background image

ICY

® 

Intravascular Heat Exchange Catheter (Custom Luer) 

Instructions for Use 

Model IC-3893A/8700-0782-01 

 

106477-001 

Revision 6 

1 of 10 

 

ZOLL 

Caution: Federal law restricts this device to sale by or on the 
order of a physician. 

 

ICY

®

 Intravascular Heat Exchange Catheter (ICY Catheter) 

Kit 

includes: 

Quantity 

English 

ICY

 

Catheter 3-Luer 

9.3 French x 38 cm  
Triple Infusion Luer  
Extension Line Clamps  
Radiopaque Shaft  
Applause Heparin Coating 

Guidewires .032” (0.81 mm) x 75 cm 

Vessel Dilator 10.5 F x 0.38” 
(3.6 mm x 1.0 mm) 

Detachable Suture Tab & Clip 

18 ga x 2 ½" (1.3 mm x 6.3 cm) Radiopaque OTN 
Catheter 

000 Silk Suture 

Chloraprep

®

  Triple Swabstick Prep Pack 

4" x 4" (10 cm x 10 cm) Gauze Sponges 

No. 11 Surgical Blade w/ long handle 

3 cc Syringe & 25 ga x 1" (0.5 mm x 2.5 cm ) 
Needle 

5 cc Syringes & 22 ga x 1

½

“ (0.7 mm x 3.8 cm) 

Needle 

Fenestrated Drape 

18 ga x 2 ¾" (1.3 mm x 67 mm) Needle 

Needle Disposal Cup 

SilvaSorb

®

  Site Antimicrobial Dressing 

SureSite Transparent Film Dressing 

Device Description 

The ICY Intravascular Heat Exchange Catheter (“ICY Catheter” or 
“catheter”) is a sterile, single-use, flexible catheter designed for 
placement in the inferior vena cava from an insertion site in the 
femoral vein. The ICY Catheter is to be connected to a ZOLL 
single use disposable Coolgard 3000

®

 /Thermogard XP

®

 Start-Up 

Kit and the Coolgard 3000/ Thermogard XP System (both supplied 
separately). An YSI-400 Temperature probe is required for the 
operation of the Coolgard 3000 or Thermogard XP. A dilator and 
guidewire are required for the percutaneous insertion of the ICY 
Catheter. Three (3) Luers are available for infusion and sampling.  

 

Infusion Port 

Flow Rate 

Priming 
Volume 

Guidewire Port (brown) 

1700 ml/hr 

0.5 cc 

Medial Port (white) 

900 ml/hr 

0.4 cc 

Proximal Port (blue) 

1200 ml/hr 

0.4 cc 

 

The ICY

 

catheter blood contact surfaces (tip, balloon, and shaft) are 

treated with an anti-thrombotic Applause heparin coating.  

Sterility 

Ethylene oxide sterilized. The ICY

 

Catheter is supplied sterile for 

single use only and should not be resterilized. The package should 
be inspected prior to use to ensure that the sterility barrier has not 
been compromised. 

Storage 

Store between 20°C-25°C. Avoid freezing and excessive heat 
above 40°C. 

Indic

a

tions for Use 

The ZOLL

®

 ICY Intravascular Heat Exchange Catheter Model IC-

3893A, connected to the ZOLL Coolgard 3000/Thermogard XP 
Thermal Regulation System, is indicated for use: 

 

In cardiac surgery adult patients to achieve and/or maintain 
normothermia during surgery and recovery/intensive care, 
and  

 

To induce, maintain and reverse mild hypothermia in 
neurosurgery adult patients in surgery and recovery/intensive 
care.   

Contraindications 

1.

 

The risks of the catheter are essentially those of a central line. 
The catheter should not be used in patients for whom central 
line placement is not indicated. 

2.

 

Bleeding diathesis. 

3.

 

Active sepsis. 

4.

 

Infection or active bleeding at the site of catheter insertion. 

5.

 

Patients with no vascular access or a vascular system that 
will not accommodate a catheter, including patients with 
vena cava filters or other implanted impediments to passage 
of the catheter. 

6.

 

Patients for whom the required temperature monitoring 
cannot be established. 

7.

 

Hypothermia is contraindicated in patients who have 
hematological diseases that will be made worse with 
hypothermia e.g. any disease that produces cryoglobulinemia, 
any hemoglobinopathy in which hemolytic anaemia can be 
precipitated by cold including Sickle Cell Disease or 
Thalassemia. 

8.

 

Not intended for pediatric or neonatal use. 

Warnings and Precautions 

Central venous catheterization should only be performed by well-
trained personnel well versed in anatomical landmarks and safe 
technique. Personnel should also have knowledge of potential 
complications.  

SINGLE USE ONLY.

 The product is designed for single use 

only. Do not resterilize or reuse. Do not reinsert, once removed 
from the patient. Do not alter the catheter in any way. 

Maximum 

use period: 4 days.

 

Potential risks with reuse of a single-use device include but are not 
limited to: 

 

Potentially life threatening infection 

 

Toxic shock due to degradation of materials 

 

Increased risk of thrombosis 

 

Reduced heat exchange power 

 

Device failures 

WARNING: The ICY Catheter and Start-Up Kit could 
potentially misconnect with other devices with small bore 
connectors.   Such connection errors could result in patient 
injury or death. 

 

Summary of Contents for ICY IC-3893A

Page 1: ...t the sterility barrier has not been compromised Storage Store between 20 C 25 C Avoid freezing and excessive heat above 40 C Indications for Use The ZOLL ICY Intravascular Heat Exchange Catheter Model IC 3893A connected to the ZOLL Coolgard 3000 Thermogard XP Thermal Regulation System is indicated for use In cardiac surgery adult patients to achieve and or maintain normothermia during surgery and...

Page 2: ...ive force in moving the catheter or guidewire If resistance is encountered an x ray should be performed to identify the reason for the resistance 10 Passage of the guidewire into the right heart can cause dysrhythmias right bundle branch block vessel wall atrial or ventricular perforation 11 Use only sterile normal saline for catheter priming It is the circulating fluid in the catheter 12 The cath...

Page 3: ... connect to standard Luer Lock syringes or other standard Luer Lock connectors 4 Remove the caps from the IN and OUT Luers With the catheter cover in place fill the syringe 5 cc or larger with sterile saline and attach the syringe to the female IN Luer 5 WARNING Never inject positive pressure into the IN Luer with the OUT Luer cap in place 6 Gently inject saline through the catheter until it begin...

Page 4: ...indwelling catheter length using the centimeter marks on the catheter shaft as reference Frequent visual reassessment should be made to ensure that the catheter has not moved 26 Attach a primed Start Up Kit to the catheter by connecting the male Luer of the Start Up Kit to the female IN Luer of the catheter and the female Luer of the Start Up Kit to the male OUT Luer of the the catheter White ZOLL...

Page 5: ...t move the catheter if resistance is felt Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped If they are capped uncap them deflate the balloon and try removing the catheter again If resistance is still encountered an x ray should be performed to identify the reason for the resistance MRI Safety Information Non clinical testing has demonstrated that the ICY catheter is ...

Page 6: ...e guidewire at all times Dispenser Every guidewire is provided in a dispenser package Remove the guidewire anti migration clip before dispensing the guidewire Remove the guidewire protective cap immediately prior to guidewire use Prepare the guidewire prior to insertion It is recommended that the dispenser be filled with heparinized solutions e g saline or dextrose to bathe the guidewire during in...

Page 7: ...xchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 7 of 10 1 OUT 2 IN 3 Medial 4 Distal 5 Proximal 6 Detail view ICY IC 3893A Catheter Shaft 7 Balloon 5 mm x 80 5 mm 8 Balloon 8 mm x 80 5 mm Balloons ...

Page 8: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 8 of 10 ...

Page 9: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 9 of 10 ...

Page 10: ...ICY Intravascular Heat Exchange Catheter Custom Luer Instructions for Use Model IC 3893A 8700 0782 01 10 of 10 ICY SilvaSorb ...

Reviews: