2x...

Page 17: ...st be mounted beforehand See 4 4 5 g Mount outer cover if the version with fixed control in the top of the cover is being installed please jump to section 4 4 4 mounting of fixed control in top of cov...

Page 18: ...ength and mount them i Mount the toilet and press pad j Finish the installation by checking the following 1 Rigidity test sit on the toilet and check screw connections are firm 2 Check that full trave...

Page 19: ...uring that no children or adults are sited under the product as severe injury from entrapment may occur 4 3 Residual risks Warning Be aware that there may be no person or body parts under the toilet w...

Page 20: ...ms a Attach housing to the profile with 3x M6x12 and 1x M6x16 Use a locking nut on M6x16 b Make sure that the plastic cover is pressed as far onto the frame as possible Mount 2x M6 buttonhead screws i...

Page 21: ...mbly onto the two screw heads and attach the assembly with the two M6x30 as shown If need be the two M6x30 can be loosened and the support arms can be adjusted in height to better fit the user The scr...

Page 22: ...can twist significantly dismount the assembly and leave 5mm between the underside of the screw head to the plastic cover There should be no clearance between the aluminum bracket and plastic cover Adj...

Page 23: ...ort long is ordered together with Ropox Toilet Lifter there will be pre drilled holes in the cover cap If the back support is ordered later for subsequent mounting new holes must be drilled Please see...

Page 24: ...e Bidet Elegant see TF 200 01 0025 4 4 4 Mounting of fixed control in top of cover a Place the cover on the floor close to the lifting unit Certain parts of this procedure can be easier if the lifting...

Page 25: ...Close the lid and re attach the controlbox d Use the attached strips to secure the wire to the belt as indicated Roll up the remaining wire and strip it together in a bundle as close to the top of th...

Page 26: ...as been ordered it must be mounted prior to mounting the original covers The extension cover comes disassembled to make transport and disposal easier It s mounted onto the original covers by clamping...

Page 27: ...oiler lifter can be ordered with increased crushing safety such that items underneath won t be crushed when driving down With the increased safety in function the lifter will never push with more than...

Page 28: ...ring of remote 4 1 6 Check that all wires are pressed into the sockets Product stopped working Check batteries of remote Check if the power is plugged in Handcontrol doesn t work Check battery power F...

Page 29: ...products Toilet lifter Tradename Solution S90 Sanirens 1 3 peracid For the Toilet lifter with a spiral cable going to its control the following agents have been recommended by its manufacturer for the...

Page 30: ...tuator s are aligned and calibrated from the factory In case they need calibration to run simultaneously use the following procedure Press and hold both UP and DOWN on the remote until the unit has be...

Page 31: ...room of a medical electrical system for magnetic resonance imaging where the intensity of electromagnetic disturbances is high 10 2 Adjacent and stacked use Warning Use of this equipment adjacent to o...

Page 32: ...onsidered to be a contributing factor to the case and 3 The event resulted or could have resulted in one of the following outcomes Death a patient a user or a third party s death A serious deteriorati...