ZOLL M Series Operation Manual Download Page 3

ZOLL M Series XL Battery

1

Product Description

The ZOLL XL Battery is a 4.5 AH lead acid battery pack intended for use in appropriately 
configured M Series and M Series CCT defibrillators. It is available as a Smart Battery or Smart 
Ready Battery. Please see the 

Smart Battery Operator’s Guide

 for additional information on Smart 

Battery functionality. 

NOTE:

The XL battery is NOT intended for use in ZOLL 900/1200/1400/1600/1700/2000 series 
units. See the "Defibrillator Compatibility"  section below for information on M Series use.

Warnings!

• Follow all recommended maintenance and operating instructions for the XL Battery and associated 

ZOLL charging/testing systems. If a problem or question arises, contact Technical Service. Do not 

use the XL Battery until it has been inspected, tested, and recalibrated by the appropriate personnel 
and you have been adequately trained in the proper use of the XL Battery.

• Locate the XL Battery and ZOLL resuscitation devices where they cannot harm the patient or user 

if they fall.

• Under certain conditions, it is possible for small amounts of hydrogen gas to be released during 

charging of the battery pack. This may present a hazard.

• Do not immerse or set the XL Battery in fluid. 
• Do not use alcohol or ketones (MEK, acetone, etc.) on the XL Battery. Do not autoclave the XL 

Battery.

• The XL Battery is protected against interference from radio frequency emissions typical of two way 

radios and cellular phones (digital and analog) used in emergency service/public safety activities. 

Users of the XL Battery should assess the device's performance in their typical environment of use 
for the possibility of radio frequency interference from high power sources.

• Keep a fully charged spare battery pack available at all times with the Monitor/Defibrillator. 

Replace the battery immediately upon low battery message.

• Do not install batteries into Monitor/Defibrillators when storage may exceed 90 days. Battery 

damage may occur.

• Do not use the XL Battery if the case is cracked or broken.
• When charging a XL Battery pack in a Base PowerCharger

4x4

 or Base PowerCharger

1x1

 unit, ensure 

unit is labeled “XL Battery Ready” otherwise the battery cannot be charged to full capacity and 

operating life may be reduced.

• ZOLL charging/testing systems used with the XL Battery should be calibrated or tested 

periodically using the appropriate Service Manual Test Check Out procedures.

• To dispose of the battery properly, place it in an acid resistant container. Dispose of the battery in 

accordance with federal, state, and local regulations. Do not incinerate batteries. Ship the batteries 

to a reclaimation facility for the recovery of metal and plastic compounds as the proper method of 
waste management. 

General Information

Defibrillator Compatibility

M Series units must have software version 30.00 or higher and be outfitted with the appropriate 
handle assembly. New M Series units may be ordered with the XL Battery option. Older M Series 
units may be upgraded if the current software version is 20.00 or higher. All M Series CCT units are 
XL Battery-ready.
The XL Battery is not intended for use in ZOLL 900/1200/1400/1600/1700/2000 series units. 

Reviews: