ZOLL M Series Invasive Blood Pressure
3
4. Verify that neither the CHECK TRANSDUCER nor the REPLACE
TRANSDUCER messages display.
If desired, repeat steps 1 - 4 to connect a second transducer. The ZERO P1 or ZERO
P2 message displays. No pressure measurements will appear until the transducer
has been zeroed.
Setting up the transducer and catheter
Follow manufacturer’s directions for use and your hospital’s protocols on how to setup, prime and
flush the transducer and make connections to the pressure monitoring catheter.
Displaying the IBP Menu
Before you can zero the transducer, you must first display the IBP menu. You can display the IBP menu
without leaving Monitor, Defib or Pacer mode.
AEDs must be in manual mode to display the IBP menu. See the
M Series Operator’s Guide
for further
details on activating the manual override mode.
To display the IBP menu:
1.
Press the
Param
softkey. If the Param softkey label is not displayed, press the
Return
softkey
until “Param” displays.
The Parameter menu displays (see Fig. 2):
Figure 2: Parameter Menu
2.
Press the
Select
softkey until “IBP” highlights.
3.
Press the
Enter
softkey.
M Series displays the IBP menu (see Fig. 3):
Figure 3: IBP Menu
You can now zero the transducer(s).
NOTE
Unless otherwise noted, all the following procedures in this insert assume that you are already in the
IBP menu.