32
Trouble Shooting
What It Means
What To Do
1. The meter does not enter test mode after
inserting a
test strip.
• Test strip is inserted upside down, or
bottom end in, or not fully inserted into the
meter.
Insert test strip print side up, top end in.
• No battery is installed
• Battery is installed incorrectly
Install battery with (+) facing up.
Dead battery
Replace battery. Reset date and time, if
necessary.
If meter still does not enter test mode,
contact Zoetis Customer Care.
• Defective test strip or meter
• Blood or foreign objects put in the test strip
port
Contact Zoetis Customer Care.
2. The test does not start after applying the
blood drop.
• Blood drop is too small
Add blood to the same sample area within
60 seconds, or repeat the test using a new
test strip and a larger blood drop.
• Defective test strip
• Sample applied after meter turns off
Repeat the test using a new test strip.
Wait for drop and strip to appear before
applying blood or control solution.
If problem persists, contact Zoetis
Customer Care.
• Defective meter
Contact Zoetis Customer Care.