ZOLL IMPACT 731 Quick Start Manual Download Page 2

Circuits:               

[Infant for 5-30kg, max 300ml TV/Adult for >20kg, min 200ml TV]

 

 

Operating Sequence: 

1. Hook up all connections: Green hose to green transducer port, clear hose to silver exhalation valve, 

circuit to gas outlet, oxygen if FiO2 > 21% desired. 

2. Turn on ventilator using power switch. Vent will start with default settings after a self check. 
3. Use default settings for initial settings: TV 500; FiO2 21; Rate 12; I:E 1:2.5; PEEP 5; Assist Control 
4. High pressure alarm is pre-set at 35cmH2O and pressure will pop off at set value  
5. Adjust FiO2 upward to maintain SpO2 > 90% 
6. For pediatric patient, start with a TV of 8ml/kg 

 

Changing Parameters from Start-up Defaults: 

 