ZOLL Aed Plus Manuallines Download Page 4

 

Automated External Defibrillator (AED) Guidelines 

 

 

A

th

le

ti

cs

 

 

Patient Monitoring 

1.

 

Once the defibrillator has been applied to the patient, do not turn off the defibrillator or 
remove the defibrillation pads unless prompted by the device (e.g., “Replace battery” or 
“Replace pads”).  

2.

 

The defibrillator will continue background monitoring of the patient’s heart rhythm and 
alert the rescuers if additional shocks are required.  

3.

 

Continue to assess the patient’s airway, breathing, and circulation and provide CPR as 
indicated. 

 
Post Rescue: 

1.

 

Prepare the AED for the next rescue. 

2.

 

Cleaning  

a.

 

Clean and disinfect unit with soft, damp cloth using 90% isopropyl alcohol or 
soap and water, or chlorine bleach (30 ml/liter water).  

b.

 

Do not immerse any part of the unit in water.  

c.

 

Do not use ketones (MEK, acetone, etc.).  

d.

 

Avoid using abrasives (e.g., paper towels) on the LCD display, if so equipped.  

e.

 

Do not sterilize the unit. 

3.

 

Connect a new pair of pads to the AED, close the lid, and verify the green check symbol 
(

) appears in status indicator window. 

4.

 

Order and replace any equipment used during the rescue. 

 

Medical Response Documentation 

 

Each incident involving the use of the AED will be documented (Attachment A) to include date, 
time, and location, reason for use, outcome, responding EMS assistance, and equipment used. 

 

AED Self-Test System 

 

The AED has a comprehensive self-test system that automatically tests the electronics, battery, 
pads, and high voltage circuitry.  Self-tests are also activated every time the AED lid is opened 
and closed.   
 

AED Indicators and Scheduled Monitoring / Maintenance 

 

Status Indicator  

The readiness indicator is the green check symbol (

) appears in status indicator window.  The 

AED is Rescue Ready [battery has an adequate charge; pads are properly connected to the AED 
and functioning; and, integrity of the internal circuitry is good]. 
 
If the readiness indicator is not displayed, refer to the troubleshooting table below. 

 

Reviews: