Chapter 20 See-Thru CPR
20-2
www.zoll.com
9650-001355-01 Rev. M
See-Thru CPR enables the rescuer to see a close approximation of the patient’s underlying ECG
rhythm while performing CPR. See-Thru CPR is available if the X Series is monitoring CPR.
Chest compressions introduce
CPR artifact
into the ECG signal. See-Thru CPR uses a filter
that relies on the correlation between CPR compressions, as detected by the ZOLL Onestep,
OneStep Complete
or CPR-D-padz, and the CPR artifact to remove much, but not all, of the
artifact from the ECG signal. Under some conditions, residual noise after filtering can obscure
the ECG rhythm, requiring the rescuer to stop CPR to assess the ECG. For example, in the case
of asystole or low amplitude PEA, the residual artifact seen after filtering may look like fine
ventricular fibrillation.
Because the filtered ECG signal may contain residual chest compression and/or filtering
artifacts,
a rescuer should always follow the standard procedure of stopping CPR to assess the
patient’s ECG rhythm before determining treatment
.
Using See-Thru CPR
To use See-Thru CPR
•
The X Series unit must be monitoring CPR.
•
ZOLL OneStep CPR pads, OneStep Complete pads, or CPR-D-padz must be attached to the
unit.
When chest compressions begin, the X Series unit
automatically
starts filtering the CPR artifact
after detecting the first 3 to 6 compressions. The filtered ECG, with the label “FIL,” may be
displayed on the second or third trace (by selecting
FILT ECG
in the Trace2 or Trace3 menu).
See-Thru CPR filtering continues as long as the OneStep CPR, OneStep Complete or
CPR-D-padz detect compressions and patient impedance is valid. When no compressions are
detected or one of the conditions noted above occurs, See-Thru CPR filtering stops, and
unfiltered ECG signals are displayed. When compressions resume, filtering automatically
restarts after 3 to 6 chest compressions.
Note:
There is a delay of approximately 1/16 second between the See-Thru CPR waveform
and the Trace 1 ECG waveform.
Summary of Contents for X Series
Page 1: ...Includes Real CPR Help and See Thru CPR X Series Operator s Guide 9650 001355 01 Rev M...
Page 64: ...Chapter 2 Product Overview 2 24 www zoll com 9650 001355 01 Rev M...
Page 74: ...Chapter 3 Monitoring Overview 3 10 www zoll com 9650 001355 01 Rev M...
Page 80: ...Chapter 4 Trends 4 6 www zoll com 9650 001355 01 Rev M...
Page 90: ...Chapter 5 Alarms 5 10 www zoll com 9650 001355 01 Rev M...
Page 110: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 001355 01 Rev M...
Page 126: ...Chapter 8 Monitoring Non Invasive Blood Pressure NIBP 8 16 www zoll com 9650 001355 01 Rev M...
Page 156: ...Chapter 10 Pulse CO Oximetry SpO2 10 18 www zoll com 9650 001355 01 Rev M...
Page 166: ...Chapter 11 Monitoring Invasive Pressures IBP 11 10 www zoll com 9650 001355 01 Rev M...
Page 198: ...Chapter 14 12 Lead ECG Interpretive Analysis 14 16 www zoll com 9650 001355 01 Rev M...
Page 218: ...Chapter 16 Advisory Defibrillation 16 6 www zoll com 9650 001355 01 Rev M...
Page 224: ...Chapter 17 Analysis CPR Protocol Defibrillation 17 6 www zoll com 9650 001355 01 Rev M...
Page 234: ...Chapter 18 External Pacing 18 10 www zoll com 9650 001355 01 Rev M...
Page 242: ...Chapter 19 Real CPR Help 19 8 www zoll com 9650 001355 01 Rev M...
Page 254: ...Chapter 21 Patient Data 21 6 www zoll com 9650 001355 01 Rev M...
Page 338: ...Specifications A 38 www zoll com 9650 001355 01 Rev M...