Maintenance Tests
9650-0807-01 Rev. C
ZOLL Propaq MD Service Manual
31
15.0 SpO
2
, SpCO, and SpMet Test
(for SpO
2
feature with SpCO, or SpCO and SpMet)
Note:
Use this test if your unit has SpO
2
and SpCO, or SpO
2
, SpCO and SpMet. If you do not
have the SpMet feature on the unit, disregard the SpMet values in the test below.
Equipment
Fluke Index 2 SpO
2
Simulator (or equivalent), Masimo Rainbow 3 ft BioTek Test Sensor
(Masimo P/N 3379) or Masimo Rainbow 3 ft Direct Connect Optical Test Sensor
Test Setup
Disconnect any ECG cable or IBP probes from the
Propaq
unit.
Do This...
Observe This...
Pass/Fail/N/A
15.1 Power on the Propaq unit without the
SpO
2
sensor attached.
Verify that the
SpO
2
numeric
displays
INITIALIZING
, then
goes blank.
15.2 Connect the
SpO
2
sensor to the Propaq
unit.
Verify that
SpO
2
numeric
displays
SEARCHING
.
15.3 Set the simulator to 75%
SpO
2
and 50
BPM. Attach the
SpO
2
sensor to the
simulator.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 72.1% and
78%.
The
SpCO
numeric displays an
SpCO
value between 13.5 and
21.5.
The
SpMet
numeric displays an
SpMet
value between 5.5 and
9.5.
Verify that the displayed pulse
rate reads between 44.5 and 55.5
BPM.
o
o
o
o
o
o
o
o
o
15.4 Set the simulator to 75%
SpO
2
and 230
BPM.
Verify that after a brief period of
initialization,
The
SpO
2
numeric displays an
SpO
2
value between 72.1% and
78%.
The
SpCO
numeric displays an
SpCO
value between 12.8 and
20.8.
The
SpMet
numeric displays an
SpMet
value between 5.2 and
9.2.
Verify that the displayed pulse
rate reads between 222.7 and 236
BPM.
o
o
o
o
o
o
o
o
o
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Service Manual 9650 0807 01 Rev C...
Page 6: ...TABLE OF CONTENTS iv www zoll com 9650 0807 01 Rev C...
Page 64: ...58 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 102: ...96 www zoll com 9650 0807 01 Rev C This page intentionally left blank...
Page 108: ...CHAPTER 4 REPLACEMENT PARTS 102 9650 0807 01 Rev C Pictures and Diagrams Picture 1 Picture 2...
Page 109: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 103 Picture 3 Picture 4...
Page 110: ...CHAPTER 4 REPLACEMENT PARTS 104 9650 0807 01 Rev C Picture 5 Picture 6...
Page 111: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 105 Picture 7 Picture 8...
Page 112: ...CHAPTER 4 REPLACEMENT PARTS 106 9650 0807 01 Rev C Picture 9 Picture 10...
Page 113: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 107 Picture 11 Picture 12...
Page 114: ...CHAPTER 4 REPLACEMENT PARTS 108 9650 0807 01 Rev C Picture 13 Picture 14...
Page 115: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 109 Picture 15 Picture 16...
Page 116: ...CHAPTER 4 REPLACEMENT PARTS 110 9650 0807 01 Rev C Picture 17 Picture 18...
Page 117: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 111 Picture 19 Picture 20...
Page 118: ...CHAPTER 4 REPLACEMENT PARTS 112 9650 0807 01 Rev C Picture 21 Picture 22...
Page 119: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 113 Picture 23 Picture 24...
Page 120: ...CHAPTER 4 REPLACEMENT PARTS 114 9650 0807 01 Rev C Picture 25 Picture 26...
Page 121: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 115 Picture 27 Picture 28...
Page 122: ...CHAPTER 4 REPLACEMENT PARTS 116 9650 0807 01 Rev C Picture 29 Picture 30...
Page 123: ...Overview 9650 0807 01 Rev C Propaq MD Service Manual 117 Picture 31 Picture 32...
Page 124: ...9650 0807 01 Rev C Propaq MD Service Manual 118...
Page 134: ...128 9650 0807 01 Rev C This page intentionally left blank...
Page 138: ...132 9650 0807 01 Rev C This page intentionally left blank...
Page 142: ...This page intentionally left blank...