Cannot feel RF, will not warm up
Pause button being
pressed down.
The thermo stat function
is working.
RF output head is not firmly
in contact with the skin.
Skin is too dry.
Apply the gel on your skin.
Place the RF output head
on your skin the gel applied.
The temperature sensor in the center of treatment
head senses skin temperature in range of
40 degrees.Therefore, sudden temperature
chage will not happen. It warms up gradually.
Too warm
Skin is too dry.
The device has been used
on the same area for
over 3 minutes continuously.
Thermostat is not working.
27
Press to resume.
Apply the gel on your skin.
Change the part of treatment.
If there is residual dirt, oil or gel on
the thermostat, the thermal sensor
will not be able to sense the temperature
correctly, giving rise to risks of low
temperature burns.After each use, please
clean with cotton bud dampened with water.
Should the anomalies continues after trying the remedies, stop
using the device immediately and contact the dealer
When requesting for maintenance and inspection,
please present the warranty card.