background image

highlight the month field. Press the Inc Value or Dec value soft key to select the appropriate month. 
Pressing the Next-Field soft key will set the selected month and move the cursor to the next field (day).  
Repeat above steps to set the correct day, year, hours and minutes field.  Finally, <Enter and Return>. 

 
5.10.

 

SP02  Check  (Massimo) 

When a good sensor is connected, check that its internal red LED is on.  After approximately 20 seconds 
using a healthy subject, the small template on the upper part of the display will fill with a value between 
91 and 99.  No waveform is displayed.

 

 

5.11.

 

NIBP Check (if present) 

Select the appropriately sized cuff and connect to a simulator.  Press the <NIBP> hard key to begin 
measurement.  Note the screen will display the real-time cuff pressure, and eventually the final 
Systolic/Diastolic reading.  As a pressure gage, the service manual specs that the unit will measure 
pr/- 5mmHg compared to a standard.  See Sections 23 thru 24 of the E Series Service Manual for 
adjustment information.  Similar pressure/service info not found in M documentation. 

 

5.12.

 

CO2   Not addressed. 

 

APPENDIX 

 

PATIENT LEAKAGE TEST / LINE LEAKAGE TEST  

Refer to the leakage tester’s manual for specific directions.  For the M Series, see p. 15 of the service manual.   For the E 
Series, see p 1-14.  The limits are the same for either model.  Should be done if unit was dropped/damaged or incurred 
moisture. 

 
CONFIGURATION SETUP  

Turn the selector switch to OFF, and wait 10 seconds. Simultaneously press and hold the right-most and left-most soft 
keys (below the monitor) while turning the selector switch to MONITOR (ON for AED units), and hold the soft keys for 
approximately 4 seconds.  The unit powers on in System Configuration mode. 
 
Enter the passcode (0 0 0 0 0 0 0 0) using the left “0” soft key.  Note: When you enter System Configuration for the first 
time, enter the default pass code of 00000000.  To change a digit, use the Delete soft key to highlight the entry and re-
enter the digit.  When all 8 digits have been entered, the unit automatically advances to the Select Language screen.  
After selecting your language, hit <Restore Defaults><Enter>.  The next screen indicates “Defaults Restored”.  Hit <Exit 
Config>.   Then select “Save Changes and Exit”, then <Enter>.  Finally “Save Changes To Card?” <Next Item>, 
<Enter>. 

 

QRS VOLUME

 (M model; E almost identical)  

Change QRS volume by hitting the <Speaker> button soft key in 

the lower left corner, and follow the prompts. Note:  If no QRS volume despite increasing the volume bar to maximum, 
remedy by reconfiguring the entire unit to “Factory Default”.  Or enter Config Mode, <English> <Enter> <Change 
Config> <Tones> <Enter> <QRS Volume on Power Up> <Change Value> <on> <Return> <Save Changes and Exit> 
<Enter>.   Reminder: turn off ECG simulator to exit configuration. 

 