Circuits:
[Infant for 5-30kg, max 300ml TV/Adult for >20kg, min 200ml TV]
Operating Sequence:
1. Hook up all connections: Green hose to green transducer port, clear hose to silver exhalation valve,
circuit to gas outlet, oxygen if FiO2 > 21% desired.
2. Turn on ventilator using power switch. Vent will start with default settings after a self check.
3. Use default settings for initial settings: TV 500; FiO2 21; Rate 12; I:E 1:2.5; PEEP 5; Assist Control
4. High pressure alarm is pre-set at 35cmH2O and pressure will pop off at set value
5. Adjust FiO2 upward to maintain SpO2 > 90%
6. For pediatric patient, start with a TV of 8ml/kg
Changing Parameters from Start-up Defaults: