background image

ICY

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

Model IC-3893A/8700-0782-01 

 

4 of 10 

 

11.

 

Using centimeter marks on the catheter as positioning 
reference points, advance catheter to at least the 18 cm mark, 
to ensure the proximal infusion port is in the vessel. 

12.

 

Hold the catheter at the desired depth and remove the 
guidewire. If resistance is encountered when attempting to 
remove the guidewire after catheter placement, the guidewire 
may be kinked at the tip of the catheter. If resistance is 
encountered, withdraw the catheter relative to the guidewire 
about 2-3 cm and attempt to remove the guidewire. If 
resistance is encountered again, remove the guidewire and 
catheter simultaneously. 

13.

 

Caution: Do not apply undue force to the guidewire.

 

14.

 

Verify that the guidewire is intact upon removal. 

15.

 

Check catheter placement by attaching a syringe to the distal 
infusion Luer and aspirate until a free flow of venous blood is 
observed. Connect the infusion Luer to the appropriate Luer-
Lock line as required. The unused infusion port may be 
"locked" through the injection cap using standard hospital 
protocol. A slide clamp is provided on the tubing to occlude 
flow through the infusion Luer during line and injection cap 
changes. 

Caution: To minimize risk of damage to the 

tubing from excessive pressure, the clamp must be opened 
prior to infusing through the Luer. 

16.

 

Caution: Do not clamp or occlude IN or OUT lines. This 
can cause line blockage and possible failure. 

17.

 

Secure and dress the insertion site and catheter temporarily. 

18.

 

Verify the catheter tip position by chest x-ray immediately 
after placement. The x-ray exam must show the catheter 
located in the IVC with the distal end of the catheter parallel 
to the vena cava wall. If the catheter tip is malpositioned, 
reposition and re-verify. 

19.

 

The proximal radiopaque marker indicates the proximal end 
of the balloon. Ensure that the balloon and proximal port 
reside completely in the vessel. If the catheter is 
malpositioned, reposition and re-verify. 

20.

 

Secure the catheter to the patient. Use the juncture Luer side 
wings as the primary suture site to minimize the risk of 
catheter migration. 

21.

 

The ZOLL suture tab and clip can also be used as an 
additional attachment point. Ensure that catheter body is 
secure and does not slide. 

22.

 

Caution: Use only the ZOLL suture tab and clip provided 
in the kit. Catheter damage may result if other tabs or 
clips are used. 

23.

 

Caution: Do not suture directly to the outside diameter of 
the catheter, to minimize the risk of cutting or damaging 
the catheter or impeding catheter flow. 

24.

 

Dress the puncture site per hospital protocol. Maintain the 
insertion site with regular meticulous redressing using aseptic 
technique. 

25.

 

Record on the patient's chart the indwelling catheter length 
using the centimeter marks on the catheter shaft as reference. 
Frequent visual reassessment should be made to ensure that 
the catheter has not moved. 

26.

 

Attach a primed Start-Up Kit to the catheter by connecting 
the male Luer of the Start-Up Kit to the female IN Luer of 
the catheter and the female Luer of the Start-Up Kit to the 
male OUT Luer of the the catheter. White "ZOLL" tags are 

fitted loosely to the IN and OUT extension tubes to help 
identify them. Ensure that a sufficient amount of sterile saline 
is present at the ends of the Luers to make an air-free 
connection. Refer to the operation manual. 

27.

 

The Start-Up Kit IN and OUT Luers are only intended to 
connect to the catheter IN and OUT Luers and are not 
intended to connect to standard Luer Lock syringes.  They 
have ZOLL custom fittings and are orange in color for easy 
identification. 

28.

 

WARNING: Failure to connect the Start-Up Kit correctly 
to the catheter could result in catheter failure. Do not 
attach the Start-Up Kit (orange) Luers to the brown, 
white or blue infusion Luers. 

29.

 

Caution: Do not place any stopcocks in line that may be 
inadvertently shut off. This can cause line blockage and 
possible failure. 

30.

 

Pump saline through the Start-Up Kit and catheter to ensure 
that all connections are secure and that there is no leaking. 
Allow any remaining air in the system to be purged out as 
described in the system operation manual. 

Disconnecting the Catheter from the System

 

1.

 

Stop circulating saline through the catheter. 

2.

 

Disconnect the Start-Up Kit from the catheter.  

3.

 

To maintain sterile connections, immediately cap off the Luer 
connectors of both the catheter and Start-Up Kit using sterile 
Luer caps or connect the IN and OUT Luers together.  

Reconnecting the Catheter to the System 

1.

 

Remove the Luer caps from Luer connectors of the catheter 
and Start-Up Kit. Discard the Luer caps or disconnect the IN 
and OUT Luers from each other.  

2.

 

Attach the Start-Up Kit to the catheter by connecting the 
male Luer of the Start-Up Kit to the female IN Luer of the 
catheter and the female Luer of the Start-Up Kit to the male 
OUT Luer of the catheter. The Start-Up Kit and catheter IN 
and OUT Luers are orange in color. Ensure that a sufficient 
amount of sterile saline is present at the ends of the Luers to 
make an air-free connection. 

3.

 

WARNING:  Failure to connect the Start-Up Kit 
correctly to the catheter could result in catheter failure. 

4.

 

WARNING:  DO NOT use the IN and OUT Luer fittings 
for standard central line infusion ports. They are for 
connection to the Coolgard 3000/Thermogard XP System 
ONLY. 

5.

 

The Start-Up Kit IN and OUT Luers are only intended to 
connect to the catheter IN and OUT Luers and are not 
intended to connect to standard Luer Lock syringes.  They 
have ZOLL custom fittings and are orange in color for easy 
identification. 

6.

 

Caution:  Do not place any extra stopcocks in line that 
may be inadvertently shut off. This can cause line 
blockage and possible failure. 

Summary of Contents for ICY IC-3893A

Page 1: ...t the sterility barrier has not been compromised Storage Store between 20 C 25 C Avoid freezing and excessive heat above 40 C Indications for Use The ZOLL ICY Intravascular Heat Exchange Catheter Model IC 3893A connected to the ZOLL Coolgard 3000 Thermogard XP Thermal Regulation System is indicated for use In cardiac surgery adult patients to achieve and or maintain normothermia during surgery and...

Page 2: ...ive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for the resistance 10 Passage of the guidewire into the right heart can cause dysrhythmias right bundle branch block vessel wall atrial or ventricular perforation 11 Use only sterile normal saline for catheter priming It is the circulating fluid in the catheter 12 The cath...

Page 3: ... connect to standard Luer Lock syringes or other standard Luer Lock connectors 4 Remove the caps from the IN and OUT Luers With the catheter cover in place fill the syringe 5 cc or larger with sterile saline and attach the syringe to the female IN Luer 5 WARNING Never inject positive pressure into the IN Luer with the OUT Luer cap in place 6 Gently inject saline through the catheter until it begin...

Page 4: ...indwelling catheter length using the centimeter marks on the catheter shaft as reference Frequent visual reassessment should be made to ensure that the catheter has not moved 26 Attach a primed Start Up Kit to the catheter by connecting the male Luer of the Start Up Kit to the female IN Luer of the catheter and the female Luer of the Start Up Kit to the male OUT Luer of the the catheter White ZOLL...

Page 5: ...t move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again If resistance is still encountered an x ray should be performed to identify the reason for the resistance MRI Safety Information Non clinical testing has demonstrated that the ICY catheter is ...

Page 6: ...e guidewire at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during in...

Page 7: ...xchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 7 of 10 1 OUT 2 IN 3 Medial 4 Distal 5 Proximal 6 Detail view ICY IC 3893A Catheter Shaft 7 Balloon 5 mm x 80 5 mm 8 Balloon 8 mm x 80 5 mm Balloons ...

Page 8: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 8 of 10 ...

Page 9: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 9 of 10 ...

Page 10: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 10 of 10 ICY SilvaSorb ...

Reviews: