background image

      Additional Remote Controllers are available.

      Contact ZOLL Medical Corporation or your distributor.

1. 

Keep Trainer2 and Remote Controller clean, undamaged, 

 

and free of excessive wear.

2. 

Verify that there are no cracks or loose parts in Trainer2 and

 Remote 

Controller.

3. 

Replace Training electrodes as necessary.

MAINTENANCE

Trainer2 does not respond

(green light in status window

is not lit) when On/Off

Button is pushed

TROUBLESHOOTING

Clean the Trainer2 unit and remote controller with soft, damp

cloth using soap and water, then wipe dry immediately.

Do not immerse any parts in water.

Do not clean with ketones (e.g., MEK, acetone).
Avoid using abrasives on any surface.
Do not sterilize the Trainer2 or the Remote Controller.

CLEANING

TRAINING USE ONLY - NOT FOR CLINICAL USE

REF 9650-0090-01 Rev. C

Product intended for indoor use only.

Do not attempt to plug AED Plus or AED Pro

simulator into connector of Trainer2.

This device is for training only. It will not deliver defibrillation energy.

Replace all four D batteries

in Trainer2 unit.

Remote Controller does not 

respond (scenario 

selection is not lit) 

when View/Enter

Button is pushed.

Replace both AA batteries in

Remote Controller.

With new batteries in both, 

Trainer2 fails to respond to 

Remote Controller. 

AED Plus Trainer2 problem/

display:

          

Action:

Contact ZOLL 

Technical Service.

Do not attempt to take Trainer2 or Remote Controller apart. 

Trainer2 does not beep or

react to Remote Controller

during CPR interval. 

Re-set scenario and make

certain that CPR metronome

light is on.

Trainer2 announces breathing

prompts, OPEN AIRWAY and
CHECK BREATHING.

Unit is configured for ERC. 

Change to AHA configuration.

(See Switching AHA/ERC 

Configuration section of this

Operator’s Guide).

SWITCHING AHA/ERC CONFIGURATION     

American Heart Association (AHA) configuration:  No prompts for 

breathing.

European Resuscitation Council (ERC) configuration: Contains breathing 

prompts OPEN AIRWAY and CHECK BREATHING.

To switch between either AHA or ERC configuration, do the following:

With the Trainer2 unit off, hold down the Shock Button. Press On/Off Button 

once.  Green light will illuminate at status window to indicate AHA 

cofiguration (prompt sequence will contain no breathing prompts).

Press Shock Button again.  Green light will turn off and red light will illuminate

at status window to indicate ERC configuration (prompt sequence will 

contain breathing ).

Press Shock Button again to revert to AHA configuration (green light on 

and red light off).

When configured as desired (green = AHA, red = ERC), hold down On/Off

Button to turn unit off to store configuration.

WARNINGS

Remove battery compartment doors only to replace batteries. 

 

Medical Electrical Equipment needs special precautions regarding  

EMC and needs to be installed and put into service according to the 

EMC information provided in accompanying documents. 
Portable and Mobile RF Communications equipment can affect 

 Medical Electrical Equipment. 

SYMBOLS

The following symbols appear on the ZOLL AED Trainer2:

Temperature limitation.

Manufacturer.

Authorized representative in the European Community.

Caution.

Return to a collection site intended for waste electrical and electronic

equipment (WEEE).  Do not dispose of in unsorted trash.

REF

Catalogue Number.

NOTE 

 

Conformité Européene - Complies with Low Voltage Directive 73/23/EEC. 

 

 

 

ZOLL Medical Europe

Newtonweg 18

6662 PV  ELST

The Netherlands

Tel:  +31 (0) 481 366410

Fax: +31 (0) 481 366411

 

 

 

 

International Service: Contact your  

local distributor. 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

VF 1 

VF 2 

VF 3 

VF 4 

Done 

Shock 

CPR 

Shock 

CPR 

Simulated heart rhythm:  Student Response: 

ASYSTOLE 1 

ASYSTOLE 2 

VF 1 

VF 2 

Done 

No Shock 

CPR 

Shock 

CPR 

Simulated heart rhythm:  Student Response: 

VF 1 

VF 2 

NSR 1 

NSR 2 

Done 

Shock 

CPR 

No Shock 

CPR 

Simulated heart rhythm:  Student Response: 

ASYSTOLE 1 

ASYSTOLE 2 

ASYSTOLE 3 

ASYSTOLE 4 

Done 

No Shock 

CPR 

No Shock 

CPR 

Simulated heart rhythm:  Student Response: 

SCENARIOS 

 

 

ACCESSORIES

You can order the following Trainer2 accessories from the ZOLL 

Customer Service Department or your distributor:

 

Item: 

      REF

1.  CPR-D padz® Training Electrodes (1 Set)  

8900-0804-01

2.  CPR-D padz Training Gel (5 Pair)   

 

8900-0803-01

3.  stat-padz® Training Electrodes (1 Set) 

 

8900-0805-01

4.  pedi-padz® II Training Electrodes (case of 6) 

8900-000861-01

5.  Trainer2 Remote Controller 

 

 

8008-0007

6.  Trainer Carrying Case 

 

 

 

8000-0375-01

7.  Trainer Demo Manikin 

 

 

 

8000-0835-01

ZOLL, AED Plus, Real CPR Help, CPR-D padz, stat-padz and pedi-padz are trademarks or 

registered trademarks of ZOLL Medical Corporation in the United States and/or other countries. 
Copyright 2012 by ZOLL Medical Corporation

.  

All rights reserved.

© 2012 ZOLL Medical Corporation 

On 
Off 

For technical support or repair: 

 

ZOLL Medical Corporation 

269 Mill Road 

Chelmsford, MA USA 01824-4105 

 

Attn:  Technical Service Department 

 

Tel:  978-421-9655     800-348-9011 

Fax:  978-421-0001 

 

 

 

 

AED Plus

Trainer2

Automated External Defibrillator  Trainer2

Operator’s Guide

AHA/ERC 2010

®

Reviews: