Product Overview
2-6
www.zoll.com
906-0330-01-01 Rev. D
Parameter Windows
These windows provide control of parameters depending on the mode you select.
When you select Continuous mode, you also specify the following parameter options:
When you select Intermittent mode, you also specify the following parameters options:
Shared Icon Area
Directly below the message area, the device displays icons that indicate:
•
Power source (external power or internal battery)
•
Battery Charge Status
•
Alarm Muted/Audible
•
Alarm Priority
Continuous Mode
Range (mm Hg)
(kPa)
Surgical
20 to 550 mm Hg
3 to 73
Pharyngeal
60 to 400 mm Hg
8 to 53
Tracheal
60 to 220 mm Hg
8 to 28
CASS
10 to 40 mm Hg
1.3 to 5.3
Note: Smart Flow (On or Off) is also a selectable option when in Continuous Mode.
Intermittent Mode
Range
kPa
Vacuum Level
10 to 200 mm Hg
1 to 27 kPa
Time On/Off
5 to 40 seconds
Not Applicable