background image

OPERATING INSTRUCTIONS 

MODEL DSA-680 TRMS 

:600A AC/DC Type 

MODEL DSA-1020 TRMS 

:1000A AC/DC Type 

DIGISNAP 

SNAP-AROUND 

1. Features 

 

Designed to meet international safety standards. 

IEC61010-1 & IEC61010-2-032 
Measurement Category (CAT.) IV 600V 
Pollution Degree 2 

 

Double molded main body provides comfortable single 

handed grip

 

Data Hold Function 

 

LCD Backlight function to facilitate working at dimly lit 

situations.

 

REL function to indicate measurement variation 

(Current, voltage, Resistance measurement)

 

MIN/MAX function enables easy reading of min & max 

value during measurement. 

 

PEAK Hold Function enables Peak value measure- 

  ment of starting current. (only at ACA Range)

 

With Continuity & Diode Check Function 

 

Capacity measurement of capacitors 

 

Temperature measurement, switchable between ºC 

and ºF (with K-type temperature sensor) 

 

NCV (

on Contact Voltage) Function for wiring check 

 

600V input protection 

 

Sleep Function to extend battery life 

 

With Bar Graph, 6039 count display

 

2. Safety Warnings 

This instrument has been designed, manufactured and 

tested according to IEC 61010: Safety requirements for 

Electronic Measuring apparatus, and delivered in the 

best condition after passed the inspection. 

This instruction manual contains warnings and safety 

rules which must be observed by the user to ensure safe 

operation of the instrument and retain it in safe condition.   

Therefore, read through these operating instructions 

before using the instrument. 

WARNING 

 

Read through and understand the instructions

 

contained in this manual before using the

instrument.

 

 

Keep the manual at hand to enable quick reference 

 whenever necessary.

 

The instrument is to be used only in its intended

 applications. 

 

Understand and follow all the safety instructions

 contained in the manual. 

 

It is essential that the above instructions are

 adhered to. 

 

Failure to follow the above instructions may cause

injury, instrument damage and/or damage to

equipment under test. 

 

DANGER is reserved for conditions and  

actions

 

that are likely to cause serious or 

fatal injury. 
WARNING is reserved for conditions and 
actions

 

  that can cause serious or fatal injury. 

CAUTION is reserved for conditions and 
actions that can cause injury or instrument 
damage. 

The symbol    indicated on the instrument means that 

the user must refer to the related parts in the manual for 

safe operation of the instrument. It is essential to read the 

instructions wherever the    symbol appears in the 

manual. 

 

Marks listed in the table below are used on this 

instrument.

User must refer to the manual. 
Instrument with double or reinforced insulation
Indicates that this instrument can clamp on   

bare conductors when measuring a voltage   

corresponding to the applicable measurement
category, which is marked next to this symbol.

AC 
DC 
AC & DC 

DANGER 

 

Never make measurement on a circuit in which voltage 

over AC600V exists. 

 

Do not attempt to make measurement in the presence 

of flammable gasses. Otherwise, the use of the 

instrument may   cause sparking, which can lead to an 

explosion. 

 

Transformer jaw tips are designed not to short the 

circuit under test. If equipment under test has exposed 

conductive parts, however, extra precaution should be 

taken to minimize the poss bility of shorting.

 

 

Never attempt to use the instrument if its surface or

your hand is wet. 

 

Do not exceed the maximum allowable input of any 

measuring range. 

 

Never open the Battery cover during a measurement. 

 

The instrument is to be used only in its intended 

applications or conditions. Otherwise, safety functions

equipped with the instrument doesn’t work, and

instrument damage or serious personal injury may be

caused.

 

  WARNING 

 

Never attempt to make measurement if any abnormal
condi ions, such as  broken case and exposed metal parts 
are found on the instrument. 

 

Do not rotate the Func ion Switch while  he test leads are 
being connected. 

 

Do not install substitute parts or make any modification to 
the instrument. For repair or re-calibra ion, return  he
instrument to your local distributor from where it was
purchased. 

 

Do not try to replace the batteries if the surface of  he
instrument is wet. 

 

Disconnect all the cords and cables from the object under 

test and power off  he instrument before opening  he
Battery Cover for Battery replacement.

 

CAUTION 

 

Set the Function Switch to an appropriate position

before starting measurement. 

 

Firmly insert the test leads. 

 

Disconnect the test leads from the instrument for

current measurement. 

 

Do not expose the instrument to the direct sun, high 

temperature and humidity or dewfall. 

 

Be sure to power off the instrument after use. When 

the instrument will not be in use for a long period, 

place it in storage after removing the batteries. 

 

Use a cloth dipped in water or neutral detergent for

cleaning the instrument. Do not use abrasives or

solvents.

Measurement categories (Over-voltage categories) 

To ensure safe operation of measuring instruments, 
IEC61010 establishes safety standards for various 
electrical environments, categorized as CAT.I to CAT.IV, 
and called measurement categories. Higher-numbered 
categories correspond to electrical environments with 
greater momentary energy, so a measuring instrument 
designed for CAT.III environments can endure greater 
momentary energy than one designed for CAT.II. 

CAT.I 

: Secondary electrical circuits connected to an AC 

electrical outlet through a transformer or similar device. 

CAT.II

: Primary electrical circuits of equipment connected 

to an AC electrical outlet by a power cord. 

CAT.III

: Primary electrical circuits of the equipment 

connected directly to the distr bution panel, and feeders 
from the distribution panel to outlets. 

CAT.IV

: The circuit from the service drop to the service 

entrance, and to the power meter and primary over 
current protection device (distr bution panel).

3. Specification

3-1. Measuring range & accuracy 

(accuracy guaranteed at 23Cº±5 Cº, humidity 45~85%) 
AC Current 600A, 1000A Function 

Accuracy 

Function 

Measuring 

Range 

DSA -680 

DSA -1020 

600A 

0-600.0A 

Peak 1500A 

CF=2.5 @ 

600A 

CF=3.0 @ 
500A 

  ±2.0%rdg±5dgt(50/60Hz) 

±3.5%rdg±5dgt(40 ~ 500Hz) 

±5.5%rdg±5dgt(500 ~ 1kHz) 

* Add 2% at CF>2

 

1000A 

0-1000A 

Peak 1500A 

CF=2.5 @ 

600A 

CF=3.0 @ 
500A 

N/A 

DC Current 600A, 1000A Function

Accuracy 

Functon 

Measuring 

Range 

DSA-680 DSA-1020 

600A 0-600 0A 

±1.5%rdg±5dgt 

1000A 600-1000A 

N/A 

±1 5%rdg±5dgt 

AC Voltage Function 
       

Auto-ranging, Input impedance: approx. 10M

 

Accuracy

 

Range 

Measuring 

Range 

DSA-680 DSA-1020 

6/60/600V 

0-600.0V 

±1 5%rdg±4dgt (50/60Hz) 

±3 5%rdg±5dgt (40~400Hz)

 

DC Voltage Function 
       

Auto-ranging, Input impedance: approx. 10M

) 

Accuracy

 

Range 

Measurng 

Range 

DSA-680 DSA-1020 

600mV/6/
60/600V 

0-600 0V 

±1 0%rdg±5dgt 

Resistance (Diode Check/ Continuity/ Capacity) Function 

Accuracy

 

Range 

Measuring 

Range 

DSA-680 DSA-1020 

600

/6k/60k/ 

600k

/6M

 

0-6 000M

 

±1 0%rdg±5dgt 

60M

 

6.00M-60 00M

 

±5%rdg±8dgt 

Cont Buzzer 

0-600 0

 

Buzzer sounds at 100

 or less 

Diode 

Test voltage: 0-2V 

Capacity Function

 

Accuracy

 

Functon 

Measuring 

Range 

DSA-680 DSA-1020 

40nF 

±3.0%±50dgt 

400nF 

4

μ

40

μ

0 01nF 
- 400

μ

Auto-ranging 

±2.5%rdg±50dgt 

Frequency/ DUTY Function

Auto-ranging for Frequency

 

Accuracy 

Range 

Measuring

Range 

DSA-660

DSA-1010

ACA 

40Hz - 400Hz

ACV 

1Hz~10kHz

±0.5%rdg±5dgt 

0.1-99.9% 

(Pulse width/Pulse period) 

±2.5%rdg±5dgt 

Note: Measurable inputs are: 40Vrms@ACV or 
50Arms@AC600A, 350A@AC1000A Range

Temperature Function 

Accuracy 

Function 

Measuring 

Range 

DSA-680

DSA-1020

- 50

º

C ~ 0

º

±5

º

C±3dgt 

º

0

º

C ~ 150

º

±3

º

C±2dgt 

-58

º

F~ 32

º

±7

º

F±2dgt 

º

32

º

F ~ 302

º

±5

º

F±2dgt 

3-2. General Specification 

 

Mode of operation 

 : 

∆∑

 mode 

 

Display 

: max. 6039 counts   

(Frequency: 9999, Capacity & Temperature:4039) & 

 Bar graph 

 

Over-range indication   : “OL” displayed when exceeding 

the measuring range. (except for AC/DCV and 1000A
Function) 

 

Range switching : 

Auto-ranging / Voltage, Resistance, Capacity Range
Single range / Continuity, Diode check, DUTY and
Temperature 

 

Sample rate 

: three times per second 

 

Functional construction : 

 OFF/ ACA/ ACV/ DCA/ DCV/ 

º

C/ 

º

 

Keys : 

SELECT(AC/DC switching & / 

/        /        /        ), PEAK 

HOLD/ Back Light, REL

,Hz/DUTY, MIN/MAX 

 

Power source : DC3V/ R03(UM-4) x 2pcs 

 

Low battery warning    : “     “ mark is displayed at 

2.4V±0.15V or less. 

 

Temperature& humidity : 23ºC±5ºC, relative humidity 

accuracy guaranteed 85% or less (no condensation) 

 

Operating temperature  : 0 ~ 40ºC, relative humidity 85% 

 & humidity range or less (no condensation) 

 

Storage temperature  : -20 ~ 60ºC, relative humidity 

 & humidity range   85% or less (no condensation) 

 

Current consumption  : approx. 25 mA 

 

Sleep Function : Automatically powered off in about 

15 min after the last Function switch operation. Rotate the 
Function Switch from OFF to any position to exit from the 
Sleep state. 

 

Applicable Standards 
IEC 61010-1:2001

Measurement CAT. IV    600V Pollution degree 2 

IEC 61010-031:2002, IEC 61010-2-032 

EMC

EN 61326 

・ 

EN 55022 

・ 

EN 61000-4-2(perfomance criterion B) 

・ 

EN 61000-4-3(perfomance criterion B) 

 

Overload Protection 

Current Range : 720A AC/ 10 sec @ DSA680 

1200A AC/DC/ 10 sec @ DSA1020 

  Voltage Range : 720V AC/DC/ 10sec 

   Resistance Range : 600V AC/DC/ 10sec 

 

Withstand Voltage 

6880V 

AC 

TRMS 50/60Hz

/ 5 sec 

 (between Jaws and electrical circuit/ between internal 
circuit and enclosure) 

 

Insulation Resistance : 10M

 or more/ 1000V 

between electrical circuit and enclosure

 

 

Conductor size 

DSA-680: approx. 33mm 
DSA-1020: approx. 40mm 

 

Dimension 

approx. 254(L)×82(W)×36(D)mm / DSA-1020 
approx. 243(L)×77(W)×36(D)mm / DSA-680 

 

Weight : approx 300 g @ DSA-680 

310g @ DSA-1020 

 

Accessories 

Test Leads 

TL

75 / 1 set 

K-type Temperature Probe 

MODEL8216/ 1 set 

Battery R03

UM-4

/ 2pcs 

Instruction manual  English, Spanish, French / 1pce 
Carrying Case 

C90 

*Effective Value (RMS) 

Most alternating currents and voltages are expressed in 
effective values, which are also referred to as RMS
(Root-Mean-Square) values. The effective value is the
square root of the average of square of alternating
current or voltage values. Many clamp meters using a
conventional rectifying circuit have "RMS" scales for AC 
measurement. The scales are, however, actually
cal brated in terms of the effective value of a sine wave 
though the clamp meter is responding to the average
value. The calibration is done with a conversion factor of 
1.111 for sine wave, which is found by dividing the
effective value by the average value. These instruments 
are therefore in error if the input voltage or current has
some other shape than sine wave. 

*CF (Crest Factor) 

is found by dividing the peak value

by the effective value. 
Examples:     Sine wave: CF=1.414
Square wave with a 1: 9 duty ratio: CF=3 

3-3. Function Keys 

The “

” mark shows available function at each Range. 

HOLD

PEAK

SELECT

ZERO 

Hz/ 

DUTY 

MAX/

MIN  

ACA 

 

 

 

 

 

 

ACV 

 -  - 

 

 

 

DCA 

 - 

 

 - 

 

DCV 

 -  - 

 - 

 

 

 - 

 

 - 

 

- - 

  - - - 

- - 

  - - - 

 - 

 

 -  - 

TEMP

 - 

 

 - 

 

4. Preparation for measurement 

4-1. Checking Battery Voltage 

Set the Function Switch to any position other than “

OFF

”. 

When the display is clear without “

BATT

” mark, showing 

battery voltage is enough. When the display is blank or

BATT

” mark is indicated, replace the batteries according 

to Section 7, Battery Replacement. 

CAUTION 

The Sleep feature automatically powers the   
instrument off in about 15 min after the last switch or   
key operation. Therefore, the display may be blank   
even with the Function Switch set to a position other   
than “OFF”. To operate the instrument in this case,   
turn the switch back to the “OFF” position, then to any   
other position. Replace the batteries if nothing was 

displayed after above operations. 

4-2. Checking Switch Setting & Operation

 

Confirm the Function Switch is set to the correct position, 
the instrument is set to the correct measurement mode,   
and the Data hold function is disabled. Otherwise,   
desired measurement cannot be made.   

5. Measurement 

5-1. AC Current Measurement 

DANGER 

 

Never make measurement on a circuit in which

 voltage over AC600V exists to avoid getting

 electrical shock.

 

Transformer jaw tips are designed not to short the 

circuit under test. If equipment under test has