(7) Measure Capacitance 

1. Connect the black test lead to the COM Terminal and connect the red test lead to the 

     VΩHz Terminal;

2. Turn the rotary switch to the Resistance Mode, press SELECT three times to toggle to 

     the Capacitance Mode;

3. Connect the red probe to the anode side and the black probe to the cathode side of 

     the capacitor being tested;

4. Read the measured capacitance value on the display once the reading is stablized.

*Caution:

a. Disconnect circuit power and discharge all capacitors before you test capacitance.

(8) Measure Frequency and Duty Cycle

1. Connect the black test lead to the COM Terminal and connect the red test lead to 

     the VΩHz Terminal;

2. Turn the rotary switch to the Frequency Mode; press SELECT once to toggle to the 

     Duty Cycle Mode if you want to measure duty cycle;

3. Touch the probes to the desired test points of the circuit;

4. Read the measured frequency/duty cycle value on the display.

*Caution:

a. The Frequency Mode only applies to measure high frequency with low voltage.

(9) Auto Power Off

1. The product  automatically powers off after 15 minutes of inactivity;

2. The built-in beeper beeps 5 times 1 minute before power off;

3. To restart the product, press SELECT botton;

4. To disable the Auto Power Off function, hold down the SELECT botton when 

     turning on the product, you will hear five beeps if you have successfully disabled 

     the function.

E. Genearl Maintenance

Beyond replacing batteries and fuses, do not attempt to repair or service the product 

unless you are qualified to do so and have the relevant calibration, performance test, and 

service instructions. 

(1) Do not operate the product around hot, wet, flammable, explosive or magnetic 

      environments.

(2) Clean the product with damp cloth and mild detergent; do not use abrasives or 

      solvents.

(3) Remove the input signals before you clean the product.

(4) Remove the batteries if you will not use the product for a long time to prevent 

      possible battery leak.

(5) When “    ” is shown on the display, batteries shall be replaced as below:

1. Loosen the screw and remove the battery cover;

2. Replace the used batteries with new batteries of the same type;

3. Place the battery cover back and fasten the screw.

(6) Replace fuses as above steps. Use only fuses of the same type as the original 

      ones.

(3) Measure AC/DC Voltage

1. Connect the black test lead to the COM Terminal and connect the red test lead to         

     the VΩHz Terminal;

2. Turn the rotary switch to the AC Voltage Mode or the DC Voltage Mode;

3. Touch the probes to the correct test points of the circuit to measure the voltage;

4. Read the measured voltage on the display.

*Caution:

a. Do not measure voltage that exceeds the MAX Value as indicated in the  

Specifications;

b. Do not touch high voltage circuit during measurements.

(4) Measure Resistance

1. Connect the black test lead to the COM Terminal and connect the red test lead 

     to the VΩHz Terminal;

2. Turn the rotary switch to the Resistance Mode, and the display will show “OL”;

3. Touch the probes to the desired test points of the circuit to measure the 

     resistance;

4. Read the measured resistance on the display.

*Caution:

a. Disconnect circuit power and discharge all capacitors before you test resistance.

b. Do not input voltage at the Resistance Mode.

(5) Measure Diode

1. Connect the black test lead to the COM Terminal and connect the red test lead to 

    the VΩHz Terminal;

2. Turn the rotary switch to the Resistance Mode, press SELECT once to toggle to 

     the Diode Mode;

3. Connect the red probe to the anode side and the black probe to the cathode side 

    of the diode being tested;

4. Read the forward bias voltage value on the display;

5. If the polarity of the test leads is reversed with diode polarity or the diode is 

     broken, the display reading shows “OL”.

*Caution:

a. Do not input voltage at the Diode Mode.

b. Disconnect circuit power and discharge all capacitors before you test diode.

(6) Measure Continuity

1. Connect the black test lead to the COM Terminal and connect the red test lead to 

    the VΩHz Terminal;

2. Turn the rotary switch to the Resistance Mode, press SELECT twice to toggle to 

     the Continuity Mode;

3. Touch the probes to the desired test points of the circuit;

4. The built-in beeper will beep when the resistance is lower than 50Ω, which 

     indicates a short circuit.

*Caution:

a. Do not input voltage at the Continuity Mode.

Warning

1. Do NOT exceed the “maximum value” indicated in the Specification;

2. Do NOT input voltage at the Current Mode, the Resistance Mode, the Diode 

     Mode, or the Continuity Mode;

3. Do NOT use the product when the batteries or the battery cover is not placed 

     properly;

4. Turn off the product and remove the test leads from the test points before 

     changing batteries or fuses.

F. Troubleshooting

If your product do not function as normal, the following steps may help you. If the 

problem still cannot be solved, please contact your dealer.

Problem

Possible Reason

Display Mulfunction

Low battery; replace batteries

      Symbol

Replace batteries

No current input

Replace fuse

LIMITED WARRANTY 

AND LIMITATION OF LIABILITY

 

Customers enjoy one-year warranty from the date of purchase. This warranty 

does not cover fuses, disposable batteries, or damage from accident, neglect, 

misuse, alternation, contamination, or abnormal conditions of operation or 

handling. 