This cleaning protocol* guarantees an efficient, hygienic cleaning process, free from
bacteria, even when the equipment is very dirty.
By means of this process, you will also contribute to the proper preventive maintenance
of your Speed Up.
Below are the steps to follow to fulfil the protocol with success.
CLEANING PROTOCOL
①
Tools required for effective cleaning:
·
Zumex Citric Active
TM
detergent
·
Disinfectant: Hydroalcoholic at a concentration of 0.91g/ml
·
Clean or new sponge (of a material that does not scratch the machine surfaces)
·
Clean disposable paper
·
Wash point
·
Brush supplied (for the cleaning points which require it)
NOTE: The brush must be cleaned and sanitised by following the instructions below.
*Speed Up (v.00) cleaning instruction validation report
performed by an independent laboratory with report number
IFSC243 / 19Rev3.
②
Steps required for effective cleaning:
In the
juice circulation zone
, follow the specific steps to guarantee maximum hygiene::
1. Apply
Zumex Citric Active
TM
to remove wax and pulp. Leave to act 30 seconds and remove the dirt with a sponge.
2. Rinse with water.
3. Apply a hydroalcoholic disinfectant with a concentration of 0.91g/ml and leave to rest for 1 minutes.
4. Rinse with water.
5. Final drying of each of the parts and recesses.