Figure 4.20. Alarm Screen
In some cases, however, the alarm may be cleared by power cycling the console. For example, such an alarm
would occur if the pump tubing were to become jammed in the peristaltic pump causing the pump to slow
down.
If the reason for the alarm is not cleared by power cycling, the alarm will repeat each time the console is
turned back on.
During an alarm, the patient temperature display and the patient temperature alarms are not active.
Note.
In the event of an alarm
: Investigate and troubleshoot the cause - refer to: Alarms and Corrective
Actions on page 85 and Troubleshooting on page 89. If the alarm persists, call ZOLL for service.
Your First Case
What you need
You will need the following for each case:
•
A new Start-Up Kit
•
A new 500ml bag of normal saline.
•
A YSI-400 compatible temperature probe e.g. a Foley catheter, rectal or esophageal temperature probe
•
The blue patient connection cable to connect the temperature probe to the console
•
A ZOLL heat exchange catheter
•
An aseptic work area to support catheter insertion.
WARNING.
Refer to catheter IFU
.
ZOLL heat exchange catheters are inserted via a Seldinger technique sim-
ilar to a central venous line. There are specific instructions for use included with each catheter. Refer to these
to understand the specific unique insertion requirements of a ZOLL heat exchange catheter.