background image

                Accurately imitating the AED Plus requires that instruc-

tors be familiar with the protocol currently followed by the 

AED Plus for 

PUSH HARDER 

and 

GOOD COMPRESSIONS

 prompts.

                For further information about the functioning of the  

clinical AED Plus, see the AED Plus Operator’s Guide and Admin-

istrator’s Guide.

               The Remote Controller lights indicate the current training

settings while the Trainer2 simulates the actual AED Plus behavior.

                The Remote Controller will turn itself off after 10 seconds. 

The Trainer2 will remember the last scenario setting which will re-

appear the next time the 

view/enter

 button is pushed.  

NOTE

SCENARIOS

VF 1

VF 2

VF 3

VF 4

Done

Shock

CPR

Shock

CPR

1

Simulated heart rhythm: Student Response:

ASYSTOLE 1

ASYSTOLE 2

VF 1

VF 2

Done

No Shock

CPR

Shock

CPR

2

Simulated heart rhythm: Student Response:

VF 1

VF 2

NSR 1

NSR 2

Done

Shock

CPR

No Shock

CPR

3

Simulated heart rhythm: Student Response:

ASYSTOLE 1

ASYSTOLE 2

ASYSTOLE 3

ASYSTOLE 4

Done

No Shock

CPR

No Shock

CPR

4

Simulated heart rhythm: Student Response:

 

 

 

 

 

 

 

 

 

SETUP and TEST

1

SELECT SCENARIO SETTINGS

2

Scenario 3: SHOCK - NO SHOCK

Scenario 2: NO SHOCK - SHOCK

Scenario 1: SHOCK - SHOCK

Manual Scenario mode

CONTROLS / INDICATORS

Handheld Remote Controller:

PUSH HARDER

 

prompt

(only when metronome is beeping)

NO

 

SHOCK

 advised on next heart analysis

(manual mode only)

Pads

 

attached/not attached to victim’s bare chest

(toggle-switch)

CPR Metronome

 

set up to beep or not beep

(Real CPR Help

®

 during CPR interval)

Push to turn Trainer2 on

Push again to start current scenario 

(as defined using Remote Controller)

Hold to turn Trainer2 off

ZOLL AED Plus Trainer:

View/Enter

 

scenario

Show current scenario selection (press once)

Start scenario (press again)

Move Up

 to next scenario.

Scenario 4: NO SHOCK - NO SHOCK

 

 

Scenario Set-up Controls

:

Training Controls:

GOOD COMPRESSIONS

 

prompt

(only when metronome is beeping)

Pause

 (resume) current scenario

 

End

 

current scenario

Green light remains on = Trainer2 is on and ready for use

Green light off = Trainer2 is turned off or has no battery power

Red Light steady = Last command from Remote Controller accepted

Red Light 3 flashes = Last command from Remote Controller is invalid

NOTE

Install 4 “D” batteries in the AED Plus Trainer2.  

Trainer2 unit:

Action:

Press 

On/Off

 button to turn Trainer2 on.  

Green light turns on, indicating Trainer2 is 

ready for use.  

Remote Controller:

Install 2 “AA” batteries in handheld remote. 

Turn the Remote Controller on by pushing the

view/enter

 button.  

Point Remote Controller at unit and press

view/enter

 button again.  Green light on Remote

Controller will show Manual Scenario Mode is

selected.

Manual Scenario Mode
View/Enter Button

Trainer2 unit:

Action:

Turn Trainer2 unit off (hold down 

On/Off

 button).  

Setup test is now complete.  Trainer2 unit and 

Remote Controller are ready for use.  

Trainer2 unit:

Action:

Turn Trainer2 on by pushing the 

On/Off

 button.  

Action:

Remote Controller:

Action:

Press 

view/enter 

button once to turn on Remote

Controller.

Use 

move up

 button to select desired scenario.

Use CPR Metronome button to turn Real CPR Help 

(beeping) on or off for this scenario.

Push 

view/enter

 button once to enter scenario for 

AED Plus Trainer2.

CPR PROMPTS

3a

Remote Controller:

Action:

The Remote Controller controls all prompts when

the Trainer2 is beeping (during the CPR interval).

NOTE

SCENARIO CONTROLS

3

Remote Controller:

Action:

VF (Ventricular Fibrillation-shock advised) is the 

default simulated heart rhythm in Manual Scenario.   

Analysis - VF

For the next analysis to result in NO Shock advised, 

push the

 NO Shock

 button. 

NOTE

                Once the No Shock button has been pushed, the next 

simulated heart analysis will advise NO Shock. This setting cannot be 

reversed for the next analysis.

In all scenarios the 

Pads 

buttton must be pressed 

once to proceed with analysis.  If pressed again, it 

will begin to prompt for attaching pads.

PAUSE/STOP SCENARIO

3b

Press the 

PUSH HARDER 

button during the CPR 

interval to force the unit to say “Push Harder”.
Press the 

GOOD COMPRESSIONS 

button during the 

CPR interval to force the unit to say “Good 

Compressions”.

Remote Controller:

Action:

In all scenarios (manual and automatic), the rescue

can be paused by pushing the 

Pause/Resume

 

button once. Push it a second time to resume the 

scenario. 

Trainer2 unit:

Action:

The PUSH HARDER and GOOD COMPRESSIONS 

buttons are only valid during the CPR interval. 

The red light on the Trainer2’s status window emits 

one long flash if the prompt is currently valid.

If the PUSH HARDER or GOOD COMPRESSIONS

buttons are pushed when the Trainer2 is not 

beeping, the unit’s red light emits three quick

flashes, indicating that the prompt is currently

invalid.

In all scenarios (manual and automatic), the rescue

can be ended by pushing the End button once.

NOTE

4

On

Off

On

Off

On

Off

On

Off

Reviews: