    Department of Education and Professional Development 

 

 

 

 

Zoll AEDpro Information 

 

Summary of Contents for aed pro

Page 1: ...Department of Education and Professional Development Zoll AEDpro Information...

Page 2: ...of circulation as indicated by Unconsciousness Absence of breathing and Absence of pulse and other signs of circulation When the victim is less than 8 years old or weighs less than 55 lb 25 kg use ZO...

Page 3: ...ng Do not use electrodes that have passed their expiration date Do not use defibrillation pads or ECG electrodes if the gel is dried or damaged patient burns or poor quality ECG signals might result f...

Page 4: ...f and the unit resumes ECG analysis After delivering a shock the unit continues analyzing the patient s ECG guiding the rescuer to perform CPR or to deliver additional shocks if needed Manual Mode Def...

Page 5: ...itoring modes only Shows the current heart rate in beats per minute The symbol flashes with each detected heartbeat Soft key labels Labels for the soft keys appear at the bottom of the display directl...

Page 6: ...ard 2 Pull the top edge of the latch toward you and then downward Pull the centre of the lid upwards as shown 3 Lift the edge of the battery compartment cover 4 Slide out the cover panel If a battery...

Page 7: ...rtment latch and tilt the top edge toward the unit 9 Press the bottom edge of the latch toward the unit to lock the cover in place Important If you change the battery while the unit is in clinical use...

Page 8: ...t the defibrillation electrodes to the patient cable connector Note If electrodes are not preconnected the unit will fail its next self test in standby state 8 Press and release the On Off button to t...

Page 9: ...nd software Preconnecting the Defibrillation Electrodes Cable WARNING To prepare for an emergency keep the defibrillation electrodes cable connected to the unit at all times WARNING Do not reuse defib...

Page 10: ...hrough patient assessment by issuing voice and text messages the unit issues the voice and text prompt DON T TOUCH PATIENT ANALYZING The unit then begins analysis of the patient s ECG to determine whe...

Page 11: ...he electrode as you go 6 Follow the AED Pro prompts Applying Defibrillation Electrode Pads Infant Child WARNING ZOLL pedi padz are for paediatric patients only these pads provide defibrillation energy...

Page 12: ...ines whether or not the patient has a shockable rhythm and then displays and voices its recommendation SHOCK ADVISED or NO SHOCK ADVISED 4 After ECG analysis which message appears If SHOCK ADVISED con...

Page 13: ...e Replace the battery pack immediately ADULT PADS PEDIATRIC PADS The unit detected the specified type of electrode pads and adjusted defibrillation energy settings accordingly PLUG IN CABLE The unit s...

Page 14: ...ED KEEP PATIENT STILL ECG rhythm analysis halted due to excessive ECG signal artifact Stop CPR and keep the patient as motionless as possible SHOCK DELIVERED A shock was delivered to the patient NO SH...

Page 15: ...illator automatically disarms itself To recharge the defibrillator you must press the CHARGE soft key again In manual mode the screen shows the shock count elapsed time mode MANUAL heart rate heartbea...

Page 16: ...charge ready tone indicates that the defibrillator is charged and ready The Shock button flashes repeatedly WARNING You have 60 seconds to perform the following step otherwise the defibrillator automa...

Page 17: ...an implanted pacemaker ECG Electrode Placement Before applying ECG electrodes be sure to Remove all clothing covering the patient s chest Clip or shave any excessive hair to ensure proper adhesion of...

Page 18: ...s on screen until defibrillation pads are attached If the detected heart rate is lower than the configured minimum heart rate limit the unit issues the voice and text prompt CHECK PATIENT This message...

Page 19: ...ge SAVING DATA appears To cancel changes and revert to the previously set date and time press the X soft key Troubleshooting and Maintenance To ensure that the AED Pro unit is functioning properly and...

Page 20: ...iautomatic mode Release the soft key ECG Monitoring Troubleshooting Symptom Corrective Action Messages CHECK ECG ELECTRODES ATTACH ECG ELECTRODES Ensure that the ECG cable is connected to each electro...

Page 21: ...Shock advised Charging 2 Shout STAND CLEAR look for patient to be clear of anyone 3 Press flashing red button 4 Immediately resume CPR Zoll Defib quick guide ECG monitor 1 Press green on off button Or...

Page 22: ...breathing Plug in cable 2 Press 2 soft keys 3 Press MANUAL 4 Assess rhythm 5 Press charge button 6 Shout STAND CLEAR look for patient to be clear of anyone 7 Press flashing red button Warning Defib wi...