35
Nature of the Alarm
Figure 4.11. “Hi” Patient Temperature Alarm
The alarms are both visual and audible. The alarms will not clear until the patient’s temperature no longer trig-
gers the alarm state.
The audible alarm may be temporarily muted for 2 minutes by pressing the Mute button. The alarm will con-
tinue after that time unless it has cleared.
The visual alarm displays the patient temperature in flashing red text. The visual alarm will not stop until the
alarm has cleared.
WARNING. Setting of patient temperature alarms.
The patient temperature alarms are programmable.
They cannot be deactivated. Press the Mute button to mute the alarm for 2 minutes.
Note.
Patient temperature alarms in
Standby.
If the console is in Standby and the T1 temperature probe is
connected to the console but not inserted into the patient, a “Lo” patient temperature alarm can be triggered
if the ambient temperature of the exposed probe is below the alarm limit. To avoid the alarm under these con-
ditions, unplug the T1 temperature probe from the console until you are ready to connect to the patient. You
may not enter Run without the probe connected to the console.
Bath Pre-set
The menu allows you to select from either: Pre-Cool, Pre-Warm or None. This selection operates only in
Standby and is canceled when the console enters Run.