Page 2-4
User Guide
10. Turn the Platform over and press the On/Off button to power it up. If the User Control Panel reports
a user advisory, check the installation of the LifeBand band clip into the slot in the driveshaft.
2.1.2
Removing the LifeBand
Note:
Please do NOT cut the LifeBand before removing it from the Platform. Cutting the LifeBand
may cause the Platform to report a Fault and will require specific steps to clear that Fault.
1.
Place the Platform with the patient surface facing down on a smooth, flat surface.
2.
Flip up the hinged belt guards of the LifeBand cover plate to disengage the LifeBand from the
rollers (see Figure 2-5). You will hear the guards “snap.” This is normal.
Figure 2-5
Flip Up the Hinged Belt Guards of the LifeBand
LifeBand Cover Plate
Belt Guard