Page 12
Instruction manual 50192010
Version 1 June-00
7. MAINTENANCE
Every indication mentioned before will lack of value if you do not realize in continuity
all the work of maintenance.
The manteinence necessary for these microscopies is the same than for other optic instru-
ments. They should always be covered to avoid dust.
The eyepieces should not be dissasembled by the user, if the is any dust in the extern surfa-
ce of the lence clean them with a soft material that does not leave fluff.
To eliminate the dust over the lence apply wind/blow with a special object or clean with a
soft brush.
In the cleaning the mechanical parts use non corrosive lubricants, this way you will not
damage the paint or the endings. Please have special caution in not touching the optical parts.
Allways use a neutral detergent.
If the lamp blows you should chainge it for an alojen lamp ZUZI of 12V 100W, having spe-
cial caution in not toching it.
IMPORTANT:
Before changing the lamp please make sure you disconect the microscope of
the electric net. Make sure you use the original ZUZI halogen lamps of 12V 100W, the usage of
other tipe of lamps can cause a bad operating of the instrument.
For the changing of the lamp, with the entrance of the light (19) we will find two black but-
tons that you have to press and pull out turning a bit the lamp, to substitute it wait until the
lamp is cold, once it is cold, move the spring that is holding it and take it out, pulling it take of
the holders of the light bulb and substitute it.
For the changing of the fuse unscrew the boton "Fuse" (11) and substitute the interior fuse.
8. TROUBLESHOOTING
Please check the following points before sending the colposcope to the technical service:
If the lamp does not go on:
Chek that there is a good conexion to an appropiate current net.
Chek that the fuse and lamp are in a good state.
If there is dust or dirt in the field of view:
Check the dust on the objective, superior lence of the eyepiece. Once you have localized
the zone that causes the problem clean it as mentioned before.