background image

 

Refine.

082012 

2.

When select “gh”, AL1

AL2 (gh), if CO2 concentration is rising, the voltage/current will decrease.

 

 

Remark: 

Under Green House Mode, the AL1 must >AL2. 
For example: AL1= 900ppm, AL1= 800ppm now, user decrease AL1= 700ppm, AL2 will become 700-Step, 
(AL2= 600ppm, if step=100ppm; AL1= 680ppm, if step= 20ppm). 

 

Specification

 

Method - NDIR 

Sample Method -Diffusion or flow through (50 ~200 ml/min)

   

 

  CO

2

 Specifications:

 

Measurement Range 

0-3,000 ppm display   

Display Resolution 

1ppm at 0~1,000ppm; 10ppm at 1,001~3,000ppm 

Accuracy 

±75ppm or ±5% of reading whichever is greater 

 

Repeatability 

±20 ppm @400ppm 

 

Temperature Dependence 

Typ.  ±0.2%  of  reading  per  °C  or  ±2  ppm  per  °C, 
whichever is greater, referenced to 25°C 

Pressure Dependence 

0.13% of reading per mm Hg 

 

Response Time 

About 2min for 90% of step change

 

Warm-Up Time 

About 60 seconds at 22°C

 

Sound Alarm 

About 65db@10cm 

Zone LED Display   

Green:<800ppm Yellow:800~1200ppm Red: >=1200ppm 
(HVAC mode) 

Green:>1000ppm Yellow:600~1000ppm Red: <=600ppm 
(GreenHouse mode) 

Power Supply 

18~26VAC RMS 50/60Hz or 18~36VDC (half-wave 

rectified, polarity protected) 

Linear Voltage Output   

  0~10VDC (100 Ohms output impedance) 

Linear Current Output   

4~20mA (Max Load is 500 Ohm) 
( Max Load is 400 Ohm while power supply

20 VDC ) 

Relay Output 

30VDC or 250VAC, max 2A ., SPST .Normally Open   

Operating Temperature 

0°C ~50°C (32°F~122°F) 

Operating Humidity Range 

0 ~ 95% RH non-condensing 

Storage Temperature 

-20°C to +60°C (-4°F to 140°F) 

  Temperature specification:

 

Temperature Range

 

Display 0°C to 50°C 
 

Display Resolution 

0.1°C 

Display Options 

°C   

Accuracy 

Relay (no action and be under 
the alarm level)

 

±1°C when the fan blows to the device directly, the accuracy 
of temperature is +/- 1.5 °C   

Accuracy 

Relay (action and exceeds the 
alarm level)

 

±2.5°C when the fan blows to the device directly, the accuracy 
of temperature is +/- 1.5 °C 

Response Time 

20-30 minutes (case must equalize with environment) 

  RH Specification:

 

Measurement Range

 

20%-90% RH

 

Display Resolution

 

1%RH 

Accuracy 

±5%RH@23°C 

Response Time 

About 5 min for 63% of step change 

  Operating Conditions:

 

Operating Temperature 

0°C to 50°C   

Humidity Range 

0 ~ 95% RH non-condensing 

 

Dimension 

 

 

 

Calibration   

Notice: Before calibrating, you need standard gas or semi-standard gas; there are 3 methods to get 
standard gas. 

  Method A: use CO2 in office/building 

-Use two Meters (One is the device for calibration. The other one is a calibrated (new) one. 
-Use  ambient  room  gas  for  calibration  in  office,  waiting  at  least  10min,  until  the  CO2  reading 

doesn’t change 

 (Notice: user must not breathe toward the CO2 monitor, CO2 from the user will affect the reading   

of CO2 monitor) 

-Use the reading of the new device as the standard 
-Calibrate the device by the Cali Mode instruction. 

  Method B: use CO2 outsides 

-Use  ambient  room  gas  for  calibration  outsides,  waiting  at  least  10min,  until  the  CO2  reading 

doesn’t change 

 (Notice: user must not breathe toward the CO2 monitor; CO2 from the user will affect the reading 

of ZGw19) 

-Use 380~420ppm as the standard reading. 
-Calibrate the device by the Cali Mode instruction. 

  Method C: use standard CO2 in the bottle 

-Pump  the standard  CO2  gas (0~1000ppm,  flux = 0.1~0.2 liter/min) into the CO2 monitor from 

the Gas Entry Hole waiting about 2~3min. 

- Calibrate the device by the Cali Mode instruction. 

 

Radiant Innovation Inc.    Http://www.ZyAura.com 

1F,  No.3,  Industrial  East  9

th

  Road,  Science-Based  Industrial 

Park, HsinChu, Taiwan 300.           

 

   mm

[inches]