If the console detects that it is more than 3 minutes since the last time it was powered down, it will con-
duct all its self tests. You will not be offered the option of operating the console at the current settings.
Downloading Data After Improper Shutdown
To help prevent the accidental loss of patient trend data, the console is designed to store patient trend data if
power to the console is interrupted or lost. If the console has not been properly shut-down, all patient trend
data will be preserved. The next time the console is switched on, the self-test program detects the saved
patient data and offers you the option to download the data to a laptop computer. If you do not wish to
download the saved data, the console will delete the data before permitting set up for a new patient.
To download patient trend data after an improper shutdown, follow these steps in the indicated order.
1. Plug in and switch on the console. The self-test screen appears on the display screen.
Note.
The Software Version shown in Figure 4.53 is for reference only; the current product may use a different
Software Version.
Figure 4.53. Self-Test Screen
2. When the self-test is finished, the System Set Up screen displays the message “Bath Pre-Set?”